MODEL T100H
UV FLUORESCENCE SO
2
ANALYZER
Addendum to T100 Operation Manual, PN 06807
© TELEDYNE ADVANCED POLLUTION INSTRUMENTATION
9480 CARROLL PARK DRIVE
SAN DIEGO, CA 92121-5201
USA
Toll-free Phone:
800-324-5190
Phone:
858-657-9800
Fax:
858-657-9816
Email:
api-sales@teledyne.com
Website:
http://www.teledyne-api.com/
Copyright 2011 PN 07265A DCN6038
Teledyne Advanced Pollution Instrumentation 25 March 2011
Page view 0
1 2 3 4 5 6 ... 69 70

Summary of Contents

Page 1 - ANALYZER

MODEL T100H UV FLUORESCENCE SO2 ANALYZER Addendum to T100 Operation Manual, PN 06807 © TELEDYNE ADVANCED POLLUTION INSTRUMENTATION 9480 CA

Page 2

TABLE OF CONTENTS Teledyne API - T100H Addendum to T100 Operation Manual x 10.3.2. Sensor Module Repair & Cleaning...

Page 3 - SAFETY MESSAGES

11 1. INTRODUCTION This addendum is based on the Model T100 Operators Manual (P/N 06807). In most ways the T100H analyzer is identical to the T1

Page 4 - CONSIGNES DE SÉCURITÉ

Introduction Teledyne API - T100H Addendum to T100 Operation Manual) 12 Operation Instructions This section includes step by step instructions for

Page 5 - WARRANTY

13 2. SPECIFICATIONS, APPROVALS & WARRANTY 2.1. Specifications Table 2-1: Model T100H Basic Unit Specifications Min/Max Range (Physical

Page 6

Specifications, Approvals & Warranty Teledyne API - T100H Addendum to T100 Operation Manual) 14 2.2. CE Mark Compliance Emissions Compliance T

Page 7 - REVISION HISTORY

15 3. GETTING STARTED 3.1. Unpacking and Initial Setup CAUTION To avoid personal injury, always use two persons to lift and carry the Model T10

Page 8

Getting Started Teledyne API - T100H Addendum to T100 Operation Manual) 16 3.1.1. Electrical Connections: For full details on the electrical conne

Page 9 - TABLE OF CONTENTS

Teledyne API - T100H Addendum to T100 Operation Manual Getting Started 17 3.2. Pneumatic Connections NOTE To prevent dust from getting into the ana

Page 10 - LIST OF APPENDICES

Getting Started Teledyne API - T100H Addendum to T100 Operation Manual) 18 3.2.1.2. Connections with Internal Valve Options Installed  There is

Page 11 - 1. INTRODUCTION

Teledyne API - T100H Addendum to T100 Operation Manual Getting Started 19 3.2.2. T100H Layout FLOW CONTROL & PRESSURE SENSOR VACUUM PRESSURE SE

Page 13 - 2.1. Specifications

Getting Started Teledyne API - T100H Addendum to T100 Operation Manual) 20 Figure 3-4: T100H Layout (Basic Unit – No Valve Options) 07265A DC

Page 14 - 2.2. CE Mark Compliance

Teledyne API - T100H Addendum to T100 Operation Manual Getting Started 21 3.3. Initial Operation With the following exceptions, the operation of th

Page 15 - 3. GETTING STARTED

Getting Started Teledyne API - T100H Addendum to T100 Operation Manual) 22 3.3.3. Interferents for SO2 Measurements Hydrocarbons are a significan

Page 16 - 3.1.1.1. External Pump

23 4. FREQUENTLY ASKED QUESTIONS (FAQs) More FAQs are included in the T100 manual, which also includes a glossary of terms. Q: How long does th

Page 17

Frequently Asked Questions (FAQs) Teledyne API - T100H Addendum to T100 Operation Manual) 24 This page intentionally left blank.07265A DCN6038

Page 18 - SAMPLE Gas

25 5. OPTIONAL HARDWARE AND SOFTWARE With the following additions, changes and exceptions, the options listed in Table 1-1 of the T100 Manual (P

Page 19 - 3.2.2. T100H Layout

Optional Hardware and Software Teledyne API - T100H Addendum to T100 Operation Manual) 26 5.2. Ambient Zero / Two Ambient Span Valve Option (OPT 5

Page 20

Teledyne API - T100H Addendum to T100 Operation Manual Optional Hardware and Software 27 Table 5-1 describes the state of each valve during the an

Page 21 - 3.3. Initial Operation

Optional Hardware and Software Teledyne API - T100H Addendum to T100 Operation Manual) 28 between the inner and outer tubes is small and the scrub

Page 22 - Measurements

29 6. OPERATING INSTRUCTIONS 6.1. Warning Messages Please refer to the T100 User’s Manual (06807), Section 4.2.2, for a list of warnings for the

Page 23

iii SAFETY MESSAGES Important safety messages are provided throughout this manual for the purpose of avoiding personal injury or instrument damage.

Page 24

Operating Instructions Teledyne API - T100H Addendum to T100 Operation Manual) 30 6.2.2. Range Units The T100H only displays concentrations in pa

Page 25 - INSTRUMENT CHASSIS

Teledyne API - T100H Addendum to T100 Operation Manual Operating Instructions 31 6.2.4. Default DAS Channels The default Data Channels included in

Page 26

Operating Instructions Teledyne API - T100H Addendum to T100 Operation Manual) 32 6.2.5.2. Control Inputs ZERO CALHI SPAN CALZERO CALCONTROL IN

Page 27

33 7. CALIBRATION AND CALIBRATION CHECK PROCEDURES Calibration procedures for the T100H are the same as those for the T100. One exception to t

Page 28

Calibration and Calibration Check Procedures Teledyne API - T100H Addendum to T100 Operation Manual) 34 STEP THREE: Perform the calibration accor

Page 29 - 6. OPERATING INSTRUCTIONS

Teledyne API - T100H Addendum to T100 Operation Manual Calibration and Calibration Check Procedures 35 7.2. Manual Calibration Check with Ambient

Page 30 - 6.2.2. Range Units

Calibration and Calibration Check Procedures Teledyne API - T100H Addendum to T100 Operation Manual) 36 This page intentionally left blank.07265A

Page 31 - 6.2.4. Default DAS Channels

37 8. INSTRUMENT MAINTENANCE 8.1. Maintenance Schedule There is no Internal IZS offered for the T100H. 8.2. Predictive Diagnostics Because the T

Page 32

Instrument Maintenance Teledyne API - T100H Addendum to T100 Operation Manual) 38 This page intentionally left blank.07265A DCN6038

Page 33 - PROCEDURES

39 9. THEORY OF OPERATION 9.1. The UV Light Path The UV light path of the T100H is similar to that of the T100 (see Section 10.2 of the T100 Man

Page 34

Teledyne API - T100H Addendum to T100 Operation Manual iv CONSIGNES DE SÉCURITÉ Des consignes de sécurité importantes sont fournies tout au long

Page 35

Theory Of Operation Teledyne API - T100H Addendum to T100 Operation Manual) 40 9.1.1. The Reference Detector A vacuum diode UV detector that conve

Page 36

Teledyne API - T100H Addendum to T100 Operation Manual Theory Of Operation 41 FLOW CONTROL & PRESSURE SENSOR VACUUM PRESSURE SENSORSAMPLE PRESS

Page 37 - 8. INSTRUMENT MAINTENANCE

Theory Of Operation Teledyne API - T100H Addendum to T100 Operation Manual) 42 9.2.2.3. Sample Flow Calculation Unlike the T100, which uses a the

Page 38

Teledyne API - T100H Addendum to T100 Operation Manual Theory Of Operation 43 9.3. Electronic Operation The following figures replace Figures 10-10

Page 39 - 9. THEORY OF OPERATION

Theory Of Operation Teledyne API - T100H Addendum to T100 Operation Manual) 44 AC POWER ENTRANCE ON/OFF SWITCH Sample Gas Pressure Sensor Motherb

Page 40 - 9.2.1. Sample Gas Flow

45 10. TROUBLESHOOTING & REPAIR For the most part the information contained in Section 11 of the T100 Manual (P/N 06807) is also applicable

Page 41 - 9.2.2. Pneumatic Sensors

TROUBLESHOOTING & REPAIR Teledyne API - T100H Addendum to T100 Operation Manual) 46 10.3. Repair Procedures 10.3.1. Repairing the Sample Gas F

Page 42

47 10.3.2. Sensor Module Repair & Cleaning NOTE: After any repair or service has been performed on the sensor module, the T100H should be a

Page 43 - 9.3. Electronic Operation

TROUBLESHOOTING & REPAIR Teledyne API - T100H Addendum to T100 Operation Manual) 48 10.3.2.1. Adjusting the UV Lamp (Peaking the Lamp) There a

Page 44

Teledyne API - T100H Addendum to T100 Operation Manual TROUBLESHOOTING & REPAIR 49 Table 10-3: Example of HVPS Power Supply Outputs UVLAMP_SI

Page 45 - 10.2. Subsystem Checkout

v WARRANTY WARRANTY POLICY (02024D) Prior to shipment, T-API equipment is thoroughly inspected and tested. Should equipment failure occur, T-API a

Page 46 - 10.3. Repair Procedures

TROUBLESHOOTING & REPAIR Teledyne API - T100H Addendum to T100 Operation Manual) 50 10.3.2.2. PMT Hardware Calibration (FACTORY CAL) The senso

Page 47

Teledyne API - T100H Addendum to T100 Operation Manual TROUBLESHOOTING & REPAIR 51 8. Set the HVPS coarse adjustment to its minimum setting (0

Page 48 - (Peaking the Lamp)

TROUBLESHOOTING & REPAIR Teledyne API - T100H Addendum to T100 Operation Manual) 52 20. Adjust gain adjustment potentiometer until the NORM P

Page 49

Models T100H, 100EH Addendum APPENDIX A - Version Specific Software Documentation (05244B DCN6038) A-1 APPENDIX A - Version Specific Software Doc

Page 50

APPENDIX A - Version Specific Software Documentation(05244B DCN6038) Model 100EH Instruction Manual A-2 07265A DCN6038

Page 51

Models T100H, 100EH Addendum APPENDIX A - Version Specific Software Documentation (05244B DCN6038) A-3 APPENDIX A-1: Software Menu Trees, Rev C.0

Page 52 - 10.4. Technical Assistance

APPENDIX A - Version Specific Software Documentation (05244B DCN6038) Model 100EH Instruction Manual A-4 LOW HIGH ZERO SPAN CONC SAMPLE <TST

Page 53

Models T100H, 100EH Addendum APPENDIX A - Version Specific Software Documentation (05244B DCN6038) A-5 CFG MODEL NAME SERIAL NUMBER SOFTWA

Page 54

APPENDIX A - Version Specific Software Documentation (05244B DCN6038) Model 100EH Instruction Manual A-6 NAME EVENT PARAMETERS REPORT PERIOD NUM

Page 55

Models T100H, 100EH Addendum APPENDIX A - Version Specific Software Documentation (05244B DCN6038) A-7 QUIET COMPUTER SECURITY HESSEN PROTOCOL

Page 56

Teledyne API - T100H Addendum to T100 Operation Manual vi This page intentionally left blank. 07265A DCN6038

Page 57

APPENDIX A - Version Specific Software Documentation (05244B DCN6038) Model 100EH Instruction Manual A-8 INSTRUMENT IP5 GATEWAY IP5 SUBNET MASK5

Page 58 - (iDAS)

Models T100H, 100EH Addendum APPENDIX A - Version Specific Software Documentation (05244B DCN6038) A-9 . YES NO PMT DET WARNINGUV LAMP WARNI

Page 59

APPENDIX A - Version Specific Software Documentation (05244B DCN6038) Model 100EH Instruction Manual A-10 5) SAMPLE LED 6) CAL LED 7) FAULT LED

Page 60

APPENDIX B - Spare Parts Note Use of replacement parts other than those supplied by Teledyne Advanced Pollution Instrumentation (TAPI) may result

Page 61 - (see Section 6.12.4.6)

This page intentionally left blank. B-207265A DCN6038

Page 62

069280000 LIST, SPARE PARTS, T100H 3/24/2011 11:04:39 AM PARTNUMBER DESCRIPTION000940400 CD, ORIFICE, .004 BLUE

Page 63 - APPENDIX B - Spare Parts

068810000 PCA, LVDS TRANSMITTER BOARD 069340100 DOM, w/SOFTWARE, STD, T100H * 069500000 PCA, SERIAL & VIDEO INTERFACE

Page 64

Recommended Spare Parts Stocking Levels (Reference 04527 24Mar2011)Part Number Description1 2-5 6-10 11-20 21-30 UNITS000940800 CD, OR

Page 65

Part NumberDescription Quantity009690100 KIT, TFE FILTER ELEMENTS, 47MM, 5UM (30) 1018080000 AKIT, DESSICANT BAGGIES (12) 1FL0000001 FILTER, SS2HW0000

Page 66

Appendix C Warranty/Repair Questionnaire Models T100H, 100EH (05029D DCN 6038) TELEDYNE API CUSTOMER SERVICE EMAIL: [email protected]

Page 67 - 1 2-5 6-10 11-20 21-30 UNITS

vii ABOUT THIS ADDENDUM This T100H addendum, PN 07265, is to be used in conjunction with the T100 manual, PN 06807; it is comprised of multiple

Page 68 - Description Quantit

Appendix C Warranty/Repair Questionnaire Models T100H, 100EH (05029D DCN 6038) TELEDYNE API CUSTOMER SERVICE EMAIL: [email protected]

Page 69 - Models T100H, 100EH

viii This page intentionally left blank. 07265A DCN6038

Page 70

ix TABLE OF CONTENTS 1. INTRODUCTION...........................

Comments to this Manuals