149-072011 Model 2002 Vacuum Gauge Page 16 of 38
4.3. TTL Outputs
TTL outputs are provided for process control. High and Low set points are entered on the front panel (see
Section 3.4) and TTL signals are generated on the I/O cable in the following manner:
1. When the pressure as indicated by the Model 2002 is above the High set point, the High signal line (pin #3) will
be set to a TTL high level (+5 VDC). When the indicated pressure is below the High set point, the High signal
line (pin#3) will be set to a TTL low level (+0 VDC).
2. When the pressure as indicated by the Model 2002 is below the Low set point, the Low signal line (pin #4) will
be set to a TTL high level (+5 VDC). When the indicated pressure is above the Low set point, the Low signal
line (pin #4) will be set to a TTL low level (+0 VDC).
These TTL signals are used for a variety of applications. With these signals, the user has the capability to
turn off/on various equipment such as valves, flow controllers, pumps, heaters, and safety equipment.
4.4. Power Entry Module
The AC power for the Model 2002 enters the instrument through a power entry module that contains a fuse,
on/off switch and an IEC 320 power inlet.
The fuse is rated for 250 V, 1/4 A. It can be accessed by unplugging the AC cord and prying the fuse
compartment open with the tab and slot in the power inlet chamber. There is a spare fuse in the
compartment within the fuse holder.
Model 2002
Rear Panel Detail
Fig 4.2
TRANSDUCER CONN.
(FEMALE)
STANDARD
I/O CONNECTOR
LOW SETPOINT OPTION BOARD HIGH SETPOINT
RELAY CONNECTOR (MALE) RELAY