instructionmanualTeledyne Analytical InstrumentsA Business Unit of Teledyne Electronic TechnologiesModel 8800PTRACE MOISTURE ANALYZER
page 85.2 Operating StateRefer to Appendix A for a flow diagram of the operating state. The 8800P is powered ON by pressing the ‘Power’ button until a
page 9back to atmospheric settings. A long press of the pressure correct button changes the unit back to the Viewing Dewpoint Mode. Pressing the ‘Mode
page 10For example serial number 12345 will be shown as:3;61 x10Pressing the ‘Mode’ button changes the unit to the Viewing Dewpoint Mode.5.3 Set-
page 116. OptionsThe instrument can be optionally equipped with an output/power supply board, to allow for the following capabilities:1. RS-232C.2. Cu
page 12If the current loop is not being used, it is safe to connect the DB9 connector directly to a PC; pins 1&9 will not be dammaged or couse dam
page 136.3 Operating from an external Power SupplyThe external power connector is a 2mm power jack located on the rear of the unit. This input can acc
page 14 7. Automatic CalibrationThe instrument is calibrated at the factory with the shipped sensor, and does not need to be re-calibrated prior to us
page 158. Changing desiccant cartridge and/or battery.In order to change desiccant cartridge and/or battery, remove the two thumb screws on the rear c
page 169. Special messages, warnings and error indicationsLegend:denotes a beeping accompaniment to the message.denotes alternately flashing message
page 17Appendix A: Flow diagram of User Interface in the Operating State {{{{{{{{unitsvaluesvaluesSensorPositionSensorReadingunitsNew calibration v
8800P Series Trace Moisture Analyzer Instruction ManualiWhen calling your representative for technical support, please have your serial numbers availa
page 18Appendix B: Flow Diagram of User Interface in the Set-Up State Legend: / separates alternatively displayed messages Buttons
page 19Appendix C: Relationship of Instrument Reading and 4-20mA output when lbs of H20/million standard cft or ppmv engineering units are selected.In
page 20Appendix D: Uncertainty in LBS and PPMV calculations Uncertainty of LBS-H2O calculation due to +/-3˚C measurment accuracyCalculated LBS-H2O in
page 21Appendix E: Battery Life
page 11. IntroductionThe 8800P is a portable, hand-held, microprocessor controlled, battery operated dew-point meter with built in dry-storage for the
page 22. Precautions• Warranty limitation: The 8800P’s sample cell/piston assembly should not be disas-sembled except by factory trained personnel. Th
page 33. Principle of OperationDewpoint sensors, like all hygroscopic materials, adsorb much faster than they desorb. Therefore, an accurate measureme
page 4A note about dry-storageThe dry-storage consist of a stainless steel container filled with desiccant. The elec-tropolished stainless steel senso
page 54. Operating the 8800P4.1. Sample Hook-up4.1.1. FittingsThe 8800P can be hooked up to the sample gas using a variety of fittings, depending on t
page 6instrument, while still blocking the sample outlet port.• Observe the change in dewpoint indication. The reading will be stable within about 3 m
page 75.0 The User InterfaceThe user interface consists of a custom LCD display, an audio indicator, 5 push buttons and a sensor actuator.5.1 Display