Portable Oxygen Monitor
Teledyne Analytical Instruments i
OPERATING & SERVICE INSTRUCTIONS FOR
MX300-I
PORTABLE OXYGEN MONITOR
0086
TYPE B APPLIED PART:
Equipment provides a degree of protection against electric shock,
particularly regarding—
Allowable Leakage Current
Reliability of the protective earth connection (if present).
P/N M75707
Sep 18, 2009
E
CO 09-015
7
Page view 0
1 2 3 4 5 6 ... 40 41

Summary of Contents

Page 1 - MX300-I

Portable Oxygen Monitor Teledyne Analytical Instruments i OPERATING & SERVICE INSTRUCTIONS FOR MX300-I PORTABLE OXYGEN MONITOR 0086

Page 2 - All Rights Reserved

Introduction MX300-I Teledyne Analytical Instruments 2 A sensor failure alarm is incorporated which warns the user if the sensor signal is lost or l

Page 3 - About This Manual

Portable Oxygen Monitor Introduction Teledyne Analytical Instruments 3 NOTE: IF THE LOW ALARM CONDITION IS TRIGGERED WITH THE OXYGEN LEVEL STAYI

Page 4 - Safety Messages

Introduction MX300-I Teledyne Analytical Instruments 4  Rugged high impact ABS plastic construction  Splash resistant case  Long life class R1

Page 5 - Table of Contents

Portable Oxygen Monitor Introduction Teledyne Analytical Instruments 5 The Class R17MED oxygen sensor is a disposable sensor. The sen-sor has a c

Page 6

Introduction MX300-I Teledyne Analytical Instruments 6 THE MX300-I OXYGEN MONITOR, R17MED OXYGEN SEN-SOR AND ASSOCIATED HARDWARE ARE INTENDED FOR US

Page 7 - List of Figures

Portable Oxygen Monitor Operating Instructions Teledyne Analytical Instruments 7 Operating Instructions Notes: Upon receipt, INSPECT THE ENTIRE

Page 8 - List of Tables

Operating Instructions MX300-I Teledyne Analytical Instruments 8 Warning: The MX300-I oxygen Monitor, R17MED oxygen sensor, and associate

Page 9 - Introduction

Portable Oxygen Monitor Operating Instructions Teledyne Analytical Instruments 9 Note: Do not use the plastic flow diverter with the R17MED senso

Page 10 - () KEY

Operating Instructions MX300-I Teledyne Analytical Instruments 10 To install V-Mount Adapter, remove the battery compartment door by pry-ing up the

Page 11 - 1.1 Features

Portable Oxygen Monitor Operating Instructions Teledyne Analytical Instruments 11 3. Install 3 “AA” alkaline batteries into the holder as shown i

Page 12 - 1.4 Operation Summary

MX300-I Teledyne Analytical Instruments ii Copyright © 2009 Teledyne Analytical Instruments. All Rights Reserved No part of this manual may be repr

Page 13 - CAUTION STATEMENTS:

Operating Instructions MX300-I Teledyne Analytical Instruments 12 play flashing 00.0. Only the CAL and ON/OFF (I/O) keys are functional at this poi

Page 14

Portable Oxygen Monitor Operating Instructions Teledyne Analytical Instruments 13 Note: At the end of the calibration countdown, the Monitor will

Page 15 - Operating Instructions

Operating Instructions MX300-I Teledyne Analytical Instruments 14 1. To set the HI alarm: Press the HI ALARM SET key once. Press the UP and DOWN

Page 16

Portable Oxygen Monitor Operating Instructions Teledyne Analytical Instruments 15 2.1.6 Output 0-1 VDC or RS232 The MX300-I Portable Oxygen Monito

Page 17

Operating Instructions MX300-I Teledyne Analytical Instruments 16 RS232 signal output has been tested with Windows Hyperterminal. The setup changes

Page 18

Portable Oxygen Monitor Operating Instructions Teledyne Analytical Instruments 17 2.2 Operation 2.2.1 Procedure Note: Prior to every use, always t

Page 19

Operating Instructions MX300-I Teledyne Analytical Instruments 18 2.3 Factors Affecting Sampling Accuracy 2.3.1 Humidity Humidity does not directly

Page 20 - Monitor

Portable Oxygen Monitor Operating Instructions Teledyne Analytical Instruments 19 When a ventilator is a part of the breathing system, the alterna

Page 21

Operating Instructions MX300-I Teledyne Analytical Instruments 20 Error Helium 50%, balance O2 0% Nitrous Oxide 80%, balance O2 0% Carbon Dioxi

Page 22

Portable Oxygen Monitor Operating Instructions Teledyne Analytical Instruments 21 2.3.6 Cleaning The MX300-I and R17MED sensor are non-sterile dev

Page 23

Portable Oxygen Monitor Teledyne Analytical Instruments iii About This Manual This Manual describes the installation, configuration, operati

Page 24

Operating Instructions MX300-I Teledyne Analytical Instruments 22  Perform alarm test before each use (press the ALARM TEST key).  Use only iso

Page 25 - 2.2 Operation

Portable Oxygen Monitor Service Instructions Teledyne Analytical Instruments 23 Service Instructions 3.1 General Service Information Without acces

Page 26

Service Instructions MX300-I Teledyne Analytical Instruments 24 3.4 Sensor Maintenance DO check the sensor for damage or leaks before use. DON’T im

Page 27 - Gas or Vapor Level

Portable Oxygen Monitor Service Instructions Teledyne Analytical Instruments 25 Table 3-1 Troubleshooting Symptom Why What To Do New sensor resp

Page 28

Service Instructions MX300-I Teledyne Analytical Instruments 26 Symptom Why What To Do C) The sensor output has fal-len to a level where it is n

Page 29 - 2.4 Dos and Don’ts

Portable Oxygen Monitor Service Instructions Teledyne Analytical Instruments 27 Symptom Why What To Do C) Calibrate. Note: In the event that n

Page 30

Service Instructions MX300-I Teledyne Analytical Instruments 28 50 5 The ADC circuit failed during POST. 6 (5) 6 (There is no indication of whic

Page 31 - Service Instructions

Portable Oxygen Monitor Appendix Teledyne Analytical Instruments 29 3.11 Return Authorization for Service For any service beyond sensor and batte

Page 32 - 3.8 Troubleshooting

Appendix MX300-I Teledyne Analytical Instruments 30 Appendix A.1 Specifications Range: 0-100% oxygen Accuracy: +2% of full scale (at constant te

Page 33 - Table 3-1 Troubleshooting

Portable Oxygen Monitor Appendix Teledyne Analytical Instruments 31 Left intentionally blank

Page 34 - Symptom Why What To Do

MX300-I Teledyne Analytical Instruments iv Safety Messages Your safety and the safety of others are very important. Please care-fully read and unde

Page 35 - 3.9 Watchdog

Appendix MX300-I Teledyne Analytical Instruments 32 A.2 Spare Parts List QTY PART NO DESCRIPTION 1 C43690-R17MED Micro-Fuel Cell R17MED with flow

Page 36

Portable Oxygen Monitor Appendix Teledyne Analytical Instruments 33 Orders should be sent to: TELEDYNE Analytical Instruments 16830 Chestnut Stre

Page 37

Portable Oxygen Monitor Teledyne Analytical Instruments v Table of Contents Safety Messages ...

Page 38 - Appendix

MX300-I Teledyne Analytical Instruments vi 2.3.3 Pressure ... 182.3.4

Page 39

Portable Oxygen Monitor Teledyne Analytical Instruments vii List of Figures Figure 1-1: MX300-I Front View ...

Page 40 - A.3 Optional Accessories

MX300-I Teledyne Analytical Instruments viii List of Tables Table 2-1: Oxygen Reading Error in a Mixture of Anesthetic Gas . 19Table 3-1 Troublesh

Page 41

Portable Oxygen Monitor Introduction Teledyne Analytical Instruments 1 Introduction The Teledyne Analytical Instruments MX300-I Oxygen Monitor, h