HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORINGSYSTEM.PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING TH
The analyzer should be inspected and calibrated each time it is used. This includes inspecting the filter/water trap and scrubber, checking the condi
calibrate the instrument with Methane, but make it read as if it were exposed to the gas of interest. For example, if you calibrate with Methane and
.32 For a mixture of 70% Hydrogen and 30% Carbon Monoxide: 0.7(2.0) + 0.3(2.0) = 3.5% .86 .32 CAUTION In most cases, the concentrati
water for at least 15 minutes. Call a physician. See Material Safety Data Sheets (MSDS). 3. The PURAFIL Media Scrubber contains Potassium Perman
shown in Figure 4. Figure 4. Location of Sensors (Top View) CO SENSOR: (Refer to Interconnection Diagram) a. Installation: To install
a. Remove the sensor from its protective bag. Carefully remove the shorting clip. b. Pull up on the oxygen sensor holder (on the right), plac
If the analyzer is subject to high positive pressures, damage may occur to the pumps or other flow systems. A regulator may be necessary upstream to
CAUTION: ADJUST SPAN GAS PRESSURE REGULATOR TO MINIMUM FLOW RATE (2-3 scfh) BEFORE CONNECTING TO ANALYZER. NOTE: To perform manual spanning of the
To store reading press ENTER. The word SPAN will light up on top of the LCD. For span calibration a known higher temperature input would be required
To restore the built in temp. calibration constants press the CAL and LOGO keys simultaneously, then press °F/°C key. Press the CAL and COMB/TEM
INSTRUCTION MANUAL MAX 5 PORTABLE COMBUSTION EFFICIENCY MONITOR SALES ORDER: SERIAL NUMBER: Teledyne Analytical Instru
To STORE current readings (Key 3) 1. Activate Special Functions. 2. Press key number 3; the display will read “5 0 0”. CAUTION: If you press ENT
3. To test, press any key other than the LOGO or ENTER keys; to resume normal operation press the LOGO key. 4. To perform a total memory reset pr
1. Turn power switch on (without O2 sensor installed). If a negative O2 reading is displayed, proceed to step 2. If a negative reading is not dis
case, and lift the hood off the unit. BATTERY RECHARGING The rechargeable Nickel-Cadmium batteries need to be recharged after about 8 hours of
To Replace Batteries: 1. Loosen the screws indicated on the back panel of the MAX 5. 2. Remove the upper and lower screws on the outside of the MA
-0.00/-0.0 The negative sign in a reading of -0.00 or -0.0 in the CO or COMB section after auto zero is not significant. bAd Indicates a bad readi
APPENDIX TO PRINT STORED READINGS (KEY 5) - OPTIONAL 1. On the back of the instrument is a “serial output” receptacle for connection to a serial
MATERIAL SAFETY DATA SHEET Section I - Product Identification Product Name: Micro-Fuel Cells, all classes except A-2, A-3, A-5, B-2F and B-2VF.
Section III - Physical Hazards Potential for Fire and Explosion: The electrolyte in micro-fuel cells is not flammable. There are no fire to explosi
Section V - Emergency and First Aid Procedures Eye Contact: Flush eyes with water for at least 15 minutes and get immediate medical attention. Skin
TABLE OF CONTENTS MAX 5 PORTABLE COMBUSTIBLE EFFICIENCY MONITOR GENERAL DESCRIPTION Introduction 1 Main Features 1 OPERATION OF C
MATERIAL SAFETY DATA SHEET Section I - Product Identification Product Name: Carbon Monoxide Sensor, Class F-1R, F-1RH and F-2 Manufacturer: Tele
Section III - Physical Hazards Potential for Fire and Explosion: The electrolyte in carbon monoxide sensors is not flammable. There are no fire or
Section V - Emergency and First Aid Procedures Eye Contact: Flush eyes with water for at least 15 minutes and get immediate medical attention. Skin
RECOMMENDED SPARE PARTS LIST MAX 5 MONITOR QTY. P/N DESCRIPTION 1 C-6689-B3 MICRO-FUEL CELL - OXYGEN SENSOR 1* C-40371-F1R ELECTROCHEMICA
SEND ORDERS TO: TELEDYNE Analytical Instruments16830 Chestnut StreetCity of Industry, CA 91748Telephone: (626) 9341500Fax: (626) 9612538Web: www.t
REFERENCE DRAWINGS MAX 5 MONITOR INSTALLATION DRAWINGS B-39437 OUTLINE DIAGRAM B-39430 PIPING DIAGRAM (STANDARD) B-43108RF PIPING DIAGRAM (
(INSERT DRAWINGS)
TABLE OF CONTENTS (continued) ROUTINE MAINTENANCE Scrubber 19 Battery Recharging 20 Battery Replacement 20 Sensor Replacemen
MAX5 PORTABLE COMBUSTION EFFICIENCY MONITOR GENERAL DESCRIPTION INTRODUCTION The MAX 5 Portable Combustion Efficiency Analyzer provides several spe
MAIN FEATURES 1. Auto Zero. MAX 5 offers automatic zero calibration for carbon monoxide and combustibles and automatic span calibration for oxyg
Temp Key (°F or °C). This key, when pressed, determines which of the two available temperature scales (toggles between degrees Fahrenheit or degrees
* continuous flow of all six readings F. Air Flowmeter with flow set valve to regulate air flow. G. Sample Flowmeter with flow set valve to regu
methane in nitrogen. In other cases, a hydrogen and carbon monoxide mixture might be used. In the presence of catalyst poisons, the combustibles sen