INSTRUCTION MANUAL MODEL 9110AH NITROGEN OXIDES ANALYZER SERIAL NO. _______________ TELEDYNE ANALYTICAL INSTRUMENTS 16830 CHESTNUT STREET
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F ix TABLE 9-1: TEST FUNCTIONS ...
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 7-12 Table 7-10: Converter Efficiency - Automatic Calculation Step Number Action
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 7-13 Use NO in nitrogen at 80% of calibration range 2% for routine calibration. E
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 7-14 INTENTIONALLY BLANK
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 8-1 8 MAINTENANCE 8.1 Maintenance Schedule NOTE The operations outlined in this c
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 8-2 8.2 Replacing the Sample Particulate Filter The particulate filter should be i
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 8-3 Figure 8-1: Replacing the Particulate Filter
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 8-4 8.3 Replacing the Converter The Converter is located in the center of the inst
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 8-5 Figure 8-2: Converter Assembly
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 8-6 8.4 Cleaning the Reaction Cell The reaction cell should be cleaned whenever tr
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 8-7 10. Reassemble in the reverse order of disassembly, being sure to seat the pape
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F x INTENTIONALLY BLANK
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 8-8 Figure 8-3: Reaction Cell Assembly
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 8-9 Figure 8-4: Sample/Bypass Flow Control Assembly
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 8-10 8.5 Pneumatic Line Inspection Particulate matter in the pneumatic lines will
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 8-11 Figure 8-5: Pneumatic Diagram
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 8-12 Figure 8-6: Pneumatic Diagram with Zero/Span Valves
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 8-13 8.6 Leak Check Procedure If a leak checker is not available, it is possible t
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 8-14 8.8 Prom Replacement Procedure Preparation: If any setup changes such as RANG
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-1 9 TROUBLESHOOTING AND ADJUSTMENTS NOTE The operations outlined in this chapte
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-2 4. Pay special attention to the slopes, the slopes are the software equivalent
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-3 9.1 Operation Verification - Diagnostic Techniques 9.1.1 Fault Diagnosis with
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 1-1 1 HOW TO USE THIS MANUAL The Model 9110AH has been designed to produce accurat
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-4 Table 9-1: Test Functions Test Function Factory Set-Up Comment RANGE 500 PP
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-5 Table 9-1: Test Functions (Continued) Test Function Factory Set-Up Comment O
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-6 Table 9-1: Test Functions (Continued) Test Function Factory Set-Up Comment N
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-7 Table 9-1: Test Functions (Continued) Test Function Factory Set-Up Comment C
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-8 9.1.2 Fault Diagnosis with WARNING Messages The M9110AH monitors several inter
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-9 Table 9-2: Front Panel Warning Messages Message Description SYSTEM RESET Ana
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-10 Table 9-2: Front Panel Warning Messages (Continued) Message Description BLOC
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-11 Table 9-3: Summary of Diagnostic Modes DIAG Mode Description SIGNAL I/O Giv
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-12 9.1.3.1 Signal I/O Diagnostics Table 9-4: Diagnostic Mode - Signal I/O No.
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-13 Table 9-4: Diagnostic Mode - Signal I/O (Continued) No. Signal Control Des
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 1-2 INTENTIONALLY BLANK
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-14 Table 9-4: Diagnostic Mode - Signal I/O (Continued) No. Signal Control Des
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-15 Table 9-4: Diagnostic Mode - Signal I/O (Continued) No. Signal Control Des
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-16 9.1.3.2 Electric Test This function injects a constant voltage between the pr
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-17 9.1.3.4 Ozone Gen Power This diagnostic manually turns the power off and on t
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-18 Table 9-5: Model 9110AH Variables No. Name Units Default Value Value Range
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-19 9.1.5 Test Channel Analog Output Many of the TEST functions have an analog vo
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-20 Table 9-6: Test Channel Readings (Continued) TEST Channel Minimum* Maximum*
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-21 PMT Calibration Procedure: NOTE In this procedure a range of 100 ppm and a spa
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-22 10. Adjust S2, the HVPS coarse adjustment, on the preamp board to the setting
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-23 Figure 9-1: Span Calibration Voltage
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 2-1 2 GETTING STARTED 2.1 Unpacking CAUTION Your safety and the safety of others
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-24 9.2 Performance Problems When the response from a span check is outside the c
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-25 9.2.2 Flow Check 1. Check TEST function RCEL - this is the absolute pressure
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-26 2. Check instrument electronics. A. Do ELEC TEST procedure in DIAGNOSTIC men
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-27 3. HVPS noisy - see HVPS test procedure. See Section 9.3.8.5. 4. Defective e
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-28 9.2.7 Unstable Zero Common causes are: 1. Leak in pneumatic system. 2. Ligh
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-29 A. A leak that admits gas containing NO. B. Contaminated reaction cell. Remo
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-30 7. Ozone concentration too low: A. Wet air in generator - need to disassembl
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-31 9.3 Subsystem Troubleshooting and Adjustments 9.3.1 Computer, Display, Keybo
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-32 Testing and operation of the CPU RS-232 port is described in Section 9.3.2.
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-33 Figure 9-2: CPU Board Jumper Settings
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 2-2 WARNING Analyzer Exhaust – O3 Scrubber – Pump Pack Danger – Analyzer exhaust
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-34 9.3.1.3 Front Panel Keyboard The keyboard consists of 8 keys and 3 LED's
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-35 Figure 9-3: RS-232 Pin Assignments
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-36 9.3.2.1 RS-232 Port If the RS-232 port is not working, check the following:
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-37 Now set the baud rate of the Analyzer to the speed required by the modem and i
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-38 A. Turn on the instrument. B. Enter the SETUP-MORE-DIAG menu. C. Select the
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-39 9.3.3.1 ADC/DAC Calibration Procedure Due to the stability of modern electron
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-40 10 V output or non-isolated current loop Switch State Comment 3 ON Use for
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-41 1. After the switches and jumpers are set, turn on instrument power and compl
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-42 Figure 9-4: V/F Board Jumper Settings
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-43 9.3.4 Status/Temp Board The Status/Temp Board is a multifunction board that:
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 2-3 Figure 2-1: Removal of Shipping Screws & Check for Correct Power
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-44 The thermocouple amplifier for the converter temperature has a gain adjustment
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-45 9.3.4.4 4-20 mA Current Output There are two types of current loop outputs av
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-46 Table 9-8: Power Supply Module Subassemblies Module Description Linear Power
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-47 Figure 9-5: Power Supply Module Layout
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-48 Figure 9-6: Electrical Block Diagram
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-49 PSM Diagnostic procedures The Linear Power Supply board can be tested by check
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-50 The generator cartridge - transformer assembly is potted to eliminate corona a
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-51 9.3.6.1 The PermaPure Drier The PermaPure drier is constructed of 2 concentri
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-52 Figure 9-7: Ozone Generator Subsystem
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-53 9.3.7 Flow/Pressure Sensor The flow/pressure sensor board consists of 2 press
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 2-4 Figure 2-2: Rear Panel
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-54 To adjust the SAMPLE PRESSURE 1. Go to DIAG mode by pressing SETUP-MORE-DIAG,
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-55 Figure 9-8: Flow/Pressure Sensor
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-56 Figure 9-9: NOx Sensor Module
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-57 Figure 9-10: NOx Sensor Module
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-58 9.3.8 NOx Sensor Module Figure 9-9 and Figure 9-10 show the assembly of the N
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-59 9.3.8.4 PMT Cooler The PMT cooler uses a Peltier cooler supplied with DC curr
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-60 Figure 9-11: PMT Cooler Subsystem
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-61 9.3.8.5 High Voltage P.S. The HVPS is located in the interior of the Sensor M
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-62 Figure 9-12: High Voltage Power Supply
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-63 9.3.9 Z/S Valves The Z/S Valves is an option in the instrument. Before troub
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 2-5 Figure 2-3: Inlet and Exhaust Venting Recommendations
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-64 9.3.10 Pneumatic System The pneumatic system is diagrammed in Figure 8-4, Fi
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 9-65 INTENTIONALLY BLANK
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 2-6 2.3 Initial Operation 1. Turn on the instrument power. 2. The display shoul
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 2-7 7. When the instrument is warmed up, re-check the TEST functions against Table
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 2-8 Step 2 - Calibrate the instrument: Zero/Span Calibration Procedure Step Number
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 2-9 Step 3 - Review Quality of calibration: Calibration Quality Check Procedure Ste
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 2-9 Figure 2-4: Front Panel
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 2-11
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 2-12 Figure 2-5: Assembly Layout
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 2-13 Table 2-1: Final Test and Calibration Values TEST Values Observed Value Units
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 2-14 Table 2-1: Final Test and Calibration Values (Continued) Parameter Observed V
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 2-15 INTENTIONALLY BLANK
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 3-1 3 SPECIFICATIONS, WARRANTY 3.1 Specifications Operating Modes NO/NOx switchin
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F ii SAFETY MESSAGES Your safety and the safety of others is very important. We have
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 3-2 3.2 Warranty WARRANTY POLICY (02024c) Prior to shipment, TELEDYNE equipment is
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 3-3 TERMS AND CONDITIONS All units or components returned to TELEDYNE should be pro
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 3-4 INTENTIONALLY BLANK
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 4-1 4 THE M9110AH NOX ANALYZER 4.1 Principle of Operation The TELEDYNE Model 9110
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 4-2 In the NO/ NOx switching mode the Analyzer samples the gas stream and measures
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 4-3 Figure 4-1: Block Diagram
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 4-4 4.2 Operation Summary 4.2.1 Sensor Module, Reaction Cell, Detector The sensor
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 4-5 4.2.3 Computer Hardware and Software The M9110AH Analyzer is controlled by a m
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 4-6 The M9110AH can operate in any of 3 different sampling modes - NO only, NOx onl
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 4-7 4.2.6 Power Supply Module The Power supply module supplies AC and DC power to
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F iii TABLE OF CONTENTS SAFETY MESSAGES...
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 4-8 NOTE On vacuum vs absolute pressure: Many vacuum gauges read relative to ambie
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 4-9 Figure 4-2: External Pump Pack
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 4-10 4.2.8 Ozone Generator Because of the instability of ozone, it is necessary to
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-1 5 SOFTWARE FEATURES The M9110AH control software has two major operating modes
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-2 Figure 5-1: Sample Menu Tree
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-3 Figure 5-2: Setup Menu Tree
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-4 5.1.1 Sample Menu Table 5-1: M9110AH Sample Menu Structure Level 1 Level 2 L
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-5 5.1.2 Set-Up Menu Table 5-2: M9110AH Setup Menu Level 1 Level 2 Level 3 Lev
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-6 Table 5-2: M9110AH Setup Menu (Continued) Level 1 Level 2 Level 3 Level 4 De
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-7 Table 5-3: M9110AH Setup Menu Level 1 Level 2 Level 3 Level 4 Level 5 Des
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F iv 5.3 SET-UP MODE ...
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-8 Table 5-4: M9110AH Setup Menu Level 1 Level 2 Level 3 Level 4 Level 5 Des
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-9 5.2 Sample Mode 5.2.1 Test Functions NOTE In any of the following TEST functi
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-10 Sample Flow (SAMP FLW) The SAMPLE FLOW test function is computed from the pres
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-11 Reaction Cell Temperature (RCEL TEMP) This is a measurement of the temperature
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-12 Reaction Cell Pressure (RCEL) The pressure in the reaction cell is measured by
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-13 5.2.2 CAL, CALS, CALZ, Calibration Functions The calibration and zero/span ch
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-14 If AutoRange is enabled, there is a different slope and offset for each of the
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-15 After every zero or span calibration, it is very important to check the QUALIT
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-16 5.3.3 Data Acquisition System (DAS) The M9110AH contains a flexible and power
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-17 The function of each of the default Data Channels is described below: CONC: S
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F v 7.7 CALIBRATION REQUIREMENTS FOR AUTORANGE OR REMOTE RANGE...
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-18 5.3.4 Range Menu The instrument operates on any full scale range from 5 to 5,
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-19 To set up Auto Range press SETUP-RNGE-MODE-AUTO, then press ENTR. To set the v
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-20 4. CALIBRATION When the above selections have been made, the instrument now m
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-21 5.3.5 Password Enable If passwords are enabled, one of the values shown in Ta
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-22 5.3.9 Variables Menu (VARS) This menu enables you to change the settings on c
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-23 Table 5-7: M9110AH Operating Modes Mode Description ZERO CAL D Automatic dy
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-24 5.4.2 NOx Only Mode In this mode only NOx is measured. The converter is in th
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-25 5.6 Status Output The status output is an option that reports Analyzer condit
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-26 5.7.1 Setting up the RS-232 Interface The baud rate is set from the front pan
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-27 Security Feature The RS-232 port can be connected to a public telephone line w
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F vi 9.3.3 Voltage/Frequency (V/F) Board ...
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-28 Table 5-10: RS-232 Switching From Terminal Mode to Computer Mode Key Functio
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-29 5.7.2 Command Summary The information contained in the rest of this section c
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-30 Table 5-12: RS-232 Command Summary Commands Definition ? Print help screen
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-31 Table 5-13: RS-232 Command Summary Terminal Mode Editing Keys Definition V M
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-32 Table 5-14: RS-232 Interface Command Types First Character Message Type C C
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-33 5.7.3 TEST Commands and Messages Table 5-15: RS-232 Test Messages Name Mess
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-34 The T command lists TEST messages. Examples of the T command are: T LIST List
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-35 interface. These messages are helpful when trying to track down a problem with
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-36 Whenever the analyzer starts or finishes a calibration, it issues a status rep
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-37 The following is a summary of the Diagnostic commands. Table 5-19: RS-232 Dia
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F vii LIST OF FIGURES FIGURE 2-1: REMOVAL OF SHIPPING SCREWS & CHECK FOR CORRECT
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-38 To print records from a DAS data channel enter: D REPORT “name” RECORDS=nnn CO
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-39 All of the default Data Channels sample more than one parameter, for these cha
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 5-40 Where: name = name of the variable value = current value of variable warning
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 6-1 6 OPTIONAL HARDWARE AND SOFTWARE Optional equipment offered with the M9110AH i
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 6-2 6.2 Zero/Span Valves Option The Zero/Span Valve option consists of a manifold
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 6-3 Table 6-2: Setup Automatic Zero/Span Checking or Calibration Step Action Com
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 6-4 Table 6-3: Action of MODE Field in AutoCal MODE CALIBRATE = ON CALIBRATE = O
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 6-5 6.4 Molybdenum Converter Option The Molybdenum Converter is used in applicatio
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 6-6 INTENTIONALLY BLANK
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 7-1 7 CALIBRATION AND ZERO/SPAN CHECKS There are several ways to check and adjust
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F viii LIST OF TABLES TABLE 2-1: FINAL TEST AND CALIBRATION VALUES...
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 7-2 Table 7-1: Types of Zero/Span Check and Calibration Section Type of Cal or Ch
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 7-3 7.1 Manual Zero/Span Check or Cal With Zero/Span Gas in the Sample Port The ca
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 7-4 Enter the expected NOx and NO span gas concentrations: Table 7-3: Enter Expect
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 7-5 Table 7-4: Manual Span Calibration Procedure - Span Gas Thru Sample Port Step
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 7-6 7.2 Manual Zero/Span Check with Zero/Span Valves Option The Zero/Span valve op
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 7-7 Table 7-6: Manual Span Check Procedure - Z/S Valves Step Number Action Comme
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 7-8 7.4 Use of Zero/Span Valves with Remote Contact Closure The Zero/Span valve op
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 7-9 7.5 NO Only Mode Calibration For details on the operation of this mode, check
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 7-10 Table 7-8: Calibration Requirements for AutoRange or Remote Range No matter w
TELEDYNE Model 9110AH NOX Analyzer Operator Manual, 01620, Rev. F 7-11 The offset value gives information about the background signal level. Check th