Teledyne Analytical Instruments QUICKSTART GUIDE MODEL 3000 TA Trace Oxygen Analyzer
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 10 9. With the O-ring in place, align the guide pin with the hole on th
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 11 BASIC OPERATION Once the analyzer has been installed, it can be configured for y
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 12 Available Functions The six function buttons are
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 13 Password Protection If a password is assigned, then setting
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 14 TRAK/HLD Auto—Cal PSWD Logout More • Use the ◄► arrow keys to scroll
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 15 Auto Mode Zero Calibration • Press Zero to enter the zero function mode. • Us
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 16 Span Calibration Make sure an appropriate span gas is connected to th
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 17 Calib. Holding time Cal hold: 5 min • Press Enter key to continue. Span Val:
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 18 • Latching? Ltch–Y/N (Yes/No). • To change any of the alarm paramet
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 19 • Use the ▲▼ arrow keys to switch from AUTO to FX/LO, FX/MED, or FX/HI to set t
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 3 GETTING STARTED This Quickstart Guide is designed to get you set up and operating
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 4 • Left & Right Arrows Select between functions currently display
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 5 • Analog Output Connection: Terminals for current and voltage analog output sig
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 6 SETUP AND INSTALL Mount the Analyzer The Model 300
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 7 4. If using remote calibration feature: • Connect zero gas to the ZERO IN port.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 8 • Connect primary input power: Insert the power cord into the power c
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 9 • Connect Range ID Contacts: There are four dedicated Range ID relay contacts on