Teledyne GFC7000E User Manual Page 203

  • Download
  • Add to my manuals
  • Print
  • Page
    / 265
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 202
Model GFC7000E Instruction Manual TROUBLESHOOTING & REPAIR PROCEDURES
04584 Rev A1 192
Table 11-5: Relay Board Status LED Failure Indications
SIGNAL I/O PARAMETER
LED FUNCTION
ACTIVATED BY VIEW RESULT
DIAGNOSTIC TECHNIQUE
D2
Yellow
WHEEL
HEATER
WHEEL_HEATER WHEEL_TEMP
Voltage displayed should change. If not:
Failed Heater
Faulty Temperature Sensor
Failed AC Relay
Faulty Connectors/Wiring
D3
Yellow
BENCH
HEATER
BENCH_HEATER BENCH_TEMP
Voltage displayed should change. If not:
Failed Heater
Faulty Temperature Sensor
Failed AC Relay
Faulty Connectors/Wiring
D4
Yellow
SPARE N/A N/A
N/A
D5
Green
SAMPLE/CAL
GAS VALVE
OPTION
CAL_VALVE N/A
Sample/Cal Valve should audibly change
states. If not:
Failed Valve
Failed Relay Drive IC on Relay Board
Failed Relay Board
Faulty +12 VDC Supply (PS2)
Faulty Connectors/Wiring
D6
Green
ZERO/SPAN
GAS VALVE
OPTION
SPAN_VALVE N/A
Zero/Span Valve should audibly change
states. If not:
Failed Valve
Failed Relay Drive IC on Relay Board
Failed Relay Board
Faulty +12 VDC Supply (PS2)
Faulty Connectors/Wiring
D7
Green
SHUTOFF
VALVE OPTION
SHUTOFF_VALVE
N/A
Shutoff Valve should audibly change
states. If not:
Failed Valve
Failed Relay Drive IC on Relay Board
Failed Relay Board
Faulty +12 VDC Supply (PS2)
Faulty Connectors/Wiring
D8
Green
IR SOURCE IR_SOURCE CO
2
_MEASURE
Voltage displayed should change. If not:
Failed IR Source
Faulty +12 VDC Supply (PS2)
Failed Relay Board
Failed IR Photo-Detector
Failed Sync/Demod Board
Faulty Connectors/Wiring
11.1.5. Gas Flow Problems
In general, flow problems can be divided into three categories:
1. Flow is too high
2. Flow is greater than zero, but is too low, and/or unstable
3. Flow is zero (no flow)
When troubleshooting flow problems, it is a good idea to first confirm that the actual flow and not
the analyzer’s flow sensor and software are in error, or the flow meter is in error. Use an
Page view 202