Series 4030 Teledyne Analytical Instruments INSTRUCTION, OPERATING AND MAINTENANCE MANUAL FOR SERIES 4030 H
Introduction Teledyne Analytical Instruments 8 2. High resolution, accurate readings of concentration. 3.
Introduction Teledyne Analytical Instruments 9 on the customer’s requirements. The appropriate power cord fo
Introduction Teledyne Analytical Instruments 10 The Series 4030 Analyzer is composed of three subsystems:
Introduction Teledyne Analytical Instruments 11 2.2.2 Gas Flow Control System The Piping diagram is in the
Introduction Teledyne Analytical Instruments 12 The anode-igniter, as its name implies, serves two functions
Introduction Teledyne Analytical Instruments 13 Figure 2-1: Flame Ionization Cell The electrode section of t
Introduction Teledyne Analytical Instruments 14 nominally 125 volts. Output tolerance is ±10 volts from the
Installation Teledyne Analytical Instruments 15 Installation Installation of the Series 4030 Analyzer includ
Installation Teledyne Analytical Instruments 16 Regardless of configuration, the analyzer/system must be ins
Installation Teledyne Analytical Instruments 17 Figure 3-1: Gas Connections Gas connections to the ins
Series 4030 Teledyne Analytical Instruments Copyright © 2004 Teledyne Analytical Instruments All Rights Res
Installation Teledyne Analytical Instruments 18 Before tubing is connected to the system, it must be deconta
Installation Teledyne Analytical Instruments 19 THE ANALYZER WITH ITS FASTENING SCREWS. ULTIMATELY, IT IS TH
Installation Teledyne Analytical Instruments 20 Both the first and third set of outputs as well as the seco
Installation Teledyne Analytical Instruments 21 3.3.3.5 ALARM RELAYS The nine alarm-circuit connector pins
Installation Teledyne Analytical Instruments 22 Table 3-2: Alarm Relay Contact Pins Pin Contact 45 Thres
Installation Teledyne Analytical Instruments 23 3.3.3.7 NETWORK I/O A serial digital input/output for local
Installation Teledyne Analytical Instruments 24 30 31 32 Exhaust Solenoid Hot 33 Sample Solenoid Hot 34
Installation Teledyne Analytical Instruments 25 Implementation: The RS-232 protocol allows some flexibilit
Installation Teledyne Analytical Instruments 26 CAUTION: THE GASES USED MUST BE OF THE HIGHEST QUALITY, ULT
Operation Teledyne Analytical Instruments 27 Operation 4.1 Placing the System in Operation 1. Plug the power
Series 4030 Teledyne Analytical Instruments i Safety Messages Your safety and the safety of others are very
Operation Teledyne Analytical Instruments 28 4.2.4 Fuel 1. Open the main valve on the tank and set the fu
Operation Teledyne Analytical Instruments 29 the temperature controller is maintaining the temperature set poi
Operation Teledyne Analytical Instruments 30 Depending on the specifics of the application, this might include
Operation Teledyne Analytical Instruments 31 4.4.1 Default Parameters The versatility of this analyzer usual
Operation Teledyne Analytical Instruments 32 particular key it will be typed as text using all caps. Example:
Operation Teledyne Analytical Instruments 33 - Dynamic (real time) FID input level - Time spent in mode - Rang
Operation Teledyne Analytical Instruments 34 saving the value. The value will revert to the original before m
Operation Teledyne Analytical Instruments 35 11. MODIFIABLE ITEM screens have a flashing value somewhere in th
Operation Teledyne Analytical Instruments 36 ENTER it. The analyzer will immediately start that function, unle
Operation Teledyne Analytical Instruments 37 Password Mismatch ----------------------- Password Chang
Series 4030 Teledyne Analytical Instruments ii CAUTION: THE ANALYZER SHOULD ONLY BE USED FOR THE PURPOSE AN
Operation Teledyne Analytical Instruments 38 SPAN ----------------------- THC Span ?: Yes SMP Span ?
Operation Teledyne Analytical Instruments 39 ----------------------- ANALOG-OUT ADJUST ------------------
Operation Teledyne Analytical Instruments 40 ----------------------- STANDBY Switch to Sleep Mode E
Operation Teledyne Analytical Instruments 41 To setup the GAIN: 1) From the Analysis Screen, press an arrow k
Operation Teledyne Analytical Instruments 42 Note however, that the instrument can still be used for analysis
Operation Teledyne Analytical Instruments 43 <Enter> The last <Enter> enters the password. <Es
Operation Teledyne Analytical Instruments 44 ________________________________________________ Reenter a new Pa
Operation Teledyne Analytical Instruments 45 Pressing <Enter> will logout the current user and protect
Operation Teledyne Analytical Instruments 46 ________________________________________________ <Enter or Es
Operation Teledyne Analytical Instruments 47 The first line determines whether a span calculation will be perf
Series 4030 Teledyne Analytical Instruments iii Additional Safety Information DANGER COMBUSTIBLE GAS USAG
Operation Teledyne Analytical Instruments 48 AL-1 20.00 ppb AL-1 SOURCE THC AL-2 ACTIVE AL-2 LOW
Operation Teledyne Analytical Instruments 49 only during the time the concentration is above the trip point.
Operation Teledyne Analytical Instruments 50 The value of the upper limit of the currently selected range (AUT
Operation Teledyne Analytical Instruments 51 4.4.14 Analog output Adjustment The software in this instrument p
Operation Teledyne Analytical Instruments 52 'Gain: ' plus the Offset value plus a value that is pro
Appendix Teledyne Analytical Instruments 53 Recommended Spare Parts List Qty. P/N Description 1 B74671
Appendix Teledyne Analytical Instruments 54 Send orders to: TELEDYNE INSTRUMENTS Analytical Instruments 16830
Series 4030 Teledyne Analytical Instruments iv system. See the procedures for purging and leak checking this
Series 4030 Teledyne Analytical Instruments v Procedure for Removal of Internal Inaccessible Shock Hazards
Series 4030 Teledyne Analytical Instruments vi DEPENDING ON THE APPLICATION, MAY ALSO USE TOXIC GASES. IT IS
Introduction Teledyne Analytical Instruments 7 Introduction Teledyne Analytical Instruments Series 4030 Tota
Comments to this Manuals