absence of these actions will usually indicate either a CPU or Display failure. To
determine which module is defective, perform the following procedure:
1. Turn off power.
2. Remove the ribbon cable from the CPU board to the Display;
3. Turn Power on.
4. A cursor character should appear in the upper left corner of the display. If
it does not, the display is defective and should be replaced. If the cursor
does appear, it is probable that the CPU is faulty.
10.2.3 Checking the keyboard
During normal Analyzer operation, depressing the right most key of the keyboard should
cause a change of display modes. If it does not, check:
- Cable connections;
- CPU and Display operation (see Section 10.2.2)
If these checks are satisfactory, it is probable that the keyboard is defective and should be
replaced.
10.3 Troubleshooting using warning messages
The most common and/or serious instrument failures will result in a warning message (or
messages) being displayed on the front panel. Table 10-1 below lists the warning
messages which the Analyzer may display, along with their meaning and the
recommended corrective action. It should be noted that if multiple (more than 2 or 3)
warning messages occur at the same time, it is often an indication that some fundamental
analyzer sub-system (power supply, V/F board, CPU) has failed rather than an indication
of the multiple failures referenced by the warnings. In this situation, it is recommended
that proper operation of power supplies (see Section 10.6.4) and the V/F Board (see
Section 10.6.3) be confirmed before addressing the specific warning messages.
WARNING MESSAGES
Warning Message Meaning Corrective Action
SOURCE WARNING The CO REF value is
greater than 5000 mV or
less than 2500 mV
Check and adjust the
Sync Demodulator and
optical alignment as
described in Section
10.6.6 and 10.6.8
SYNC ERROR No modulation is present
on the output of the IR
detector.
Check IR source, IR
detector/pre-amp, and
Opto interrupter
P/N 02163G1 Teledyne API Model 300 CO Analyzer Instruction Manual - Page 85