Trace Oxygen Analyzer
Teledyne Analytical Instruments i
OPERATING INSTRUCTIONS FOR
MODEL OT-3MODEL OT-3
Trace Oxygen Analyzer
DANGER
Toxic gases and or flammable liquids may be present in this monitoring system.
Personal protective equipment may be required when servicing this instrument.
Hazardous voltages exist on certain components internally which may persist
for a time even after the power is turned off and disconnected.
Only authorized personnel should conduct maintenance and/or servicing.
Before conducting any maintenance or servicing, consult with authorized
supervisor/manager.
P/N M74173
10/28/2001
ECO#
Page view 0
1 2 3 4 5 6 ... 49 50

Summary of Contents

Page 1 - MODEL OT-3MODEL OT-3

Trace Oxygen AnalyzerTeledyne Analytical Instruments iOPERATING INSTRUCTIONS FORMODEL OT-3MODEL OT-3Trace Oxygen AnalyzerDANGERToxic gases and or flam

Page 2 - Model OT-3

Introduction OT-3Teledyne Analytical Instruments 2The Model 3300/OT-3 Control Unit has its own long list of featuresincluding:• Keypad lockout featu

Page 3 - Specific Model Information

Trace Oxygen Analyzer IntroductionTeledyne Analytical Instruments 3Figure 1-1: Front PanelFunction Keys:Seven push-button membrane switches are used

Page 4 - Important Notice

Introduction OT-3Teledyne Analytical Instruments 4Meter Readout: As the meter readout, it displays the oxygenconcentration currently being measured.

Page 5

Trace Oxygen Analyzer IntroductionTeledyne Analytical Instruments 5Figure 1-2: Rear Panel (AC Version)Figure 1-3: Rear Panel (DC Version)

Page 6 - Table of Contents

Installation OT-3Teledyne Analytical Instruments 6Operational Theory2.1 IntroductionThe analyzer is composed of two subsystems:• 3300/OT-3 control u

Page 7

Trace Oxygen Analyzer InstallationTeledyne Analytical Instruments 7between a battery and a true fuel cell. (All of the reactants are storedexternally

Page 8 - List of Figures

Installation OT-3Teledyne Analytical Instruments 8electrical circuitry, to one of the external contacts in the connectorreceptacle, which is on the to

Page 9 - Introduction

Trace Oxygen Analyzer InstallationTeledyne Analytical Instruments 92.2.4 The Effect of PressureIn order to state the amount of oxygen present in the s

Page 10 - 1.3 Front Panel Description

Installation OT-3Teledyne Analytical Instruments 10Figure 2-2: Characteristic MFC Input/Output Curve2.3 Electronics2.3.1 GeneralThe signal processin

Page 11

Trace Oxygen Analyzer InstallationTeledyne Analytical Instruments 112.3.2 Signal ProcessingFigure 2-3 is a block diagram of the signal processing elec

Page 12 - 1.4 Rear Panel Description

Model OT-3Teledyne Analytical Instruments iiCopyright © 2000 Teledyne Analytical InstrumentsAll Rights Reserved. No part of this manual may be reprodu

Page 13

Installation OT-3Teledyne Analytical Instruments 12The same digital information is also sent to a 12-bit digital to analogconverter (DAC) that produce

Page 14 - Operational Theory

Trace Oxygen Analyzer InstallationTeledyne Analytical Instruments 13InstallationInstallation of the analyzer includes:• Unpacking the system.• Insta

Page 15

Installation OT-3Teledyne Analytical Instruments 14CAUTION: DO NOT DISTURB THE INTEGRITY OF THE CELLPACKAGE UNTIL THE CELL IS TO ACTUALLY BEUSED. IF T

Page 16

Trace Oxygen Analyzer InstallationTeledyne Analytical Instruments 15Figure 3-1: Electrical Connectors for AC Control Unit3.3.1 Primary Input PowerA d

Page 17 - in the electronics. (The

Installation OT-3Teledyne Analytical Instruments 16• 4–20 mA % Range: Current increases with increasing oxygenconcentration, from 4 mA at 0 percent o

Page 18 - 2.3 Electronics

Trace Oxygen Analyzer InstallationTeledyne Analytical Instruments 17The specific descriptions for each type of alarm are as follows:Power Fail Alarm:

Page 19

Installation OT-3Teledyne Analytical Instruments 18any unknown power source, or any load or device that may exceed thesafe use ratings for Cl 1 Div 2

Page 20

Trace Oxygen Analyzer InstallationTeledyne Analytical Instruments 19• Vent connection• Coalescing filter drainNote: Zero gas is not required however

Page 21 - Installation

Installation OT-3Teledyne Analytical Instruments 20The vent line must be installed in a manner which will prevent dirtand water from accumulating with

Page 22 - 3.2 Location and Mounting

Trace Oxygen Analyzer InstallationTeledyne Analytical Instruments 21A flowmeter and flow restrictor are parts of the sample system andassist in settin

Page 23

Trace Oxygen AnalyzerTeledyne Analytical Instruments iiiSpecific Model InformationThe instrument for which this manual was supplied mayincorporate one

Page 24

Installation OT-3Teledyne Analytical Instruments 22

Page 25 - REMOVED FROM THE INSTRUMENT

Trace Oxygen Analyzer OperationTeledyne Analytical Instruments 23Operation4.1 IntroductionOnce the analyzer has been mounted, the gas lines connected

Page 26 - 3.4 Gas Connections

Operation OT-3Teledyne Analytical Instruments 244.2 Using the Function and Data Entry ButtonsWhen no buttons on the Analyzer are being pressed, the i

Page 27 - FITTING ON THE INSTRUMENT

Trace Oxygen Analyzer OperationTeledyne Analytical Instruments 254.3.1 HI RangeSetting the HI Range full-scale value defines the LEAST sensitiveanalys

Page 28 - 3.5 Sample System

Operation OT-3Teledyne Analytical Instruments 26until the display reads the desired percent concentration. (ifwithin 5 seconds no key is pressed, the

Page 29 - 3.6 Installation Checklist

Trace Oxygen Analyzer OperationTeledyne Analytical Instruments 27When the alarms are set for delay and an alarm condition exists, thedisplay will alte

Page 30 - Installation OT-3

Operation OT-3Teledyne Analytical Instruments 28will cause the display to flash NO PASS and the keys willhave no effect.4. To unlock the keypad, repea

Page 31 - Operation

Trace Oxygen Analyzer OperationTeledyne Analytical Instruments 29Note: If the analyzer output goes above the high alarm setpointor below the low alarm

Page 32 - • SET SPAN

Operation OT-3Teledyne Analytical Instruments 30OR ALARM BUTTONS ARE PRESSED DURINGSPAN. THIS MAY RESULT IN FALSE ALARMS IFTHE SPAN GAS HAS NOT BEEN

Page 33 - ∆ or ∇ button to

Trace Oxygen Analyzer MaintenanceTeledyne Analytical Instruments 31MaintenanceAside from normal cleaning and checking for leaks at the gasconnections,

Page 34 - 4.5 Alarm Delay

Model OT-3Teledyne Analytical Instruments ivImportant NoticeModel OT-3 complies with all of the requirements of theCommonwealth of Europe (CE) for Rad

Page 35 - 4.6 Keypad Lockout

Maintenance OT-3Teledyne Analytical Instruments 32Figure 5-1: AC Fuse ReplacementDC UnitsOn DC units, the fuse is located in a fuse holder on the rea

Page 36 - 4.8 Calibration

Trace Oxygen Analyzer MaintenanceTeledyne Analytical Instruments 33CAUTION: READ THE SECTION CELL WARRANTYCONDITIONS, BELOW, BEFORE REPLACING THECELL.

Page 37 - (THE USER’S) DISCRETION

Maintenance OT-3Teledyne Analytical Instruments 342. Open the analyzer door.3. Unscrew the cell-retainer cap from the cell block by turning itcounterc

Page 38 - Operation OT-3

Trace Oxygen Analyzer MaintenanceTeledyne Analytical Instruments 355.2.5 Cell Warranty ConditionsThe B-2C , A-2C, and Insta-Trace Class Micro-fuel Cel

Page 39 - Maintenance

Appendix OT-3Teledyne Analytical Instruments 36AppendixA.1 SpecificationsRanges: Two user selectable ranges can be setbetween 0-10 ppm and 0-9999 ppmo

Page 40

Trace Oxygen Analyzer AppendixTeledyne Analytical Instruments 37A.2 Spare Parts ListQTY P/N DESCRIPTION1 C65220-A PC Board, Main (std.)1 C70740-A PC B

Page 41

Appendix OT-3Teledyne Analytical Instruments 38A.3 Reference DrawingD-74110 Outline DiagramB-74106 Piping DiagramD-65305 Schematic main PCBD-70739 Sch

Page 42

Trace Oxygen Analyzer IndexTeledyne Analytical Instruments 39Indexaccuracy, 2, 36address. See company addressalarmconnection, 16delay, 26HI, 16LO, 16p

Page 43

Index OT-3Teledyne Analytical Instruments 40international power, 10jumper, 18keypad lockout, 2, 27lead, 7, 8linear curve, 9LO Range, 25location, 14mai

Page 44 - Appendix

Trace Oxygen Analyzer IndexTeledyne Analytical Instruments 41terminal block, 14, 15thermistor, 11universal power supply. See powersupply moduleVAC sym

Page 45 - A.2 Spare Parts List

Trace Oxygen AnalyzerTeledyne Analytical Instruments vThis instrument is designed for CL1 DIV II hazardous locations. Itis the customer's respon

Page 46 - A.4 Miscellaneous

Index OT-3Teledyne Analytical Instruments 42

Page 47 - Trace Oxygen Analyzer Index

Model OT-3Teledyne Analytical Instruments viTable of ContentsList of Figures...

Page 48 - Index OT-3

Trace Oxygen AnalyzerTeledyne Analytical Instruments vii3.4 Gas Connections 183.5 Sample System 203.6 Installation Checklist 21Operation...

Page 49

Model OT-3Teledyne Analytical Instruments viiiList of FiguresFigure 1-1: Front Panel.....

Page 50

Trace Oxygen Analyzer IntroductionTeledyne Analytical Instruments 1Introduction1.1 OverviewThe Teledyne Electronic Technologies Analytical Instrument