Introduction OT-3
Teledyne Analytical Instruments 2
The Model 3300/OT-3 Control Unit has its own long list of features
including:
Keypad lockout feature
Two user settable fail-safe alarms (with a user settable 0-20
min. time delay )
Two user settable ranges (from 0-10PPM – 0-9999PPM)
with user selectable auto-ranging.
Power fail alarm
SSR power output (Solid State Relay power available on AC
control units)
Micro processor based electronics 8 bit ADC & 18 bit DAC
Large 3 ½ digit LED O2 concentration read-out
7 button user interface keypad
Alarm condition indicator LED
100-240 VAC 50/60 Hz power (optional 9-36VDC )
0-10 VDC output
4-20 mA output
0-10 VDC range ID output
Electronic temperature compensation
Optional RS323 interface
Accuracy: +/- 2% of range (at constant temp.)
Response time: 90% in 65 sec of less @ 25 degrees C
Temperature range (0-50 degrees C)
1.3 Front Panel Description
All controls are accessible from the front panel. See Figure 1-1.
The front panel has seven push-button membrane switches, a digital
meter, and an alarm indicator LED for operating the analyzer. These
features are described briefly here and in greater detail in Chapter 4,
Operation.