Teledyne 7300a - NDIR analyzer User Manual

Browse online or download User Manual for Equipment Teledyne 7300a - NDIR analyzer. Teledyne 7300a - NDIR analyzer User Manual [fr]

  • Download
  • Add to my manuals
  • Print
  • Page
    / 85
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
i
Infrared Gas Analyzer
Teledyne Analytical Instruments
OPERATING INSTRUCTIONS
Model 7300A
Infrared Gas Analyzer
P/N M7300A
11/06/13
HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM.
PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM.
HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST FOR A
TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED.
ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND/OR SERVICING. BEFORE
CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED SUPERVISOR/MANAGER.
DANGER
7
3
0
0
S
E
R
I
E
S
I
N
F
R
A
-
R
E
D
A
N
A
L
Y
Z
E
R
Page view 0
1 2 3 4 5 6 ... 84 85

Summary of Contents

Page 1 - Model 7300A

iInfrared Gas AnalyzerTeledyne Analytical InstrumentsOPERATING INSTRUCTIONS Model 7300AInfrared Gas AnalyzerP/N M7300A11/06/13HIGHLY TOXIC AND OR F

Page 2 - Model 7300A

1-21 Introduction Model 7300ATeledyne Analytical Instruments1.3 Main Features of the AnalyzerThe Model 7300A Infrared Gas Analyzer is sophisticated

Page 3

1-3Infrared Gas Analyzer Introduction 1Teledyne Analytical Instrumentsbeing measured. There is a direct correlation between absorption and theconcen

Page 4

1-41 Introduction Model 7300ATeledyne Analytical Instruments

Page 5 - Model 7300A

Infrared Gas Analyzer Installation 2 2-1Teledyne Analytical Instruments2.0 InstallationInstallation of the Model 7300A Infrared Gas Analyzer include

Page 6

2 Installation Model 7300A2-2Teledyne Analytical Instrumentsplaced as close to the sample point as possible and bolted to its supportingsurface. Whe

Page 7

Infrared Gas Analyzer Installation 2 2-3Teledyne Analytical Instruments2.2.4 Pipe ConnectionsRefer to Appendix Piping Drawings for information about

Page 8

2 Installation Model 7300A2-4Teledyne Analytical Instruments2.3.1 Primary Input PowerThe power cord receptacle and fuse block are located in the same

Page 9 - 1.0 Introduction

Infrared Gas Analyzer Installation 2 2-5Teledyne Analytical Instruments4–20 mA dc % Range: Current rises linearly with concentration, from 4mA at 0 c

Page 10 - 1.4 General

2 Installation Model 7300A2-6Teledyne Analytical Instruments6 0.6 13.67 0.7 15.28 0.8 16.89 0.9 18.410 1.0 20.0To provide an indication of the range,

Page 11 - 1.5 NDIR Analyzer

Infrared Gas Analyzer Installation 2 2-7Teledyne Analytical InstrumentsSystem Alarm: Actuates when DC power supplied to circuits isunacceptable in on

Page 12 - 1 Introduction Model 7300A

iiModel 7300ATeledyne Analytical InstrumentsCopyright © 2013 Teledyne Analytical InstrumentsAll Rights Reserved. No part of this manual may be reprod

Page 13 - 2.0 Installation

2 Installation Model 7300A2-8Teledyne Analytical InstrumentsTable 2-6: Remote Calibration ConnectionsPin Function9 + Remote Zero11 – Remote Zero10 +

Page 14

Infrared Gas Analyzer Installation 2 2-9Teledyne Analytical Instruments2.3.3.4 Range ID RelaysFour dedicated Range ID relay contacts. For any single

Page 15

2 Installation Model 7300A2-10Teledyne Analytical InstrumentsThe voltage from these outputs is nominally 0 V for the OFF and15 V dc for the ON condit

Page 16

Infrared Gas Analyzer Installation 2 2-11Teledyne Analytical InstrumentsTable 2-8: Commands via RS-232 InputCommand Descriptionas<enter> Immedi

Page 17

2 Installation Model 7300A2-12Teledyne Analytical Instruments2.5 Testing the SystemBefore plugging the instrument into the power source:• Check the i

Page 18

Infrared Gas Analyzer Installation 2 2-13Teledyne Analytical Instruments1. Introduce zero fluid and set zero as referred in section 4.4.1NOTE: Whe

Page 19

2 Installation Model 7300A2-14Teledyne Analytical Instruments

Page 20

Infrared Gas Analyzer Start-up and Operation 3-1Teledyne Analytical Instruments3.0 Start-up and Operation3.1 PreliminaryBefore applying power to the s

Page 21

3 Start-up and Theory of Operation Model 7300A3-2Teledyne Analytical Instrumentsand sample flow between 0.1 to 0.6 SCFH (50-250 cc/min) .Zero standa

Page 22

Infrared Gas Analyzer Start-up and Operation 3-3Teledyne Analytical Instrumentsare not used during automatic sampling.3.3 Theory of operation3.3.1 Ge

Page 23 - 2.4 Gas Requirements

iiiInfrared Gas AnalyzerTeledyne Analytical InstrumentsTechnician Record SheetTechnician Record SheetTechnician Record SheetTechnician Record SheetTec

Page 24 - 2.6 Calibration

3 Start-up and Theory of Operation Model 7300A3-4Teledyne Analytical InstrumentsThese lamps are typically rated for 20,000 hours continuous operation

Page 25

Infrared Gas Analyzer Start-up and Operation 3-5Teledyne Analytical Instruments3.4 Circuit DescriptionThe Teledyne Analytical Instruments IR bench is

Page 26 - 2 Installation Model 7300A

3 Start-up and Theory of Operation Model 7300A3-6Teledyne Analytical Instruments3.5 Digital Signal Processing & ElectronicsThe Model 7300A uses a

Page 27 - 3.0 Start-up and Operation

Infrared Gas Analyzer Start-up and Operation 3-7Teledyne Analytical InstrumentsThe CPU also provides appropriate control signals to the Displays,Alarm

Page 28 - 3.2.2 Operational Calibration

3 Start-up and Theory of Operation Model 7300A3-8Teledyne Analytical InstrumentsPiece-wise approximation is the method used to linearize the signal,

Page 29 - 3.3.2 Analyzer

Infrared Gas Analyzer Start-up and Operation 3-9Teledyne Analytical Instruments3.7 AUTOMATIC FUNCTION (Optional)The events talking place during a z

Page 30 - Processing

3 Start-up and Theory of Operation Model 7300A3-10Teledyne Analytical Instruments

Page 31 - IR Bench Block Diagram

Infrared Gas Analyzer Operation 44-1Teledyne Analytical InstrumentsOperation4.1 IntroductionAlthough the Model 7300A is usually programmed to your a

Page 32

4 Operation Model 7300A4-2 Teledyne Analytical Instruments4.2 Using the ControlsTo get the proper response from these controls, press the desired k

Page 33 - Figure I

Infrared Gas Analyzer Operation 44-3Teledyne Analytical InstrumentsFigure 4-1: Hierarchy of Functions and SubfunctionsSpan/ZeroTimingSpan/ZeroOff/O

Page 34 - Linearizer Input

ivModel 7300ATeledyne Analytical Instruments

Page 35

4 Operation Model 7300A4-4 Teledyne Analytical Instruments4.2.1.2 Setup ModeThe MAIN MENU consists of 14 functions you can use to customize andche

Page 36

Infrared Gas Analyzer Operation 44-5Teledyne Analytical Instruments15 OFFSET: This function helps set a non-zero offset to the zerocalibration. It

Page 37 - Operation

4 Operation Model 7300A4-6 Teledyne Analytical InstrumentsIf you do not wish to continue a function, you can abort the session byescaping . Escapin

Page 38 - 4.2 Using the Controls

Infrared Gas Analyzer Operation 44-7Teledyne Analytical Instruments4.3.3 Password ProtectionBefore a unique password is assigned, the system assigns

Page 39 - ALOGORITHM

4 Operation Model 7300A4-8 Teledyne Analytical InstrumentsPress Escape to move on, or proceed as in Changing the Password,below.4.3.3.2 Installing

Page 40

Infrared Gas Analyzer Operation 44-9Teledyne Analytical InstrumentsWhen you have finished typing the new password, press Enter. A verifica-tion scre

Page 41

4 Operation Model 7300A4-10 Teledyne Analytical InstrumentsThe self diagnostics are run automatically by the analyzer whenever theinstrument is tur

Page 42

Infrared Gas Analyzer Operation 44-11Teledyne Analytical InstrumentsDpt INPUT OUTPUT Ø Ø.ØØ Ø.ØØThe leftmost digit (under Dpt) is the numb

Page 43

4 Operation Model 7300A4-12 Teledyne Analytical InstrumentsHard_offset_C=3015 (This is the raw DAC count of the Coarse zeroadjustment. It should re

Page 44

Infrared Gas Analyzer Operation 44-13Teledyne Analytical Instruments6 1607 3208 6409 128010 2360The response time listed above can and will change d

Page 45

iTable of ContentsTeledyne Analytical InstrumentsOPERATING INSTRUCTIONSModel 7300AInfrared Gas AnalyzerTable of ContentsGeneral Purpose

Page 46

4 Operation Model 7300A4-14 Teledyne Analytical InstrumentsHow the offset value is selected:To find out what the offset value should be, the intend

Page 47 - mode : AUTO

Infrared Gas Analyzer Operation 44-15Teledyne Analytical InstrumentsAdjst output to 4ma<ENT> to OK: -7Use the Up or Down key to adjust

Page 48

4 Operation Model 7300A4-16 Teledyne Analytical Instruments4.4 The Zero and Span Functions(1) The Model 7300A can have as many as three analysis ra

Page 49

Infrared Gas Analyzer Operation 44-17Teledyne Analytical InstrumentsIf you want to calibrate ALL of the ranges at once(multiple application analyzer

Page 50 - 4.3.11 CAL-OUT Function

4 Operation Model 7300A4-18 Teledyne Analytical Instruments4.4.1.2 Manual Mode ZeroingScroll to Zero and enter the Zero function. The screen that a

Page 51

Infrared Gas Analyzer Operation 44-19Teledyne Analytical Instrumentsb. Check for leaks downstream from the sensor, where contamina-tion may be leaki

Page 52 - Functions

4 Operation Model 7300A4-20 Teledyne Analytical InstrumentsENTER will move the blinking field to units (%/ppm). Use UP/DOWN keyto select the units,

Page 53

Infrared Gas Analyzer Operation 44-21Teledyne Analytical Instrumentsis pressed, the Span reading changes to the correct value. The instrument thenau

Page 54

4 Operation Model 7300A4-22 Teledyne Analytical Instrumentsalarm status will terminate when process conditions revert to non-alarm conditions.4. Ar

Page 55 - <ENT> To begin span

Infrared Gas Analyzer Operation 44-23Teledyne Analytical Instrumentsthe number. Holding down the key speeds up the incrementing ordecrementing.• Aft

Page 56

iiModel 7300A Infrared Gas AnalyzerTeledyne Analytical InstrumentsTable of Contents1 Introduction1.1 Overview ...

Page 57 - Function

4 Operation Model 7300A4-24 Teledyne Analytical InstrumentsNote: If all three ranges are currently defined for different applica-tion gases, then t

Page 58

Infrared Gas Analyzer Operation 44-25Teledyne Analytical Instrumentsrange readings are accurate UNLESS the application uses linearization over these

Page 59 - Select Function

4 Operation Model 7300A4-26 Teledyne Analytical Instruments• Ranges that overlap• Ranges whose limits are entirely within the span of an adjoiningr

Page 60

Infrared Gas Analyzer Operation 44-27Teledyne Analytical InstrumentsIf the concentration detected is over range, the first line of the display blink

Page 61 - Select range

4 Operation Model 7300A4-28 Teledyne Analytical InstrumentsIn the autoranging mode, the microprocessor automatically responds toconcentration chang

Page 62 - 4.7 The

Infrared Gas Analyzer Operation 44-29Teledyne Analytical InstrumentsNote: The ranges must be increasing from low to high, for example,if Range 1 is

Page 63 - Application

4 Operation Model 7300A4-30 Teledyne Analytical InstrumentsAlgorithm setup: VERIFY SETUPUP/DOWN to select and Enter VERIFY to check whether the

Page 64 - FR:Ø TO:1Ø %

Infrared Gas Analyzer Operation 44-31Teledyne Analytical InstrumentsFrom the MAIN MENU Screen—1. Use UP/DOWN to select ALGORITHM , and Enter.2. Sele

Page 65 - MAIN MENU

4 Operation Model 7300A4-32 Teledyne Analytical InstrumentsNote: The span gas use to span the analyzer must be >90% of therange being analyzed.B

Page 66

Infrared Gas Analyzer Operation 44-33Teledyne Analytical Instrumentstivity of the binary gas, the analyzer would normally be set up so that the 100%

Page 67 - 0 Ø.ØØ Ø.ØØ

iiiTable of ContentsTeledyne Analytical Instruments3 Start-up and Theory of Operation ...3-13.1 Preliminary...

Page 68 - 4.9 Special Function Setup

4 Operation Model 7300A4-34 Teledyne Analytical InstrumentsSelect STANDBY to restart the system.4.9.3 Gain PresetNOTE: This function will apply o

Page 69 - PLICATION sceen

Infrared Gas Analyzer Maintenance 5 5-1Teledyne Analytical Instruments5.0 MaintenanceAside from normal cleaning and checking for leaks at the gas

Page 70

5 Maintenance Model 7300A5-2Teledyne Analytical Instruments4. Reassemble Housing as shown in Figure 5-1. American Fuses European FusesFigur

Page 71 - 5.0 Maintenance

Infrared Gas Analyzer Maintenance 5 5-3Teledyne Analytical Instruments5.3 NDIR Analyzer Measurement CellThe Infrared Analyzer contains an auto ze

Page 72 - Figure 5-2: Installing Fuses

5 Maintenance Model 7300A5-4Teledyne Analytical InstrumentsTable 5-1: Self Test Failure CodesPower0OK1 5 V Failure2 15 V Failures3 Both FailedAnalo

Page 73

Infrared Gas Analyzer Maintenance 5 5-5Teledyne Analytical InstrumentsFigure 5-3: Cell AssemblyWARNING: HAZARDOUS VOLTAGES EXIST ON CERTAINCOMPONE

Page 74 - 5.5 Major Internal Components

5 Maintenance Model 7300A5-6Teledyne Analytical Instruments5.6 Troubleshooting5.7 GeneralThis section contains information on the assembly and th

Page 75 - Figure 5-3: Cell Assembly

Infrared Gas Analyzer Maintenance 5 5-7Teledyne Analytical Instruments5.8 TROUBLE SHOOTING CHART SYMPTOM CORRECTIVE ACTIONA. Unit does no

Page 76 - 5.7 General

5 Maintenance Model 7300A5-8Teledyne Analytical InstrumentsC. Meter reads Negative (a) ZERO control may bemisadjusted. (with zero gas flowing Ad

Page 77 - 5.8 TROUBLE SHOOTING CHART

Infrared Gas Analyzer Maintenance 5 5-9Teledyne Analytical Instruments(b) Chopper hitting housing.(c) Noisy detector.(d) AC pickup at printed circ

Page 78

ivModel 7300A Infrared Gas AnalyzerTeledyne Analytical Instruments4.4.2 Span Cal... 4-194

Page 79

5 Maintenance Model 7300A5-10Teledyne Analytical Instruments

Page 80 - 5 Maintenance Model 7300A

A-1Infrared Gas Analyzer AppendixTeledyne Analytical InstrumentsAppendixA-1 Specifications7300A Digital Control Portion:Ranges: Three Programmable Ran

Page 81 - A-1 Specifications

A-2Appendix Model 7300ATeledyne Analytical Instruments7300 Specifications Chemical and petrochemical processes Gas Analysis• Combustion and flue ga

Page 82 - Drawing List

A-3Infrared Gas Analyzer AppendixTeledyne Analytical InstrumentsRecommended 2-Year Spare Parts ListNOTE: Part list may or not be part of your system,

Page 83

A-4Appendix Model 7300ATeledyne Analytical InstrumentsATTACHMENT 7300AQuote "Exceptions" and "GAS PHASE Conditions" for thisapp

Page 84

A-5Infrared Gas Analyzer AppendixTeledyne Analytical Instruments10 If condensables occur, such as moisture, acid gases, solvents, etc.,sample system m

Page 85

1-1Infrared Gas Analyzer Introduction 1Teledyne Analytical Instruments1.0 Introduction1.1 OverviewThe Teledyne Analytical Instruments Model 7300A A

Comments to this Manuals

No comments