iTeledyne Analytical InstrumentsOPERATING INSTRUCTIONS FORModel 3300PBPercent Oxygen AnalyzerHIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRE
1-41 Introduction Model 3300PBTeledyne Analytical Instruments• Power Connection AC version: 100–240 V ac, at 50/60Hz.The connector housing includ
2-1Percent Oxygen Analyzer Operational Theory 2Teledyne Analytical InstrumentsOperational Theory2.1 IntroductionThe analyzer is composed of two subsy
2-22 Operational Theory Model 3300PBTeledyne Analytical InstrumentsMicro-Fuel Cell is therefore a hybrid between a battery and a true fuel
2-3Percent Oxygen Analyzer Operational Theory 2Teledyne Analytical Instrumentsconductor connecting them, through some electrical circuitry, to one of
2-42 Operational Theory Model 3300PBTeledyne Analytical Instrumentsexternal current, will increase, even though the proportion of oxygen h
2-5Percent Oxygen Analyzer Operational Theory 2Teledyne Analytical Instruments2.3 Electronics2.3.1 GeneralThe signal processing uses an Intel microc
2-62 Operational Theory Model 3300PBTeledyne Analytical InstrumentsIn the presence of oxygen the cell generates a current. The sensor has
3-1Percent Oxygen Analyzer Installation 3Teledyne Analytical InstrumentsInstallation of the analyzer includes:1. Unpacking the system.2. Mounting the
3 Installation Model 3300PB3-2Teledyne Analytical Instruments3.2 Location and Mounting3.2.1 Control Unit InstallationThe 3300PB Control Unit is de
3-3Percent Oxygen Analyzer Installation 3Teledyne Analytical InstrumentsPrimary Input Power: The power strip supplied inside the analyzer.Connect gro
iiTeledyne Analytical InstrumentsCopyright © 1999 Teledyne Analytical InstrumentsAll Rights Reserved. No part of this manual may be reproduced, trans
3 Installation Model 3300PB3-4Teledyne Analytical InstrumentsThe alarm relay circuits are designed for failsafe operation, meaning therelays are e
3-5Percent Oxygen Analyzer Installation 3Teledyne Analytical Instruments3.3.1 RS-232 Port OptionThe digital signal output is a standard RS-232 seri
3 Installation Model 3300PB3-6Teledyne Analytical InstrumentsA switching valve is provided to feed the analysis unit with eithersample or gas.A fl
3-7Percent Oxygen Analyzer Installation 3Teledyne Analytical Instruments3.4.1 Vacuum Service OptionFor vacuum service pluming reference figure 3-4.
3 Installation Model 3300PB3-8Teledyne Analytical Instruments3.5 Installation ChecklistBefore connecting the instrument to the power source and tu
Percent Oxygen Analyzer Operation 44-1Teledyne Analytical InstrumentsOperation4.1 IntroductionOnce the analyzer has been mounted, the gas lines conne
4 Operation Model 3300PB4-2Teledyne Analytical Instruments4.2 Using the Function and Data EntryButtonsWhen no buttons on the Analyzer are being pre
Percent Oxygen Analyzer Operation 44-3Teledyne Analytical Instruments4.3.1 HI RangeSetting the HI Range fullscale value defines the LEAST sensitivean
4 Operation Model 3300PB4-4Teledyne Analytical Instrumentsuntil the display reads the desired percent concentration. (ifwithin 5 seconds no key is
Percent Oxygen Analyzer Operation 44-5Teledyne Analytical Instruments4.5 Selecting a Fixed Range or AutorangingThe Model 3300PB can operate in fixed
iiiTeledyne Analytical InstrumentsContentsIntroduction1.1 Overview... 1-11.2 Main
4 Operation Model 3300PB4-6Teledyne Analytical InstrumentsNote: When you stop pressing the UP/DOWN keys, the rapid flashing willcease and the new s
5-1Percent Oxygen Analyzer Maintenance 5Teledyne Analytical InstrumentsMaintenanceAside from normal cleaning and checking for leaks at the gas conne
5-25 Maintenance Model 3300PBTeledyne Analytical Instrumentsfuse holder will slide out. There are two fuses in use and are visible inthe cl
5-3Percent Oxygen Analyzer Maintenance 5Teledyne Analytical InstrumentsWARNING: The sensor used in the Model 3300PB uses electrolyteswhich contain s
5-45 Maintenance Model 3300PBTeledyne Analytical Instruments5.2.5 Cell Warranty ConditionsThe Class I-22 Micro-Fuel cell is used in the Mo
A-1Percent Oxygen Analyzer AppendixTeledyne Analytical InstrumentsAppendixA.1 SpecificationsRanges: 0-3 % and 0-10 % oxygen (StandardRanges), and 0-25
A-2Appendix Model 3300PBTeledyne Analytical InstrumentsA.2 Spare Parts ListQTY. P/N DESCRIPTION1 C-65220E PC Board, Main1 C-65220F PC Board,
A-3Percent Oxygen Analyzer AppendixTeledyne Analytical InstrumentsA.3 Reference DrawingA-69265 Piping DiagramD-69229 Outline DiagramA.4 MiscellaneousT
A-4Appendix Model 3300PBTeledyne Analytical Instruments
A-5Percent Oxygen Analyzer AppendixTeledyne Analytical InstrumentsA.5 Material Safety Data SheetSection II – Physical and Chemical DataChemical and Co
ivTeledyne Analytical Instruments4.4 Setting the Alarm Setpoints... 4-34.4.1 Alarm 1 ...
A-6Appendix Model 3300PBTeledyne Analytical InstrumentsSection III – Physical HazardsSection IV – Health Hazard DataPotential for fire and e
A-7Percent Oxygen Analyzer AppendixTeledyne Analytical InstrumentsSection V – Emergency and First Aid ProceduresNOTE: The oxygen sensors are sealed, a
vTeledyne Analytical InstrumentsCOMBUSTIBLE GAS USAGE WARNINGDANGERThis is a general purpose instrument designed for usage in anonhazardous area. It
viTeledyne Analytical InstrumentsModel 3300PB complies with all of the requirements of theCommonwealth of Europe (CE) for Radio Frequency Interferenc
1-1Percent Oxygen Analyzer Introduction 1Teledyne Analytical InstrumentsIntroduction1.1 OverviewThe Teledyne Electronic Technologies Analytical Inst
1-21 Introduction Model 3300PBTeledyne Analytical Instruments• Operator can select Autoranging, which allows the analyzer toautomatically select
1-3Percent Oxygen Analyzer Introduction 1Teledyne Analytical Instruments• Set Alarm 2 Set the Alarm 2 Hi or Low, and theconcentration to which alarm