Total Organic Carbon Analyzer Maintenance
Teledyne Analytical Instruments 45
5.3 Module Service
5.3.1 NDIR (Gas Calibration)
1. From the Menu, select [CALIBRATE].
The following screen will appear:
2. Select [INFRARED ANALYZER]. The following
Screen will appear:
3. Disconnect tubing from reactor output and connect IR
calibration adapter (P/N ST20026) as shown in Figures 5-
1, 5-2, and 5-3.
4. Connect facility tubing to “Zero” and “Span” Gas Bottles
Page view 56