iTeledyne Analytical InstrumentsTED 200-TPORTABLE OXYGEN MONITORTYPE B EQUIPMENT: Equipment providing a particular degree of protec-tion against elect
1 Introduction1-4Teledyne Analytical Instruments
Operations 22-1Teledyne Analytical InstrumentsOperationsNote: Upon receipt, INSPECT THE ENTIRE UNIT FOR DAMAGE. Check the unitand all included access
2 OperationsTeledyne Analytical Instruments2-2Figure 2: TED 200-T rear viewSensor cable receptacleBattery compartment2.1.1 Battery Installation or Re
Operations 22-3Teledyne Analytical InstrumentsFigure 3: Mounting the sensor in the tee adapter6. Close the compartment door. The front panel will now
2 OperationsTeledyne Analytical Instruments2-42. Plug one end of the coiled cable into the telephone jack receptacle on the endof the T-7 sensor. The
Operations 22-5Teledyne Analytical Instruments2. If calibrating for the first time, press the CAL key once. After initialcalibration, press the
2 OperationsTeledyne Analytical Instruments2-6NOTE:The HI alarm may be defeated by setting the limit one step above 100%.The display will show HI ALA
Operations 22-7Teledyne Analytical InstrumentsWhen monitoring for oxygen in breathing circuits, the diverter must be used. Thediverter should be scre
2 OperationsTeledyne Analytical Instruments2-8Gas or Vapor Level(Balance: Mixture of 30% O2 / 70% N20, except where noted)OxygenGas or Vapor Test Le
Operations 22-9Teledyne Analytical Instruments2.3 Do’s and Don’ts– DO –• Read all of the directions before using for the first time.• Calibrate every
Teledyne Analytical InstrumentsiiCopyright © 1999 Teledyne Electronic TechnologiesAll Rights ReservedNo part of this manual may be reproduced, transmi
2 OperationsTeledyne Analytical Instruments2-10
Maintenance and Troubleshooting 33-1Teledyne Analytical InstrumentsMaintenance and Troubleshooting3.1 TED 200-T Overall MaintenanceThe TED 200-T unit
3 Maintenance and TroubleshootingTeledyne Analytical Instruments3-2If the unit is left on after this, the LCD display will eventually shut down, the
Maintenance and Troubleshooting 33-3Teledyne Analytical InstrumentsSymptomA new sensor will not calibratein air. The TED 200-T displayflashes REPLACE
3 Maintenance and TroubleshootingTeledyne Analytical Instruments3-4SymptomThe oxygen reading fluctu-ates or appears to be incor-rect.The unit appears
Maintenance and Troubleshooting 33-5Teledyne Analytical InstrumentsSymptomAlarm sounds/flashes con-tinually, but keys are func-tioning normally.No di
3 Maintenance and TroubleshootingTeledyne Analytical Instruments3-6
AppendicesA-1Teledyne Analytical InstrumentsAppendicesA.1 SpecificationsRange: 0-100% oxygenAccuracy:+2% of full scale (at constant temperature)Respo
AppendicesA-2Teledyne Analytical Instruments1 B384 Battery holder1 C885 Cable assembly1 A268 Tee adapter (22 mm)A.3 Optional Accessories1 B34102 Mount
AppendicesA-3Teledyne Analytical InstrumentsA.4 Repair ServicePLEASE READ CAREFULLYIn the event that your TED 200-T Portable Oxygen Monitor needs serv
iiiTeledyne Analytical InstrumentsWarrantyTeledyne warrants that the goods are free from defects of material and of constructionfor a period of 2 year
AppendicesA-4Teledyne Analytical InstrumentsA.5.2 HumidityHumidity does not directly affect the accuracy of the sensor's measurement.However, whe
AppendicesA-5Teledyne Analytical InstrumentsThe BATT TEST control of the TED 200-T immediately determines if thebatteries need replacing. The TED 200-
AppendicesA-6Teledyne Analytical Instruments
AppendicesA-7Teledyne Analytical InstrumentsA.6 Material Safety Data SheetSection I – Product IdentificationProduct Name: Micro–Fuel CellsMini–Micro–F
AppendicesA-8Teledyne Analytical InstrumentsSection III – Physical HazardsPrimary route of entry: Ingestion, eye/skin contactExposure limits: OSHA PEL
AppendicesA-9Teledyne Analytical InstrumentsNOTE:The oxygen sensors are sealed, and under normal circumstances, thecontents of the sensors do not pres
Teledyne Analytical InstrumentsivTable of Contents1 Introduction1.1 Applications ...
vTeledyne Analytical InstrumentsHow To Use This ManualThis manual is designed to walk you through the initial set-up of the TED 200-T.After you have u
Teledyne Analytical Instrumentsvi
Introduction 11-1Teledyne Analytical InstrumentsIntroductionThe Teledyne Electronics Technologies 200-T Portable Oxygen Monitor is aneasy-to-use, po
1 Introduction1-2Teledyne Analytical InstrumentsThe monitor may be used for verifying oxygen concentrations in gas mixturesused in:• Anesthesia• Resp
Introduction 11-3Teledyne Analytical InstrumentsCAUTION: Remove and save the diverter when the sensor is used inconfined volume applications.1.2.2 Si