Portable Oxygen Analyzer Teledyne Analytical Instruments i OPERATING & SERVICE INSTRUCTIONS FOR AX300-I PORTABLE OXYGEN ANALYZER 00
Introduction AX300-I Teledyne Analytical Instruments 2 This alarm condition is also triggered when the oxygen concentration at the point of analysis
Portable Oxygen Analyzer Introduction Teledyne Analytical Instruments 3 1.2 Options The following options are available for the AX 300-I analyzer
Introduction AX300-I Teledyne Analytical Instruments 4 The diverter is not necessary, and should not be used, when the sensor is placed directly in
Portable Oxygen Analyzer Operating Instructions Teledyne Analytical Instruments 5 Operating Instructions Notes: Upon receipt, INSPECT THE ENTIR
Operating Instructions AX300-I Teledyne Analytical Instruments 6 Figure 2-1: Installing the R17MED Sensor 3. Plug the other end of the cable
Portable Oxygen Analyzer Operating Instructions Teledyne Analytical Instruments 7 Figure 2-3: Mounting the Sensor in the Tee Adapter 2.1.2 Mounti
Operating Instructions AX300-I Teledyne Analytical Instruments 8 2.1.2.2 Universal Mounting Clamp Installation The optional Universal Mounting Clam
Portable Oxygen Analyzer Operating Instructions Teledyne Analytical Instruments 9 Figure 2-6: Installing Batteries 4. Re-install the battery co
Operating Instructions AX300-I Teledyne Analytical Instruments 10 Figure 2-7: Calibration Sequence To calibrate the analyzer
Portable Oxygen Analyzer Operating Instructions Teledyne Analytical Instruments 11 key and press the CAL key to repeat calibration. Note: The AX3
AX300-I Teledyne Analytical Instruments ii Copyright © 2009 Teledyne Analytical Instruments. All Rights Reserved No part of this manual may be repr
Operating Instructions AX300-I Teledyne Analytical Instruments 12 optional RS 232 Interface Cable (P/N B-75555) available from Teledyne for connect
Portable Oxygen Analyzer Operating Instructions Teledyne Analytical Instruments 13 If your analyzer is set for analog (0-1 VDC) output, you can
Operating Instructions AX300-I Teledyne Analytical Instruments 14 In confined volumes such as in incubators or tents. When analyzing for oxygen i
Portable Oxygen Analyzer Operating Instructions Teledyne Analytical Instruments 15 2.3.2 Temperature The R17MED oxygen sensor compensates for amb
Operating Instructions AX300-I Teledyne Analytical Instruments 16 troubleshooting section of this manual may provide additional assistance in lo-ca
Portable Oxygen Analyzer Operating Instructions Teledyne Analytical Instruments 17 CAUTION: THE AX300-I SHOULD NOT BE USED IN THE PRESENCE OF FL
Operating Instructions AX300-I Teledyne Analytical Instruments 18 Read and understand the manual before using the analyzer Calibrate every 8 ho
Portable Oxygen Analyzer Service Instructions Teledyne Analytical Instruments 19 Service Instructions 3.1 General Service Information Without acc
Service Instructions AX300-I Teledyne Analytical Instruments 20 DON’T open or try to repair the sensor. Before every use, the sensor, cable and con
Portable Oxygen Analyzer Service Instructions Teledyne Analytical Instruments 21 Symptom Why What To Do Sensor will not read be-low 22 % after
Portable Oxygen Analyzer Teledyne Analytical Instruments iii About This Manual This Manual describes the installation, configuration, operatio
Service Instructions AX300-I Teledyne Analytical Instruments 22 Symptom Why What To Do The oxygen reading fluctuates or appears to be incorrect.
Portable Oxygen Analyzer Service Instructions Teledyne Analytical Instruments 23 3.8 Watchdog The AX300-I is equipped with a watchdog circuit th
Service Instructions AX300-I Teledyne Analytical Instruments 24 7(2) 7 (There is no indication of the type of DAC failure) A failure has occurred d
Portable Oxygen Analyzer Appendix Teledyne Analytical Instruments 25 Appendix A.1 Specifications Range: 0-100% oxygen Accuracy: +2% of full s
Appendix AX300-I Teledyne Analytical Instruments 26 A.2 Spare Parts List QTY PART NO DESCRIPTION 1 C43690-R17MED Micro-Fuel Cell R17MED with flow
Portable Oxygen Analyzer Appendix Teledyne Analytical Instruments 27 Orders should be sent to: TELEDYNE Analytical Instruments 16830 Chestnut Str
AX300-I Teledyne Analytical Instruments iv Safety Messages Your safety and the safety of others are very important. Please care-fully read and unde
Portable Oxygen Analyzer Teledyne Analytical Instruments v Table of Contents Safety Messages ...
AX300-I Teledyne Analytical Instruments vi 3.1 General Service Information 19 3.2 Overall Maintenance 19 3.3 Battery Maintenance 19 3.4 Sensor M
Portable Oxygen Analyzer Teledyne Analytical Instruments vii List of Figures Figure 1-1: AX300-I Front View ...
AX300-I Teledyne Analytical Instruments viii List of Tables Table 2-1: Oxygen Reading Error in a Mixture of Anesthetic Gas . 16 Table 3-1 Troubles
Portable Oxygen Analyzer Introduction Teledyne Analytical Instruments 1 Introduction The Teledyne Analytical Instruments AX300-I Oxygen Analyzer,