OPERATING INSTRUCTIONS
Model 3010MA
Paramagnetic Oxygen Purity
Analyzer
DANGER
Toxic gases and or flammable liquids may be present in this monitoring system.
Personal protective equipment may be required when servicing this instrument.
Hazardous voltages exist on certain components internally which may persist for
a time even after the power is turned off and disconnected.
Only authorized personnel should conduct maintenance and/or servicing. Before
conducting any maintenance or servicing, consult with authorized
supervisor/manager.
P/N M3010MA
05/07/01
ECO# 01-0055
Page view 0
1 2 3 4 5 6 ... 87 88

Summary of Contents

Page 1 - Model 3010MA

Teledyne Analytical Instruments i OPERATING INSTRUCTIONS Model 3010MA Paramagnetic Oxygen Purity Analyzer DANGER Toxic gases and or flammabl

Page 2 - Model 3010MA

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments x Preface Overview The Teledyne Analytical Instruments Model 3010MA Paramagnetic

Page 3 - Safety Messages

Paramagnetic O2 Purity Analyzer Teledyne Analytical Instruments xi • High resolution, accurate readings of oxygen content • Paramagnetic oxygen

Page 4 - CONSEQUENCES

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments xii

Page 5 - Table of Contents

Paramagnetic O2 Purity Analyzer Introduction Teledyne Analytical Instruments 13 OPERATING INSTRUCTIONS Model 3010MA Paramagnetic Oxygen Purity An

Page 6

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 14

Page 7

Paramagnetic O2 Purity Analyzer Introduction Teledyne Analytical Instruments 15 Introduction 1.1 Overview The Teledyne Analytical Instruments M

Page 8

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 16 • Alarms—Set the alarm setpoints and attributes. • Range— Set up the 3 user d

Page 9 - Figures

Paramagnetic O2 Purity Analyzer Introduction Teledyne Analytical Instruments 17 Standby Button: The Standby feature turns off the display and out

Page 10 - Preface

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 18 • Network—For future expansion. Not implemented at this printing. • 50-Pin Eq

Page 11

Paramagnetic O2 Purity Analyzer Operational Theory Teledyne Analytical Instruments 19 Operational Theory 2.1 Introduction The analyzer is compos

Page 12

Model 3010MA Teledyne Analytical Instruments ii Copyright © 2001 TELEDYNE Analytical Instruments All Rights Reserved. No part of this manual may be

Page 13

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 20 Figure 2-1: Block Diagram for Model 3010MA Control Unit Figure 2-2: Block Diag

Page 14

Paramagnetic O2 Purity Analyzer Operational Theory Teledyne Analytical Instruments 21 2.2 Electronics and Signal Processing The Model 3010MAParama

Page 15 - Introduction

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 22 When an oxygen level above 95% is present, the paramagnetic sensor in the Analy

Page 16

Paramagnetic O2 Purity Analyzer Operational Theory Teledyne Analytical Instruments 23 Figure 2-4: Block Diagram of the Model 3010MA Electronics

Page 17

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 24

Page 18

Paramagnetic O2 Purity Analyzer Installation Teledyne Analytical Instruments 25 Installation Installation of the Model 3010MA Analyzer includes: 1

Page 19 - Operational Theory

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 26 Figure 3-1: Rack Mountable Control and Analysis Units

Page 20

Paramagnetic O2 Purity Analyzer Installation Teledyne Analytical Instruments 27 All operator controls are mounted on the control panel which is sh

Page 21

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 28 3.3 Rear Panel Connections—Control Unit Figure 3-4 shows the Model 3010MA rear

Page 22 - 2.5 Temperature Control

Paramagnetic O2 Purity Analyzer Installation Teledyne Analytical Instruments 29 3.3.2 Electrical Connections—Control Unit CAUTION: FOR SAFE CONN

Page 23

Paramagnetic O2 Purity Analyzer Teledyne Analytical Instruments iii Safety Messages Your safety and the safety of others is very important. We have

Page 24

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 30 are given where each function is described in the paragraphs below. At the end

Page 25 - Installation

Paramagnetic O2 Purity Analyzer Installation Teledyne Analytical Instruments 31 contact connections. The contact connections are shown in Table 3-

Page 26

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 32 36 System Alarm, normally closed contact 20 System Alarm, moving contact

Page 27

Paramagnetic O2 Purity Analyzer Installation Teledyne Analytical Instruments 33 When the contact is OPEN, the analyzer is analyzing, the Remote Ca

Page 28

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 34 39 Range 2 ID Contact 19 Range 3 ID Contact 18 Range 3 ID Contact 34 No

Page 29 - USE INSTALLATION

Paramagnetic O2 Purity Analyzer Installation Teledyne Analytical Instruments 35 Figure 3-7: FET Series Resistance Below is a table of the complet

Page 30

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 36 12 (-) Remote Span 37 System Alarm, NO Contact 13 38 Range 1 ID Contact 14

Page 31

Paramagnetic O2 Purity Analyzer Installation Teledyne Analytical Instruments 37 Table 3-6: Commands via RS-232 Input Command Description as<en

Page 32

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 38 The Analysis Unit is installed adjacent to the Control Unit usually on the same

Page 33

Paramagnetic O2 Purity Analyzer Installation Teledyne Analytical Instruments 39 3.5.1 Gas Connections—Analysis Unit When making gas connections t

Page 34

Model 3010MA Teledyne Analytical Instruments iv POSSIBLY ACCOMPANIED WITH HAZARDOUS CONSEQUENCES. This manual provides information designed to guid

Page 35

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 40 4. Check that sample flow rate is between 300 and 1000 SCCM. CAUTION: FLOW R

Page 36

Paramagnetic O2 Purity Analyzer Operation Teledyne Analytical Instruments 41 Operation 4.1 Introduction Once the analyzer has been installed, it

Page 37

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 42 The Enter button is used to accept any new entries on the VFD screen. The Escap

Page 38

Paramagnetic O2 Purity Analyzer Operation Teledyne Analytical Instruments 43 • Setup Auto-Cal • Assign Password • Initiate Self-Test • Check softw

Page 39 - 3.6 Testing the System

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 44 After a password is assigned, the operator must log out to activate it. Until

Page 40 - SENSOR!

Paramagnetic O2 Purity Analyzer Operation Teledyne Analytical Instruments 45 To setup an Auto–Cal cycle: Choose System from the Function buttons.

Page 41 - Operation

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 46 4.3.3.1 ENTERING THE PASSWORD To install a new password or change a previously

Page 42

Paramagnetic O2 Purity Analyzer Operation Teledyne Analytical Instruments 47 4.3.3.2 INSTALLING OR CHANGING THE PASSWORD If you want to install a

Page 43 - 4.3 The System Function

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 48 Wait a moment for the entry screen. You will be given clearance to proceed. Us

Page 44 - 4.3.1 Setting the Display

Paramagnetic O2 Purity Analyzer Operation Teledyne Analytical Instruments 49 4.3.5 System Self-Diagnostic Test The Model 3010MA has a built-in se

Page 45 - 4.3.3 Password Protection

Paramagnetic O2 Purity Analyzer Teledyne Analytical Instruments v Table of Contents Safety Messages ...

Page 46 - NTERING THE PASSWORD

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 50 4.3.6 Version Screen Move the < > arrow key to More and press Enter. Wit

Page 47

Paramagnetic O2 Purity Analyzer Operation Teledyne Analytical Instruments 51 If you are using password protection, you will need to enter your pas

Page 48 - 4.3.4 Logout

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 52 The beginning span value is shown in the upper left corner of the display. As t

Page 49

Paramagnetic O2 Purity Analyzer Operation Teledyne Analytical Instruments 53 4.5 The Alarms Function The Model 3010MA is equipped with 2 fully ad

Page 50 - 4.3.7 Display Sensor Output

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 54 4. Are either of the alarms to be defeated? The defeat alarm mode is incorpo

Page 51 - 4.4.1 Span Cal

Paramagnetic O2 Purity Analyzer Operation Teledyne Analytical Instruments 55 To reset a latched alarm, go to Dft– and then press either ∆ two time

Page 52 - ANUAL MODE SPANNING

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 56 Use the ∆∇ arrow keys to enter the upper value of the range (all ranges end at

Page 53 - 4.5 The Alarms Function

Paramagnetic O2 Purity Analyzer Operation Teledyne Analytical Instruments 57 4.7 The Analyze Function Normally, all of the functions automaticall

Page 54

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 58 The analog output signal has a voltage which depends on the oxygen concentratio

Page 55 - 4.6 The Range Function

Paramagnetic O2 Purity Analyzer Maintenance Teledyne Analytical Instruments 59 Maintenance 5.1 Routine Maintenance Aside from normal cleaning an

Page 56 - 4.6.3 Fixed Range Analysis

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments vi 4.1 Introduction 41 4.2 Using the Data Entry and Function Buttons 41 4.3 T

Page 57 - 4.8 Signal Output

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 60 2. To change between American and European fuses, remove the single retaining

Page 58

Paramagnetic O2 Purity Analyzer Maintenance Teledyne Analytical Instruments 61 Preamp 0 OK 1 Zero Too High 2 Amplifier output doesn’t match test

Page 59 - Maintenance

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 62 The Control Unit Contains: • 2 Power Supplies • Motherboard (with microproces

Page 60 - European Fuses

Paramagnetic O2 Purity Analyzer Maintenance Teledyne Analytical Instruments 63 To access the Universal Power Supply and electronic PCB’s, the rea

Page 61

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments 64

Page 62

Paramagnetic O2 Purity Analyzer Analysis Unit Teledyne Analytical Instruments 65 OPERATING INSTRUCTIONS Model 3010MA Paramagnetic Oxygen Purity An

Page 63 - 5.5 Cleaning

Part 2 Analysis Unit Model 3010MA Teledyne Analytical Instruments 66

Page 64

Paramagnetic O2 Purity Analyzer Analysis Unit Teledyne Analytical Instruments 67 Analysis Unit 6.1 Overview The Model 3010MA Paramagnetic Oxygen

Page 65

Part 2 Analysis Unit Model 3010MA Teledyne Analytical Instruments 68 6.2 Paramagnetic Sensor The paramagnetic sensor is sensitive to pressure and

Page 66

Paramagnetic O2 Purity Analyzer Analysis Unit Teledyne Analytical Instruments 69 Figure 6-1: Side view of Paramagnetic Sensor Figure 6-2: Front

Page 67 - Analysis Unit

Paramagnetic O2 Purity Analyzer Teledyne Analytical Instruments vii 6.2.2 Cross Interference 71 6.3 Oven and Temperature Controller 74 6.4

Page 68 - 6.2 Paramagnetic Sensor

Part 2 Analysis Unit Model 3010MA Teledyne Analytical Instruments 70 The Paramagnetic sensor and heating elements are contained within the Analysis

Page 69

Paramagnetic O2 Purity Analyzer Analysis Unit Teledyne Analytical Instruments 71 4 Not used 5 Preamplifier output ~0- 1 Vdc 6 Not used 7 24

Page 70

Part 2 Analysis Unit Model 3010MA Teledyne Analytical Instruments 72 Figure 6-4: 15-Pin Sensor Cable and Pin Identification

Page 71 - 6.2.2 Cross Interference

Paramagnetic O2 Purity Analyzer Analysis Unit Teledyne Analytical Instruments 73 Example: Cross sensitivity calculation for a gas mixture: The gas

Page 72

Part 2 Analysis Unit Model 3010MA Teledyne Analytical Instruments 74 Cyclo hexane C6H12 -1.56 Nitric oxide NO +40.00 Ethane C2H6 -0.43 n-Octane C8

Page 73

Paramagnetic O2 Purity Analyzer Analysis Unit Teledyne Analytical Instruments 75 6.4 Sample System The paramagnetic sensor used in the Model 3010

Page 74

Part 2 Analysis Unit Model 3010MA Teledyne Analytical Instruments 76 The sample system delivers gases to the sensor from the gas inlet located on

Page 75 - 6.4 Sample System

Paramagnetic O2 Purity Analyzer Analysis Unit Teledyne Analytical Instruments 77 side of the sensor for the pressure compensation. This line is fe

Page 76

Part 2 Analysis Unit Model 3010MA Teledyne Analytical Instruments 78 If remote gas selection is desired, the user must supply 3 two-way 24 V dc sol

Page 77

Paramagnetic O2 Purity Analyzer Analysis Unit Teledyne Analytical Instruments 79 As mentioned in Chapter 2, the paramagnetic sensor used in this i

Page 78

Part 1 Control Unit Model 3010MA Teledyne Analytical Instruments viii Tables Table 3-1: Analog Output Connections ...

Page 79

Part 2 Analysis Unit Model 3010MA Teledyne Analytical Instruments 80 arrow keys to blink More, then press Enter. Use the < > arrow keys again

Page 80

Paramagnetic O2 Purity Analyzer Appendix Teledyne Analytical Instruments 81 Appendix A-1 Model 3010MA Specifications Sensor: Paramagnetic Ran

Page 81 - Appendix

Appendix Model 3010MA Teledyne Analytical Instruments 82 Approvals: CE mark Area Classification: General Purpose

Page 82 - Approvals: CE mark

Paramagnetic O2 Purity Analyzer Appendix Teledyne Analytical Instruments 83 A-2 Recommended 2-Year Spare Parts List Qty Part Number Description 1

Page 83

Appendix Model 3010MA Teledyne Analytical Instruments 84 A-3 Drawing List D-71698 Outline Drawing B-71699 Piping Diagram A-4 19-inch Relay Rack P

Page 84 - A-3 Drawing List

Paramagnetic O2 Purity Analyzer Index Teledyne Analytical Instruments 85 Index AC power connection. See power connection access door, 17, 27, 62 a

Page 85

Index Model 3010MA Teledyne Analytical Instruments 86 default values, 41 defeat alarm mode, 54 digital communications device. See RS-232 port digit

Page 86

Paramagnetic O2 Purity Analyzer Index Teledyne Analytical Instruments 87 paramagnetic sensor, 19, 62, 68, 69 password, 43, 45 PC board main, 15 pi

Page 87

Index Model 3010MA Teledyne Analytical Instruments 88 Table of Contents, v Tables, list of, viii temperature controller, 15, 19, 22, 68, 74 tempera

Page 88

Paramagnetic O2 Purity Analyzer Teledyne Analytical Instruments ix Figures Figure 1-1: Model 3010MA Control Unit Front Panel ......

Comments to this Manuals

No comments