13
319R Appendix
TELEDYNE BROWN ENGINEERING
Analytical Instruments
Appendix
Specifications
Ranges: 0–200 PPM, 0–1999 PPM O
2
, and 0–
25% calibration range
Sensitivity: 1% of full scale
Signal Output: Standard: 0–10 VDC, negative ground
4-20 mA, negative ground
plus contact closures for range
identification.
Display: Digital 7 segment, 3½-digit LCD with
over range indicator.
Alarms: One high alarm, adjustable & fail-safe.
One low alarm, adjustable & fail-safe.
One low alarm, fixed at 2% full scale and
fail-safe.
Accuracy: ±2% of full scale at constant temperature
and pressure of calibration
±5% of full scale throughout operating
temperature range (once temperature
equilibrium has been reached).
Response: 90% of full scale in 35 seconds.
Operating Temp. Range: 32 °F to 122 °F (0°C to 50°C)
Power: 85-264 VAC, 47-440 Hz or 10-36 VDC,
user specified, 10 W max.
Enclosure: General purpose, panel mount
Cell Type: Z-2
Cell Warranty: 6 months
Cell Life (Expected): 6 months
Dimensions: 2.75 "H × 6.0 "W × 2.75 "D
(63.5 mm × 152.4 mm × 63.5 mm)
Faceplate: 3.75 "H × 7.0 "W
(95.25 mm × 177.8 mm)
Probe unit: 4.0 "H × 6.0 "W × 3 "D
(101.6 mm × 152.4 mm × 76.2 mm)