62.0 Installation & Operation 319RTELEDYNE BROWN ENGINEERINGAnalytical InstrumentsSetting AlarmsThe alarms can be adjusted over the full scale ra
7319R Installation & Operation 2.0TELEDYNE BROWN ENGINEERINGAnalytical InstrumentsCalibrationTo calibrate the 319R:1. Place the power switch in t
82.0 Installation & Operation 319RTELEDYNE BROWN ENGINEERINGAnalytical Instruments
9319R Maintenance 3.0TELEDYNE BROWN ENGINEERINGAnalytical InstrumentsMaintenanceReplacing the FuseIf the unit will not turn on when plugged in and sw
3.0 Maintenance 319R10TELEDYNE BROWN ENGINEERINGAnalytical Instruments5. Replace the fuse holder into its receptacle, pushing in firmlyuntil it click
11319R Maintenance 3.0TELEDYNE BROWN ENGINEERINGAnalytical InstrumentsWarranty ConditionsThe Class Z-2 cell used in the 319R is warranted for six (6)
3.0 Maintenance 319R12TELEDYNE BROWN ENGINEERINGAnalytical Instruments
13319R AppendixTELEDYNE BROWN ENGINEERINGAnalytical InstrumentsAppendixSpecificationsRanges: 0–200 PPM, 0–1999 PPM O2, and 0–25% calibration rangeSens
14Appendix 319RTELEDYNE BROWN ENGINEERINGAnalytical InstrumentsSpare Parts ListQTY. P/N DESCRIPTION1 C57304 PC board, main1 A33748 Thermistor assembly
15319R AppendixTELEDYNE BROWN ENGINEERINGAnalytical InstrumentsDrawing ListC-57490 Outline diagramA-58154 Probe interconnection diagramD-57302 Electro
TELEDYNE BROWN EGINEERINGAnalytical InstrumentsCopyright © 1993 Teledyne Brown Engineering Analytical InstrumentsAll Rights Reserved.No part of this m
16Appendix 319RTELEDYNE BROWN ENGINEERINGAnalytical InstrumentsMaterial Safety Data SheetProduct Name: Micro-Fuel Cells and Super Cells, all classes e
17319R AppendixTELEDYNE BROWN ENGINEERINGAnalytical InstrumentsFlash Point: na Flammable Limits: na LEL: na UEL: naExtinguishing Media: Use extinguish
18Appendix 319RTELEDYNE BROWN ENGINEERINGAnalytical InstrumentsContact of electrolyte with skin or eyes will cause aburning sensation and/or feel soap
19319R AppendixTELEDYNE BROWN ENGINEERINGAnalytical InstrumentsSection VIII – Control MeasuresEye Protection: Chemical splash gogglesHand Protection:
TELEDYNE BROWN EGINEERINGAnalytical InstrumentsTable of Contents1.0 IntroductionMain Features ...
TELEDYNE BROWN EGINEERINGAnalytical Instruments
1319R Introduction 1.0TELEDYNE BROWN ENGINEERINGAnalytical InstrumentsIntroductionThe Teledyne Brown Engineering Analytical Instruments (TBE)Model 319
1.0 Introduction 319R2TELEDYNE BROWN ENGINEERINGAnalytical Instruments4. Proven Sensor. The 319R uses the Class Z-2 Micro-Fuel Cell,warranted for 6 mo
3319R Installation & Operation 2.0TELEDYNE BROWN ENGINEERINGAnalytical InstrumentsInstallation & OperationPrecautions1. Read this chapter in
42.0 Installation & Operation 319RTELEDYNE BROWN ENGINEERINGAnalytical InstrumentsThe analysis unit should be installed in the process at a maxim
5319R Installation & Operation 2.0TELEDYNE BROWN ENGINEERINGAnalytical InstrumentsAlarm Terminal ConnectionSPDT relay contacts are provided for e