MANUAL ADDENDUM
M200AU
NOy CONVERTER OPTION
(for use with M200AU Operators Manual, 02293)
Teledyne Advanced Pollution Instrumentation
9480 Carroll Park Dr.
San Diego, CA 92121-5201
TOLL-FREE: 800-324-5190
FAX: 858-657-9818
TEL: 858-657-9800
E-MAIL:
WEB SITE: www.teledyne-api.com
02808
REV. E
Copyright 2007 API Inc. July 2007
Page view 0
1 2 3 4 5 6 ... 36 37

Summary of Contents

Page 1 - NOy CONVERTER OPTION

MANUAL ADDENDUM M200AU NOy CONVERTER OPTION (for use with M200AU Operators Manual, 02293) Teledyne Advanced Pollution Instrumentati

Page 2 - SAFETY MESSAGES

Model 200AU-NOy Option Manual Addendum 1.3 Initial Operation 1. After confirming proper supply voltage, turn on the instrument power. If you are

Page 3 - Table of Contents

Model 200AU-NOy Option Manual Addendum 2.0 Specifications, Warranty 2.1 Specifications Converter Remotely mounted molybdenum converter with tempera

Page 4

Model 200AU-NOy Option Manual Addendum 2.2 Warranty ADVANCED POLLUTION INSTRUMENTATION, INC. 02024c Prior to shipment, API equipment is thoroughly

Page 5 - List of Figures

Model 200AU-NOy Option Manual Addendum 3.0 The M501Y NOy Converter 3.1 Principle of Operation The API Model 200AU with NOy Option is designed to me

Page 7 - 1.0 Getting Started

Model 200AU-NOy Option Manual Addendum 4.0 CALIBRATION AND ZERO/SPAN CHECKS Unlike most NOx analyzers, the M200AU with the NOy option does not have

Page 8

Model 200AU-NOy Option Manual Addendum NOTE The M200AU with the NOy Option does NOT have equivalency approval, and may not be used for EPA monitorin

Page 9 - 02808 Rev. E Page 9

Model 200AU-NOy Option Manual Addendum 4.1 Calibration or Cal Check Procedure The calibration of the instrument can be checked or adjusted using gas

Page 10 - 1.3 Initial Operation

Model 200AU-NOy Option Manual Addendum Enter the expected NOx and NO span gas concentrations: Table 4.2 - Enter Expected Span Gas Concentrations Pr

Page 11 - 2.0 Specifications, Warranty

Model 200AU-NOy Option Manual Addendum Table 4.3 - Span Calibration Procedure Step Number Action Comment 1. Press CAL The M200AU enters the cali

Page 12 - 2.2 Warranty

Model 200AU-NOy Option Manual Addendum SAFETY MESSAGES Your safety and the safety of others is very important. We have provided many important saf

Page 14

Model 200AU-NOy Option Manual Addendum 02808 Rev. E Page 21 5.0 Maintenance 5.1 Maintenance Schedule See Table 5.1 below.

Page 16

Page 23 Table 5.1- Preventative Maintenance Calendar Model 200AU-NOy Option Manual Addendum02808 Rev. E Item Jan Feb M

Page 17

Model 200AU-NOy Option Manual Addendum 5.2 Replacing the Sample Particulate Filter The particulate filter should be inspected often for signs of plu

Page 18

Model 200AU-NOy Option Manual Addendum Figure 5.1 - Replacing the Particulate Filter 02808 Rev. E Page 25

Page 19

Model 200AU-NOy Option Manual Addendum 5.3 Check Bypass Flow Rates The external sample pump is capable of maintaining the cell pressure at less tha

Page 20

Model 200AU-NOy Option Manual Addendum Sample In NO Out NOy Out Figure 5.2 - Converter Assembly 02808 Rev. E Page 27

Page 21 - 5.0 Maintenance

Model 200AU-NOy Option Manual Addendum 5.5 Pneumatic Line Inspection Particulate matter in the pneumatic lines will affect both flow rate and respon

Page 22

Model 200AU-NOy Option Manual Addendum Figure 5.3 - Pneumatic Diagram Figure 5.3 - Converter Assembly 02808 Re

Page 23

Model 200AU-NOy Option Manual Addendum Table of Contents SAFETY MESSAGES...

Page 24

Model 200AU-NOy Option Manual Addendum 5.6 Leak Check Procedure If a leak checker is not available, it is possible to leak check the instrument usi

Page 25

Model 200AU-NOy Option Manual Addendum 6.0 Troubleshooting, Adjustments NOTE THE OPERATIONS OUTLINED IN THIS CHAPTER ARE TO BE PERFORMED BY QUALI

Page 26 - 5.4 Replacing the Converter

Model 200AU-NOy Option Manual Addendum 6.1.2 Leak Check Refer to Section 5.6 for the leak check procedure. 6.1.3 Bypass Pump Diagnostic Procedures

Page 27

Model 200AU-NOy Option Manual Addendum 4. Press the "up-arrow" under the digit or the "down-arrow" at the left to scroll the di

Page 28

Model 200AU-NOy Option Manual Addendum SEL END RETURNS TO SV. SET TO 315 (315oC SET VALUE) NOTE: DO NOT SET THE TEMPERATURE (SV) HIGHER THAN 320oC

Page 29

Model 200AU-NOy Option Manual Addendum Figure 6.1 – Electrical Diagram 02808 Rev. E Page 35

Page 31 - 6.1.1 Pneumatic System

Model 200AU-NOy Option Manual Addendum 7.0 M501Y Spare Parts List Note: Use of replacement parts other than those supplied by API may result in non

Page 32 - 6.1.2 Leak Check

Model 200AU-NOy Option Manual Addendum 6.1 Operation Verification - M501Y Diagnostic Techniques...

Page 33 - PRESS DISPLAY ACTION

Model 200AU-NOy Option Manual Addendum List of Figures FIGURE 1.1 - REAR PANEL PNEUMATIC CONNECTIONS ...

Page 34

Model 200AU-NOy Option Manual Addendum 02808 Rev. D List of Tables TABLE 2.1 - FINAL TEST AND CALIBRATION VALUES...

Page 35

Model 200AU-NOy Option Manual Addendum 1.0 Getting Started The M501Y has been designed to operate with the M200AU low level NOx analyzer. The M200A

Page 36

Model 200AU-NOy Option Manual Addendum WARNING • Lethal voltages are present inside case • Do not operate with cover off during normal operati

Page 37 - 7.0 M501Y Spare Parts List

Model 200AU-NOy Option Manual Addendum Figure 1.1 - Rear Panel Pneumatic Connections 02808 Rev. E Page 9

Comments to this Manuals

No comments