Portable Oxygen Monitor Teledyne Analytical Instruments i OPERATING & SERVICE INSTRUCTIONS FOR AD300 PORTABLE OXYGEN ANALYZER MD300 PORT
Introduction AD300/MD300 Teledyne Analytical Instruments 10 Figure 1-2: AD300 Front View The LCD display consists of
Oxygen Analyzer Introduction Teledyne Analytical Instruments 11 1.1 Features The AD300/MD300 are rugged, versatile instruments capable of rapidl
Introduction AD300/MD300 Teledyne Analytical Instruments 12 Note: Contact the factory for retrofitting an existing instrument for 3 digit LCD d
Portable Oxygen Monitor Operation Teledyne Analytical Instruments 13 Operation Note: Upon receipt, INSPECT THE ENTIRE UNIT FOR DAMAGE. Check
Operation AD300/MD300 Teledyne Analytical Instruments 14 WARNING: THE SENSOR ELECTROLYTE IS CAUSTIC. DO NOT LET IT COME IN CONTACT WITH SKIN.
Portable Oxygen Monitor Operation Teledyne Analytical Instruments 15 Note: When the AD/MD300 instrument is used for diffusion sampling the p
Operation AD300/MD300 Teledyne Analytical Instruments 16 Figure 2-4: V-Mount Adapter Installation 2.1.2.2 UNIVERSAL MOUNTING CLAMP INSTALLATION
Portable Oxygen Monitor Operation Teledyne Analytical Instruments 17 CAUTION: IMPROPER INSTALLATION OF THE BATTERIES MAY RESULT IN DAMAGE TO
Operation AD300/MD300 Teledyne Analytical Instruments 18 Figure 2-7: Calibration Sequence readout display, should show a bar graph of the remai
Portable Oxygen Monitor Operation Teledyne Analytical Instruments 19 connect to a supply of pure dry oxygen flowing at 1-2 liters per minute.
AD300/MD300 Teledyne Analytical Instruments ii Copyright © 2010 Teledyne Analytical Instruments All Rights Reserved No part of this manual may
Operation AD300/MD300 Teledyne Analytical Instruments 20 Figure 2-8: Setting the Alarms (MD300) 8. (MD300 only) Adjust alarm set points to des
Portable Oxygen Monitor Operation Teledyne Analytical Instruments 21 5. To disable the alarms set the high alarm set point above 100%. Press
Operation AD300/MD300 Teledyne Analytical Instruments 22 Figure 2-9: 0-1 VDC or RS 232 Digital Output Port If your instrument is set for analo
Portable Oxygen Monitor Operation Teledyne Analytical Instruments 23 From the low pressure inflator hose using the sample adapter as shown
Operation AD300/MD300 Teledyne Analytical Instruments 24 Figure 2-11: High Pressure Gas Sampling When monitoring for oxygen in confined volu
Portable Oxygen Monitor Operation Teledyne Analytical Instruments 25 for more than a few minutes can also affect the temperature tracking whi
Operation AD300/MD300 Teledyne Analytical Instruments 26 2.3.4 Discrepancy in Readings When a discrepancy in oxygen readings is suspected, the
Portable Oxygen Monitor Operation Teledyne Analytical Instruments 27 2.4 Do’s and Don’ts – DO – Read all of the directions before using for t
Operation AD300/MD300 Teledyne Analytical Instruments 28 Freeze the sensor or instrument. Open or try to repair a leaking or broken sensor. Imme
Portable Oxygen Monitor Service Manual Teledyne Analytical Instruments 29 Service Manual 3.1 General Service Information The Teledyne Model A
Portable Oxygen Monitor Teledyne Analytical Instruments iii About This Manual The AD300/MD300 Operator's Manual provides both introduct
Service Manual AD300/MD300 Teledyne Analytical Instruments 30 DO: Recalibrate after replacing batteries. The MD300 instrument incorporates a bat
Portable Oxygen Monitor Service Manual Teledyne Analytical Instruments 31 unit, the discrepancy in oxygen readings persists, the problem is m
Service Manual AD300/MD300 Teledyne Analytical Instruments 32 Symptom Why What To Do higher than 21%. Take the instrument to a well-ventilated a
Portable Oxygen Monitor Service Manual Teledyne Analytical Instruments 33 Symptom Why What To Do The unit has stopped working and the LCD is
Service Manual AD300/MD300 Teledyne Analytical Instruments 34 Symptom Why What To Do Cannot adjust calibration or alarm settings Critical settin
Portable Oxygen Monitor Service Manual Teledyne Analytical Instruments 35 Table 3-2 Error Codes Error Code Audio Beeps Error Indeterminate 2
Service Manual AD300/MD300 Teledyne Analytical Instruments 36 3.8 Other Problems with the Instrument Most other problems arise from either mecha
Portable Oxygen Monitor Appendix Teledyne Analytical Instruments 37 Appendix A.1 Specifications Range: 0-100% Oxygen Accuracy: +2% of
Appendix AD300/MD300 Teledyne Analytical Instruments 38 A.2 Spare Parts List QTY PART NO DESCRIPTION 1 C72144-R17S Micro-Fuel Cell R17S 1
Portable Oxygen Monitor Appendix Teledyne Analytical Instruments 39 Orders should be sent to: TELEDYNE Analytical Instruments 16830 Chestnut
AD300/MD300 Teledyne Analytical Instruments iv Safety Messages Your safety and the safety of others are very important. Please carefully read t
Portable Oxygen Monitor Teledyne Analytical Instruments v Table of Contents Safety Messages ...
AD300/MD300 Teledyne Analytical Instruments vi 2.4 Do’s and Don’ts 27 Service Manual ...
Portable Oxygen Monitor Teledyne Analytical Instruments vii List of Figures Figure 1-1: MD300 Front View ...
AD300/MD300 Teledyne Analytical Instruments viii List of Tables Table 3-1 Troubleshooting .....
Oxygen Analyzer Introduction Teledyne Analytical Instruments 9 Introduction The Teledyne Analytical Instruments AD300 Oxygen Analyzer here after