Model GFC7000E Instruction Manual Operating Instructions
04584 Rev A1 71
6.9.4.1. Analog Output Signal Type and Range Span Selection
To select an output signal type (DC Voltage or current) and level for one output channel, activate
the ANALOG I/O CONFIGURATION MENU (see Section 6.9.1) then press:
DIAG AIO CONC_OUT_2 RANGE: 5V
SET> EDIT EXIT
DIAG ANALOG I / O CONFIGURATION
PREV NEXT ENTR EXIT
DIAG AIO AOUTS CALIBRATED: NO
< SET SET> CAL EXIT
FROM ANALOG I/O CONFIGURATION MENU
DIAG AIO OUTPUT RANGE: 5V
0.1V 1V 5V 10V CURR ENTR EXIT
These keys set
the signal level
and type of the
selected channel
DIAG AIO CONC_OUT_2:5V, CAL
< SET SET> EDIT EXIT
Press SET> to select the
analog output channel to be
configured. Press EDIT to
continue
Pressing ENT
R
records the new setting
and returns to the previous menu.
Pressing
EXIT
ignores the new setting and
returns to the previous menu.
DIAG AIO OUTPUT RANGE: 10V
0.1V 1V 5V 10V CURR ENTR EXIT
6.9.4.2. Analog Output Calibration Mode
The analog outputs can be calibrated automatically or manually. In its default mode, the
instrument is configured for automatic calibration of all channels. Manual calibration should be
used for the 0.1V range or in cases where the outputs must be closely matched to the
characteristics of the recording device. Outputs configured for automatic calibration can be
calibrated as a group or individually. Calibration of the analog outputs needs to be carried out on
first startup of the analyzer (performed in the factory as part of the configuration process) or
whenever re-calibration is required.