7.5 Test measurements
All the test measurements which can be displayed by pressing the TEST button are also
available to the host computer via the RS-232 interface. The host computer should issue
a request for a test measurement, and then the CPU will send the current value of the test
measurement to the RS-232 output. The format of the test measurement message is:
"T DDD:HH:MM IIII TEST MEASUREMENT<CRLF>"
For example, the format of the DC Power Supply output in millivolts would be:
"T 194:11:29 0000 DCPS= 2500 MV<CRLF>"
To request a test measurement, the host must issue a command of the form:
"T MEASUREMENT<CRLF>"
For a summary of all test functions issue the command "T LIST". The table below lists
the commands and the corresponding test measurements which will be returned.
TEST MEASUREMENT REQUEST COMMANDS
Command Test measurement
"?<CRLF>" RS-232 HELP screen
"T LIST<CRLF>" Summary of all TEST's
"T CO<CRLF>" Current CO reading
"T COMEAS<CRLF>" Current CO MEAS mV
"T COREF<CRLF>" Current CO REF mV
"T MRRATIO<CRLF>" Current MR RATIO
"T SPRESS<CRLF>" Sample pressure
"T SFLOW<CRLF>" Sample flow rate
"T STEMP<CRLF>" Sample temperature
"T BNCHTEMP<CRLF>" Optical Bench temperature
"T WHEELTEMP<CRLF>" Filter Wheel temperature
"T BOXTEMP<CRLF>" Internal box temperature
"T DCPS<CRLF>" DC power supply output
"T COSLOPE<CRLF>" Slope value
"T COFFSET<CRLF>" Offset value
"T CLKTIME<CRLF>" Current time-of-day