Teledyne Analytical Instruments INSTRUCTION MANUAL Model 9110TH Nitrogen Oxides Analyzer TELEDYNE ELECTRONIC TECHNOLOGIES Analytical Instrume
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 80 4.8. SETUP RNGE: RANGE UNITS AND DILUTION CONFIGURATION
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 81 4.8.2. DILUTION RATIO The dilution ratio is a software op
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 82 4.9. SETUP PASS: PASSWORD FEATURE The 9110TH/M provides
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 83 Example: If all passwords are enabled, the following menu
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 84 4.10. SETUP CLK: SETTING THE INTERNAL TIME-OF-DAY CLOCK
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 85 In order to compensate for CPU clocks which run fast or sl
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 86 4.11. SETUP MORE COMM: SETTING UP THE ANALYSER’S COMM
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 87 4.11.3. COMMUNICATION MODES, BAUD RATE AND PORT TESTING U
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 88 Press the following buttons to select a communication mod
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 89 4.11.3.2. COM Port Baud Rate To select the baud rate of on
Manual Information Model 9110TH NOx Analyzer Teledyne Analytical Instruments xi ABOUT THIS MANUAL This manual describes operation, specificati
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 90 4.11.3.3. COM Port Testing The serial ports can be tested
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 91 4.11.4. ANALYZER ID Each type of TAI analyzer is configure
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 92 4.11.5. RS-232 COM PORT CABLE CONNECTIONS In its default
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 93 Figure 4-6: CPU COM1 & COM2 Connector Pin-Outs for R
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 94 4.11.6. RS-485 CONFIGURATION OF COM2 Opting to use RS-48
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 95 4.11.7.1. Configuring Ethernet Communication Using DHCP 1.
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 96 Table 4-13: LAN/Internet Configuration Properties PROPER
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 97 4.11.7.2. Configuring Ethernet Communication Manually (Sta
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 98 Internet Configuration Button Functions BUTTON
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 99 4.11.7.3. Changing the Analyzer’s HOSTNAME The HOSTNAME
Model 9110TH NOx Analyzer Manual Information Teledyne Analytical Instruments xii This page intentionally left blank.
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 100 4.11.8. USB PORT SETUP The analyzer can be operated thro
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 101 5. In the instrument’s SETUP>MORE>COMM>COM2 men
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 102 4.11.9. MULTIDROP RS-232 SET UP When the RS-232 Multidro
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 103 Figure 4-8: Jumper and Cables for Multidrop Mode Note
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 104 Analyzer Analyzer Analyzer Last Analyzer Female DB9M
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 105 4.11.10. MODBUS SETUP The following set of instructions a
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 106 Example Read/Write Definition window: Example Connect
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 107 4.12. SETUP MORE VARS: INTERNAL VARIABLES (VARS) The
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 108 To access and navigate the VARS menu, use the following
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 109 4.12.1. SETTING THE GAS MEASUREMENT MODE In its standard
Table of Contents Model 9110TH NOx Analyzer Teledyne Analytical Instruments xiii TABLE OF CONTENTS Information About the Specific Configurati
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 110 4.13. SETUP MORE DIAG: DIAGNOSTICS MENU A series of
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 111 4.13.1. ACCESSING THE DIAGNOSTIC FEATURES To access the D
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 112 To enter the signal I/O test mode, press: SAMPLE
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 113 4.13.3. ANALOG OUTPUT STEP TEST This test can be used to
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 114 4.13.4. ANALOG OUTPUTS AND REPORTING RANGES 4.13.4.1. An
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 115 Additionally A1, A2 andA3 may be equipped with optional
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 116 selects which physical range is in effect based on the a
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 117 4.13.5. ANALOG I/O CONFIGURATION 4.13.5.1. The Analog I/O
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 118 5. Set the reporting range scale for the data type chos
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 119 4.13.5.2. Analog Output Signal Type and Range Selection T
Model 9110TH NOx Analyzer Table of Contents Teledyne Analytical Instruments xiv 4.8.1. Range Units ...
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 120 4.13.5.3. Turning the Analog Output Over-Range Feature O
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 121 To Turn the over-range feature on or off, press: From the
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 122 around the zero point. This can be achieved in the 9110
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 123 4.13.5.5. Assigning a DAS Parameter to an Analog Output C
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 124 The available gas concentration DAS parameters for outpu
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 125 To assign a DAS parameter to a specific analog output cha
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 126 4.13.5.8. Setting the Reporting Range Scale for an Analo
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 127 To set the reporting range for an analog output, press: E
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 128 4.13.5.9. Setting Data Update Rate for an Analog Output
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 129 To change the update rate for an individual analog output
Table of Contents Model 9110TH NOx Analyzer Teledyne Analytical Instruments xv 7.2.1. 9110TH Internal Gas Flow Diagrams ...
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 130 4.13.5.10. Turning an Analog Output On or Off Each outpu
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 131 4.13.6. ANALOG OUTPUT CALIBRATION Analog calibration need
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 132 4.13.6.1. Automatic Analog Output Calibration To calibra
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 133 To initiate an automatic calibration for an individual ou
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 134 4.13.6.2. Manual Calibration of Analog Output Configured
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 135 To manually adjust the signal levels of an analog output
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 136 As before, adjustment of the output is performed unti
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 137 Table 4-25: Current Loop Output Calibration with Resist
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 138 4.13.6.4. AIN Calibration This is the sub-menu calibrate
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 139 offset (volts) engineering units to be represented
Model 9110TH NOx Analyzer Table of Contents Teledyne Analytical Instruments xvi 8.3.8. Ozone Supply Air Filter ...
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 140 4.13.7. OTHER DIAG MENU FUNCTIONS 4.13.7.1. Display Sequ
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 141 The default settings for this feature are: Table 4-27: 9
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 142 To delete an entry in the display value list, press:
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 143 4.13.7.2. Optic Test The optic test function tests the re
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 144 4.13.7.3. Electrical Test The electrical test function c
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 145 4.13.7.4. Ozone Generator Override This feature allows th
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 146 To access this feature press the following menu sequence
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 147 4.13.7.5. Flow Calibration The flow calibration allows th
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 148 4.14. SETUP – ALRM: USING THE OPTIONAL GAS CONCENTRATION
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 149 4.15. Remote Operation 4.15.1. REMOTE OPERATION USING THE
Table of Contents Model 9110TH NOx Analyzer Teledyne Analytical Instruments xvii Figure 3-19: 9110TH – Internal Pneumatics with Ambient Zero-
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 150 Table 4-29: Status Output Pin Assignments CONNECTOR PIN
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 151 There are two methods to activate control inputs. The in
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 152 4.15.2.2. Help Commands in Terminal Mode Table 4-31: Te
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 153 4.15.2.4. Data Types Data types consist of integers, hexa
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 154 possible messages, and this section for information on c
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 155 To change this setting press: SETUP X.X SECONDARY
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 156 To initialize the modem press: Select which COM Port i
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 157 LOGON 940331 940331 is the default password. To change t
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 158 Figure 4-18: APICOM Remote Control Program Interface A
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 159 only respond to commands from the host computer and only
Model 9110TH NOx Analyzer Table of Contents Teledyne Analytical Instruments xviii Figure 8-4: Reaction Cell During the AutoZero Cycle ...
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 160 The first step in configuring the 9110TH/M to operate o
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 161 and COM2, the TYPE selection affects both ports. 4.15.4.4
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 162 4.15.4.5. Hessen Protocol Gas ID Since the 9110TH/M meas
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 163 4.15.4.6. Setting Hessen Protocol Status Flags TAI implem
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 164 To assign or reset the status flag bit assignments, pres
Calibration Procedures Model 9110TH NOx Analyzer Teledyne Analytical Instruments 165 5. CALIBRATION PROCEDURES This section describes calibrat
Model 9110TH NOx Analyzer Calibration Procedures Teledyne Analytical Instruments 166 5.2. CALIBRATION PREPARATIONS 5.2.1. REQUIRED EQUIPMENT
Calibration Procedures Model 9110TH NOx Analyzer Teledyne Analytical Instruments 167 5.2.3. SPAN CALIBRATION GAS STANDARDS & TRACEABILITY
Model 9110TH NOx Analyzer Calibration Procedures Teledyne Analytical Instruments 168 5.2.4. DATA RECORDING DEVICES A strip chart recorder, d
Calibration Procedures Model 9110TH NOx Analyzer Teledyne Analytical Instruments 169 5.2.5.1. Determining / Updating the NO2 Converter Efficie
Table of Contents Model 9110TH NOx Analyzer Teledyne Analytical Instruments xix Table 4-21: DIAG - Analog I/O Functions ...
Model 9110TH NOx Analyzer Calibration Procedures Teledyne Analytical Instruments 170 STEP TWO: Set the expected NO2 span gas concentration:
Calibration Procedures Model 9110TH NOx Analyzer Teledyne Analytical Instruments 171 STEP THREE Activate NO2 measurement stability function. S
Model 9110TH NOx Analyzer Calibration Procedures Teledyne Analytical Instruments 172 STEP FOUR: Perform the converter efficiency calculatio
Calibration Procedures Model 9110TH NOx Analyzer Teledyne Analytical Instruments 173 5.3. MANUAL CALIBRATION The following section describes
Model 9110TH NOx Analyzer Calibration Procedures Teledyne Analytical Instruments 174 VENT here if input is pressurized Source ofSAMPLE Gas P
Calibration Procedures Model 9110TH NOx Analyzer Teledyne Analytical Instruments 175 STEP TWO: Set Expected NO and NOX Span Gas Concentration
Model 9110TH NOx Analyzer Calibration Procedures Teledyne Analytical Instruments 176 STEP THREE: Perform Zero/Span Calibration: Press ENTR
Calibration Procedures Model 9110TH NOx Analyzer Teledyne Analytical Instruments 177 5.4. CALIBRATION CHECKS Informal calibration checks, whic
Model 9110TH NOx Analyzer Calibration Procedures Teledyne Analytical Instruments 178 5.5. MANUAL CALIBRATION WITH ZERO/SPAN VALVES Zero and
Calibration Procedures Model 9110TH NOx Analyzer Teledyne Analytical Instruments 179 STEP TWO: Set Expected NO and NOX Span Gas Concentration
Model 9110TH NOx Analyzer Teledyne Analytical Instruments ii Copyright © 2013 Teledyne Analytical Instruments All Rights Reserved. No part of
Model 9110TH NOx Analyzer Table of Contents Teledyne Analytical Instruments xx This page intentionally left blank.
Model 9110TH NOx Analyzer Calibration Procedures Teledyne Analytical Instruments 180 STEP THREE: Perform Zero/Span Calibration: Press ENTR
Calibration Procedures Model 9110TH NOx Analyzer Teledyne Analytical Instruments 181 5.6. CALIBRATION CHECKS WITH ZERO/SPAN VALVES Zero and sp
Model 9110TH NOx Analyzer Calibration Procedures Teledyne Analytical Instruments 182 5.7. CALIBRATION WITH REMOTE CONTACT CLOSURES Contact c
Calibration Procedures Model 9110TH NOx Analyzer Teledyne Analytical Instruments 183 5.8. AUTOMATIC CALIBRATION (AUTOCAL) The AutoCal feature
Model 9110TH NOx Analyzer Calibration Procedures Teledyne Analytical Instruments 184 The following example sets sequence #2 to carry out a z
Calibration Procedures Model 9110TH NOx Analyzer Teledyne Analytical Instruments 185 To program the sample sequence shown above, follow this f
Model 9110TH NOx Analyzer Calibration Procedures Teledyne Analytical Instruments 186 5.9. CALIBRATION QUALITY ANALYSIS After completing one
Instrument Maintenance Model 9110TH NOx Analyzer Teledyne Analytical Instruments 187 6. INSTRUMENT MAINTENANCE Predictive diagnostic functions
Model 9110TH NOx Analyzer Instrument Maintenance Teledyne Analytical Instruments 188 This page intentionally left blank.
Instrument Maintenance Model 9110TH NOx Analyzer Teledyne Analytical Instruments 189 Table 6-1: 9110TH/M Preventive Maintenance Schedule ITEM
Introduction Model 9110TH NOx Analyzer Teledyne Analytical Instruments 1 1. INTRODUCTION, FEATURES, AND OPTIONS 1.1. OVERVIEW The Models 9110T
Model 9110TH NOx Analyzer Instrument Maintenance Teledyne Analytical Instruments 190 This page intentionally left blank.
Instrument Maintenance Model 9110TH NOx Analyzer Teledyne Analytical Instruments 191 6.3. PREDICTIVE DIAGNOSTICS The analyzer’s test function
Model 9110TH NOx Analyzer Instrument Maintenance Teledyne Analytical Instruments 192 and tweezers. We recommend not to touch any part of th
Instrument Maintenance Model 9110TH NOx Analyzer Teledyne Analytical Instruments 193 6.4.2. CHANGING THE O3 DRYER PARTICULATE FILTER The air
Model 9110TH NOx Analyzer Instrument Maintenance Teledyne Analytical Instruments 194 6.4.3. MAINTAINING THE EXTERNAL SAMPLE PUMP 6.4.3.1. Re
Instrument Maintenance Model 9110TH NOx Analyzer Teledyne Analytical Instruments 195 6.4.4. CHANGING THE NO2 CONVERTER The NO2 converter is l
Model 9110TH NOx Analyzer Instrument Maintenance Teledyne Analytical Instruments 196 9. Re-attach the tube fittings to the converter and re
Instrument Maintenance Model 9110TH NOx Analyzer Teledyne Analytical Instruments 197 Figure 6-4: Reaction Cell Assembly 4. The reaction ce
Model 9110TH NOx Analyzer Instrument Maintenance Teledyne Analytical Instruments 198 10. After cleaning, the analyzer span response may drop
Instrument Maintenance Model 9110TH NOx Analyzer Teledyne Analytical Instruments 199 5. Take out the components of the assembly: a spring, a
Model 9110TH NOx Analyzer Introduction Teledyne Analytical Instruments 2 1.4. OPTIONS Option Option Number Description/Notes Reference Pumps
Model 9110TH NOx Analyzer Instrument Maintenance Teledyne Analytical Instruments 200 This page intentionally left blank.
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 201 7. TROUBLESHOOTING & REPAIR This section conta
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 202 Suspect a leak first! Technical Support data i
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 203 The analyzer also issues an alert via the serial
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 204 SAMPLE SYSTEM RESET NOX = XXX.
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 205 Table 7-1: Test Functions - Possible Causes for O
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 206 SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 207 7.1.4. STATUS LED’S Several color-coded, light-emi
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 208 7.1.4.3. Relay Board and Status LEDs The most imp
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 209 Table 7-2: Relay Board Status LEDs LED COLOR FUN
Introduction Model 9110TH NOx Analyzer Teledyne Analytical Instruments 3 Option Option Number Description/Notes Reference Communication Cables
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 210 SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 211 7.2.1. 9110TH INTERNAL GAS FLOW DIAGRAMS Figure
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 212 Figure 7-6: 9110TH – Internal Gas Flow with Am
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 213 Figure 7-7: 9110TH – Internal Gas Flow with O
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 214 7.2.2. 9110TM INTERNAL GAS FLOW DIAGRAMS VACUUM
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 215 Figure 7-9: 9110TM – Internal Gas Flow with A
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 216 VACUUM PRESSURE SENSORSAMPLE PRESSURE SENSORO3 FL
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 217 If the pump is operating but the unit reports a XX
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 218 Check the actual flow rate through the ozone d
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 219 re-calibrate the flow electronically using the pro
Model 9110TH NOx Analyzer Introduction Teledyne Analytical Instruments 4 This page intentionally left blank.
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 220 A corruption of the Auto-zero filter may also
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 221 Check for disconnected cables to the sensor mod
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 222 Verify that the expected concentration is set
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 223 The standard gases may be mislabeled as to type
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 224 NO2 impurities in the NO calibration gas. NO
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 225 Wrong materials in contact with sample - use gl
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 226 A dirty reaction cell can cause high AutoZero
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 227 surfaces. Do not exceed 15 psi when pressurizing t
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 228 Attach the outlet port of a flow meter to the
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 229 Functions of the Relay PCA include: handling
Specifications Model 9110TH NOx Analyzer Teledyne Analytical Instruments 5 2. SPECIFICATIONS AND APPROVALS 2.1. 9110TH/M OPERATING SPECIFICATI
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 230 110 VAC /115 VAC 220 VAC /240 VAC 1 2 3 4 5 67
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 231 7.5.4.2. AC Configuration – Standard Heaters (JP2)
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 232 7.5.4.3. AC Configuration –Heaters for Option Pac
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 233 7.5.5. DC POWER SUPPLY TEST POINTS Table 7-6: DC
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 234 7.5.7. TOUCH SCREEN INTERFACE Verify the function
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 235 7.5.10. MOTHERBOARD 7.5.10.1. A/D functions A basi
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 236 7.5.10.3. Status Outputs The procedure below can
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 237 7.5.10.4. Control Inputs The control input bits ca
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 238 7.5.12. RS-232 COMMUNICATION 7.5.12.1. General RS
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 239 Use the RS-232 test function to send “w” charac
Model 9110TH NOx Analyzer Specifications Teledyne Analytical Instruments 6 2.2. APPROVALS AND CERTIFICATIONS The TAI Nitrogen Oxides Analyze
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 240 Divide the displayed HVPS voltage by 10 and test
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 241 7.5.16.2. Sample Pressure Measure the voltage acro
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 242 Temperature is reported properly but heater do
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 243 To avoid this problem, we recommend the use of a
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 244 7.5.20. PMT TEMPERATURE PMT temperature should be
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 245 7.6. REPAIR PROCEDURES This section contains some
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 246 7.6.2. O3 GENERATOR REPLACEMENT The ozone generat
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 247 6. Attach the replacement dryer to the mounting b
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 248 9. Set the HVPS coarse adjustment switch to the
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 249 7.6.5. REPLACING THE PMT, HVPS OR TEC The photo mu
Getting Started Model 9110TH NOx Analyzer Teledyne Analytical Instruments 7 3. GETTING STARTED 3.1. UNPACKING AND INITIAL SETUP CAUTION THE 9
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 250 Figure 7-18: 9110TH/M Sensor Assembly Figure 7
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 251 9. Discard the plastic screws and replace with ne
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 252 7.7. REMOVING / REPLACING THE RELAY PCA FROM THE
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 253 7.8. FREQUENTLY ASKED QUESTIONS The following list
Model 9110TH NOx Analyzer Troubleshooting & Repair Teledyne Analytical Instruments 254 QUESTION ANSWER Can I use the DAS system in place
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 255 This page intentionally left blank.
Model 9110TH NOx Analyzer Principles of Operation Teledyne Analytical Instruments 256 8. PRINCIPLES OF OPERATION The 9110TH/M Nitrogen Oxides
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 257 νhNONO +2*2→ (Equation 9-2) All things being const
Model 9110TH NOx Analyzer Principles of Operation Teledyne Analytical Instruments 258 8.1.2. NOX AND NO2 DETERMINATION The only gas that is t
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 259 8.2. CHEMILUMINESCENCE DETECTION 8.2.1. THE PHOTO
Model 9110TH NOx Analyzer Getting Started Teledyne Analytical Instruments 8 b. Remove the 2 screws fastening the top cover to the unit (o
Model 9110TH NOx Analyzer Principles of Operation Teledyne Analytical Instruments 260 8.2.3. AUTO ZERO Inherent in the operation of any PMT i
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 261 8.2.4. MEASUREMENT INTERFERENCES It should be d th
Model 9110TH NOx Analyzer Principles of Operation Teledyne Analytical Instruments 262 Table 8-1: List of Interferents GAS INTERFERENCE TYPE
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 263 8.3. PNEUMATIC OPERATION Note It is important tha
Model 9110TH NOx Analyzer Principles of Operation Teledyne Analytical Instruments 264 Figure 8-5: External Pump Pack Finally, the 9110TH/M
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 265 8.3.2.1. NO/NOx and AutoZero cycles For the routin
Model 9110TH NOx Analyzer Principles of Operation Teledyne Analytical Instruments 266 VACUUM PRESSURE SENSORSAMPLE PRESSURE SENSORO3 FLOW SE
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 267 VACUUM PRESSURE SENSORSAMPLE PRESSURE SENSORO3 FLO
Model 9110TH NOx Analyzer Principles of Operation Teledyne Analytical Instruments 268 8.3.3.1. Critical Flow Orifice The most important compo
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 269 The following table lists the gas flow rates of th
Getting Started Model 9110TH NOx Analyzer Teledyne Analytical Instruments 9 Figure 3-1: 9110TH Front Panel Figure 3-2: Display Screen and
Model 9110TH NOx Analyzer Principles of Operation Teledyne Analytical Instruments 270 CE Mark safety screw), and should be changed according
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 271 8.3.6. O3 GENERATOR The 9110TH/M uses a corona dis
Model 9110TH NOx Analyzer Principles of Operation Teledyne Analytical Instruments 272 8.3.7. PERMA PURE® DRYER The air supplied to the O3 gen
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 273 very large and the dryer’s efficiency is low at fi
Model 9110TH NOx Analyzer Principles of Operation Teledyne Analytical Instruments 274 8.3.8. OZONE SUPPLY AIR FILTER The 9110TH/M uses ambien
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 275 8.3.10. PNEUMATIC SENSORS Note The 9110TH/M displ
Model 9110TH NOx Analyzer Principles of Operation Teledyne Analytical Instruments 276 flow condition (2:1 pressure ratio) through the sample
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 277 Figure 8-13: Dilution Manifold Please inquire wi
Model 9110TH NOx Analyzer Principles of Operation Teledyne Analytical Instruments 278 Figure 8-14: Oxygen Sensor - Principle of Operation 8
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 279 8.5. ELECTRONIC OPERATION Figure 8-15 shows a bloc
Model 9110TH NOx Analyzer Teledyne Analytical Instruments iii INFORMATION ABOUT THE SPECIFIC CONFIGURATION OF YOUR MODEL 9110T NOX ANALYZER Note
Model 9110TH NOx Analyzer Getting Started Teledyne Analytical Instruments 10 The front panel liquid crystal display screen includes touch co
Model 9110TH NOx Analyzer Principles of Operation Teledyne Analytical Instruments 280 commands to a variety of peripheral devices through the
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 281 In the unlikely event that the flash chip should f
Model 9110TH NOx Analyzer Principles of Operation Teledyne Analytical Instruments 282 8.5.3. PHOTO MULTIPLIER TUBE (PMT) The 9110TH/M uses a
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 283 A typical PMT is a vacuum tube containing a variet
Model 9110TH NOx Analyzer Principles of Operation Teledyne Analytical Instruments 284 8.5.4. PMT COOLING SYSTEM The performance of the analyz
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 285 be adjusted through a potentiometer. These adjust
Model 9110TH NOx Analyzer Principles of Operation Teledyne Analytical Instruments 286 8.5.6. PNEUMATIC SENSOR BOARD The flow and pressure sen
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 287 Thermistor(s) – Low Temperature Sensing:(e.g. Samp
Model 9110TH NOx Analyzer Principles of Operation Teledyne Analytical Instruments 288 Table8-4: Thermocouple Configuration Jumper (JP5) Pin-
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 289 Table 8-5: Typical Thermocouple Settings TC TYP
Getting Started Model 9110TH NOx Analyzer Teledyne Analytical Instruments 11 Figure 3-3: Display/Touch Control Screen Mapped to Menu Charts
Model 9110TH NOx Analyzer Principles of Operation Teledyne Analytical Instruments 290 8.5.8. STATUS LEDS & WATCH DOG CIRCUITRY Thirteen L
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 291 8.5.9. MOTHERBOARD This is the largest electronic
Model 9110TH NOx Analyzer Principles of Operation Teledyne Analytical Instruments 292 PMT TEMPERATURE: This signal is the output of the therm
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 293 8.5.10. ANALOG OUTPUTS The analyzer comes equipped
Model 9110TH NOx Analyzer Principles of Operation Teledyne Analytical Instruments 294 8.6. POWER DISTRIBUTION & CIRCUIT BREAKER The analy
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 295 8.7. FRONT PANEL/DISPLAY INTERFACE ELECTRONICS Use
Model 9110TH NOx Analyzer Principles of Operation Teledyne Analytical Instruments 296 8.8. SOFTWARE OPERATION The instrument’s core module is
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 297 8.8.1. ADAPTIVE FILTER The 9110TH/M NOX analyzer s
Model 9110TH NOx Analyzer Principles of Operation Teledyne Analytical Instruments 298 8.8.3. TEMPERATURE/PRESSURE COMPENSATION (TPC) The soft
Troubleshooting & Repair Model 9110TH NOx Analyzer Teledyne Analytical Instruments 299 (Equation 9-9) The preset gain parameters are set
Model 9110TH NOx Analyzer Getting Started Teledyne Analytical Instruments 12 Figure 3-4: 9110TH/M Rear Panel Layout Table 3-2: Rear Panel
ESD Primer Model 9110TH NOx Analyzer Teledyne Analytical Instruments 301 9. A PRIMER ON ELECTRO-STATIC DISCHARGE TAI considers the prevention
Model 9110TH NOx Analyzer ESD Primer Teledyne Analytical Instruments 302 Table 9-1: Static Generation Voltages for Typical Activities MEANS
ESD Primer Model 9110TH NOx Analyzer Teledyne Analytical Instruments 303 can occur from a charged conductive object, such as a metallic tool
Model 9110TH NOx Analyzer ESD Primer Teledyne Analytical Instruments 304 still trapped and can still build to high enough levels to cause dam
ESD Primer Model 9110TH NOx Analyzer Teledyne Analytical Instruments 305 Simply touching a grounded piece of metal is insufficient. While
Model 9110TH NOx Analyzer ESD Primer Teledyne Analytical Instruments 306 1. Attach your anti-ESD wrist strap to ground before doing anything
ESD Primer Model 9110TH NOx Analyzer Teledyne Analytical Instruments 307 9.4.2.3. Transferring Components from Rack to Bench and Back When tr
Model 9110TH NOx Analyzer ESD Primer Teledyne Analytical Instruments 308 9.4.2.4. Opening Shipments from Teledyne Packing materials such as
ESD Primer Model 9110TH NOx Analyzer Teledyne Analytical Instruments 309 3. Place the item in the container. 4. Seal the container. If usi
Getting Started Model 9110TH NOx Analyzer Teledyne Analytical Instruments 13 Figure 3-5: 9110TH/M Internal Layout
Glossary Model 9110TH NOx Analyzer Teledyne Analytical Instruments 311 This page intentionally left blank.
Model 9110TH NOx Analyzer Glossary Teledyne Analytical Instruments 312 10. GLOSSARY Note: Some terms in this glossary may not occur elsewhere
Glossary Model 9110TH NOx Analyzer Teledyne Analytical Instruments 313 Term Description/Definition include nitrate (NO3), PAN, N2O and other
Model 9110TH NOx Analyzer Glossary Teledyne Analytical Instruments 314 Term Description/Definition GFC Gas Filter Correlation I2C bus a cloc
Glossary Model 9110TH NOx Analyzer Teledyne Analytical Instruments 315 Term Description/Definition handle gases that may react on other surfa
Model 9110TH NOx Analyzer Getting Started Teledyne Analytical Instruments 14 3.4. ELECTRICAL CONNECTIONS Note To maintain compliance with E
Getting Started Model 9110TH NOx Analyzer Teledyne Analytical Instruments 15 3.4.2. ANALOG INPUTS (OPTION 64) CONNECTIONS The Analog In connec
Model 9110TH NOx Analyzer Getting Started Teledyne Analytical Instruments 16 Table 3-1: Analog Output Data Type Default Settings PARAMETER
Getting Started Model 9110TH NOx Analyzer Teledyne Analytical Instruments 17 EMITTER BUS FOR PINS 1-8 STATUS 1 2 3 4
Model 9110TH NOx Analyzer Getting Started Teledyne Analytical Instruments 18 3.4.5. CURRENT LOOP ANALOG OUTPUTS (OPT 41) SETUP This option a
Getting Started Model 9110TH NOx Analyzer Teledyne Analytical Instruments 19 3.4.5.1. Converting Current Loop Analog Outputs to Standard Volta
Model 9110TH NOx Analyzer Teledyne Analytical Instruments iv Power Requirements This Model 9110E is configured to operate from the following
Model 9110TH NOx Analyzer Getting Started Teledyne Analytical Instruments 20 3.4.6. CONNECTING THE CONTROL INPUTS Control Inputs are used to
Getting Started Model 9110TH NOx Analyzer Teledyne Analytical Instruments 21 3.4.7. CONNECTING THE ALARM RELAY OPTION (OPT 61) The 9110TH/M ca
Model 9110TH NOx Analyzer Getting Started Teledyne Analytical Instruments 22 Table 3-7: Concentration Alarm Relay Output Operation RELAY FU
Getting Started Model 9110TH NOx Analyzer Teledyne Analytical Instruments 23 3.4.8.3. Connecting to a Multidrop Network The multidrop option i
Model 9110TH NOx Analyzer Getting Started Teledyne Analytical Instruments 24 3.5. PNEUMATIC CONNECTIONS Note To prevent dust from getting i
Getting Started Model 9110TH NOx Analyzer Teledyne Analytical Instruments 25 Even though NO gas in nitrogen could be used as a span gas, the m
Model 9110TH NOx Analyzer Getting Started Teledyne Analytical Instruments 26 3.5.2. PNEUMATIC CONNECTIONS TO 9110TH/M BASIC CONFIGURATION Fi
Getting Started Model 9110TH NOx Analyzer Teledyne Analytical Instruments 27 Source of SAMPLE GASRemoved during calibrationInstrumentChassisSA
Model 9110TH NOx Analyzer Getting Started Teledyne Analytical Instruments 28 1. Attach a 1/4" exhaust line between the external pump a
Getting Started Model 9110TH NOx Analyzer Teledyne Analytical Instruments 29 Figure 3-15: 9110TH Internal Pneumatic Block Diagram - Standar
Model 9110TH NOx Analyzer Teledyne Analytical Instruments v Valve Options No Valves Internal SS Zero/Span Valves Second Range Span Valve Internal
Model 9110TH NOx Analyzer Getting Started Teledyne Analytical Instruments 30 VACUUM PRESSURE SENSORSAMPLEPRESSURE SENSORO3 FLOW SENSORFLOW
Getting Started Model 9110TH NOx Analyzer Teledyne Analytical Instruments 31 3.5.3. CONNECTIONS WITH INTERNAL VALVE OPTIONS INSTALLED If your
Model 9110TH NOx Analyzer Getting Started Teledyne Analytical Instruments 32 3.5.3.1. Ambient Zero/Ambient Span Valves (OPT 50A) The Model 9
Getting Started Model 9110TH NOx Analyzer Teledyne Analytical Instruments 33 Figure 3-20: 9110TM – Internal Pneumatics with Ambient Zero-Spa
Model 9110TH NOx Analyzer Getting Started Teledyne Analytical Instruments 34 All supply lines should be vented outside of the analyzer’s enc
Getting Started Model 9110TH NOx Analyzer Teledyne Analytical Instruments 35 Table 3-11: Two-Point Span Valve Operating States MODE VALVE CO
Model 9110TH NOx Analyzer Getting Started Teledyne Analytical Instruments 36 This page intentionally left blank.
Teledyne API - Model T200H/T200M Operation Manual Getting Started Teledyne Analytical Instruments 37 Figure 3-21: 9110TH - Internal Pneumatics
Model 9110TH NOx Analyzer Getting Started Teledyne Analytical Instruments 38 Figure 3-22: 9110TM - Internal Pneumatics for Zero Scrubber/D
Getting Started Model 9110TH NOx Analyzer Teledyne Analytical Instruments 39 3.5.3.3. Internal Flow for O2 Sensor Option 65A Please see Sectio
Model 9110TH NOx Analyzer Teledyne Analytical Instruments vi This page intentionally left blank.
Model 9110TH NOx Analyzer Getting Started Teledyne Analytical Instruments 40 Figure 3-24: 9110TM – Internal Pneumatics with O2 Sensor Opti
Getting Started Model 9110TH NOx Analyzer Teledyne Analytical Instruments 41 3.6. INITIAL OPERATION CAUTION! If the presence of ozone is dete
Model 9110TH NOx Analyzer Getting Started Teledyne Analytical Instruments 42 3.6.3. FUNCTIONAL CHECK After the analyzer’s components have wa
Getting Started Model 9110TH NOx Analyzer Teledyne Analytical Instruments 43 3.7. CALIBRATION An initial calibration and functional check shou
Model 9110TH NOx Analyzer Getting Started Teledyne Analytical Instruments 44 STEP 1 - Set Units: To select the concentration units of measur
Getting Started Model 9110TH NOx Analyzer Teledyne Analytical Instruments 45 STEP 2 - Dilution Ratio: If the dilution ratio option is enabled
Model 9110TH NOx Analyzer Getting Started Teledyne Analytical Instruments 46 STEP 3 – Set NOx and NO span gas concentrations : Set the expe
Getting Started Model 9110TH NOx Analyzer Teledyne Analytical Instruments 47 STEP 4 – Zero/Span Calibration : To perform the zero/span calibra
Model 9110TH NOx Analyzer Getting Started Teledyne Analytical Instruments 48 3.7.2. BASIC O2 SENSOR CALIBRATION PROCEDURE If your instrument
Getting Started Model 9110TH NOx Analyzer Teledyne Analytical Instruments 49 3.7.2.2. O2 Calibration Method STEP 1 – Set O2 span gas concentr
Model 9110TH NOx Analyzer Safety Messages Teledyne Analytical Instruments vii SAFETY MESSAGES Important safety messages are provided throughou
Model 9110TH NOx Analyzer Getting Started Teledyne Analytical Instruments 50 STEP 2 – Activate O2 sensor stability function To change the st
Getting Started Model 9110TH NOx Analyzer Teledyne Analytical Instruments 51 STEP 4 – O2 ZERO/SPAN CALIBRATION : To perform the zero/span cali
Model 9110TH NOx Analyzer Getting Started Teledyne Analytical Instruments 52 This page intentionally left blank.
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 53 4. OPERATING INSTRUCTIONS To assist in navigating the anal
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 54 The mode field of the front panel display indicates to th
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 55 in one of the DAS channels (Section 4.7) for data analysis
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 56 A1:NXCNC1=100 PPM1 A2:NOCNC1=100 PPM1 A3 :N2 CNC1=2 5 P
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 57 Table 4-3: List of Warning Messages MESSAGE DEFINITION A
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 58 4.3. CALIBRATION MODE 4.3.1. CALIBRATION FUNCTIONS Pressi
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 59 Table 4-5: Secondary Setup Mode Features and Functions MO
Model 9110TH NOx Analyzer Safety Messages Teledyne Analytical Instruments viii CONSIGNES DE SÉCURITÉ Des consignes de sécurité importantes
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 60 SAMPLE A1:NXCNC1=100PPM NOX=XXX.X
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 61 4.7. SETUP DAS - USING THE DATA ACQUISITION SYSTEM (DAS
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 62 4.7.1. DAS STRUCTURE The DAS is designed around the featu
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 63 4.7.1.2. DAS Parameters Data parameters are types of data
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 64 WARNINGS: Some data may be useful when stored if one o
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 65 The default data channels can be used as they are, or they
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 66 Table 4-9: 9110TH/M Default DAS Configuration CHANNELS w
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 67 4.7.2.1. Viewing DAS Data and Settings DAS data and settin
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 68 4.7.2.2. Editing DAS Data Channels DAS configuration is m
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 69 To edit the name of a data channel, follow the above key s
Model 9110TH NOx Analyzer Safety Messages Teledyne Analytical Instruments ix This page intentionally left blank.
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 70 4.7.2.4. Editing DAS Parameters Data channels can be edit
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 71 To configure the parameters for a specific data parameter,
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 72 4.7.2.5. Sample Period and Report Period The DAS defines
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 73 4.7.2.6. Report Periods in Progress when Instrument Is Pow
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 74 To define the REPORT PERIOD, follow the instruction shown
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 75 4.7.2.7. Number of Records The DAS is capable of capturing
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 76 4.7.2.8. RS-232 Report Function The 9110TH/M DAS can auto
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 77 4.7.2.11. Disabling/Enabling Data Channels Data channels c
Model 9110TH NOx Analyzer Operating Instructions Teledyne Analytical Instruments 78 4.7.2.12. HOLDOFF Feature The DAS HOLDOFF feature allows
Operating Instructions Model 9110TH NOx Analyzer Teledyne Analytical Instruments 79 4.7.3. REMOTE DAS CONFIGURATION Editing channels, paramete