  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 43
NOTE
REMOTE ZERO AND SPAN ADJUSTMENT ALTER THE
FORMULAS USED TO COMPUTE THE CARBON
MONOXIDE READINGS IF DYN_ZERO OR DYN_SPAN ARE
ENABLED. THIS METHOD OF CALIBRATION IS NOT
APPROVED BY USEPA AND IS NOT INTENDED TO
REPLACE THE USEPA APPROVED CALIBRATION.
3.7 Remote zero/span check or adjustment
(RS-232)
Besides Z/S checking from the front panel, automatic Z/S checking, and remote Z/S
checking via the contact closure inputs, the Analyzer can also be checked via the RS-232
interface. Remote checking via the RS-232 interface supports zero, span, and zero
followed by span check, and is identical to remote check via the contact closure inputs.
This RS-232 control feature is provided mainly so that a host computer at another
location can control the Analyzer. See Section 7.0 for detailed information on using the
RS-232 interface to do a remote Z/S check.
3.8 Power-on hold off
Whenever the Model 300 is powered on it will go through a HOLD-OFF sequence (see
Section 3.10 below) like it does after a zero/span check.
3.9 Hold off
Every type of check or adjustment (zero, span, manual, remote, etc.) is followed by a
hold off period of from 1 to 20 minutes, during which time the internal data acquisition
system (DAS) does not accumulate CO readings into the DAS average. To set the hold
off time, press SETUP-MORE-VARS and press NEXT until the variable
HOLDOFF_TIME is displayed. To change the setting, enter a number from 1 to 20 and
then press ENTR to store the new value or EXIT to leave the hold off time unchanged.
P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 38
Page view 43