Teledyne Analytical Instruments OPERATING INSTRUCTIONS FOR Model 3300ZB Trace Oxygen Analyzer DANGER Toxic gases and or flammable liquids ma
Trace Oxygen Analyzer Introduction Teledyne Analytical Instruments 1 Introduction 1.1 Overview The Teledyne Analytical Instruments (TAI) Model
Introduction Model 3300ZB Teledyne Analytical Instruments 2 Compact and rugged control unit, wall-mounting NEMA-4 rated enclosure. External
Trace Oxygen Analyzer Introduction Teledyne Analytical Instruments 3 Set LO Alarm Set the concentration BELOW which an alarm activates.
Introduction Model 3300ZB Teledyne Analytical Instruments 4 Figure 1-2: Rear Panel Power Connection AC Power: 100–240 VAC, 50/60 Hz. The
Trace Oxygen Analyzer Operational Theory Teledyne Analytical Instruments 5 Operational Theory 2.1 Introduction The analyzer is composed of three
Operational Theory Model 3300ZB Teledyne Analytical Instruments 6 The sensor is designed for long term stability with an operating life of 5 ye
Trace Oxygen Analyzer Operational Theory Teledyne Analytical Instruments 7 2.3.2 Signal Processing Figure 2-1 is a block diagram of the signal p
Operational Theory Model 3300ZB Teledyne Analytical Instruments 8 RS-232 (optional) output. The same digital information is also sent to a 8-bit
Trace Oxygen Analyzer Installation Teledyne Analytical Instruments 9 Installation Installation of the analyzer includes: 1. Unpacking the system
Model 3300ZB Teledyne Analytical Instruments ii Copyright © 2014 Teledyne Instruments/ Analytical Instruments All Rights Reserved. No part of thi
Installation Model 3300ZB Teledyne Analytical Instruments 10 3.2 Location and Mounting 3.2.1 Control Unit Installation The 3300ZB control unit is
Trace Oxygen Analyzer Installation Teledyne Analytical Instruments 11 Figure 3-1: Rear Panel Electrical Connectors Analog Outputs: There are th
Installation Model 3300ZB Teledyne Analytical Instruments 12 The specific descriptions for each type of alarm are as follows: HI Alarm Configured
Trace Oxygen Analyzer Installation Teledyne Analytical Instruments 13 The RS-232 protocol allows some flexibility in implementation in the choice
Installation Model 3300ZB Teledyne Analytical Instruments 14 Figure 3-3: Piping Diagram 3.5 Installation Checklist Before connecting the instru
Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 15 Operation 4.1 Introduction Once the analyzer has been mounted, the gas lines c
Operation Model 3300ZB Teledyne Analytical Instruments 16 4.2 Using the Function and Data Entry Buttons When one of the function buttons is being
Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 17 4.3.1 HI Range Setting the HI Range full-scale value defines the LEAST sensiti
Operation Model 3300ZB Teledyne Analytical Instruments 18 1. Press the SET LO ALARM function button once. 2. Within 5 seconds press either the
Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 19 3. As the display is blinking, once the readout appears stable, press either
Trace Oxygen Analyzer Teledyne Analytical Instruments iii Safety Messages Your safety and the safety of others is very important. We have provid
Operation Model 3300ZB Teledyne Analytical Instruments 20 (This page intentionally left blank)
Trace Oxygen Analyzer Maintenance Teledyne Analytical Instruments 21 Maintenance Aside from normal cleaning and checking for leaks at the gas con
Maintenance Model 3300ZB Teledyne Analytical Instruments 22 To open the spare fuse compartment, push on one end until it slides out. 4. Remove th
Trace Oxygen Analyzer Maintenance Teledyne Analytical Instruments 23 Figure 5-2: ZrO2 Sensor and Cell Block Figure 5-3 (A&B): Cell Removal/
Maintenance Model 3300ZB Teledyne Analytical Instruments 24 (This page intentionally left blank)
Trace Oxygen Analyzer Appendix Teledyne Analytical Instruments 25 Appendix A.1 Specifications Ranges: Two user selectable ranges can be set bet
Appendix Model 3300ZB Teledyne Analytical Instruments 26 A.2 Spare Parts List 1 C-64586A PC Board, Power Supply 1 C-76491C PCB Board, Main
Trace Oxygen Analyzer Appendix Teledyne Analytical Instruments 27 A.3 Drawing List D80672 Outline Diagram A69265 Piping Diagram
Model 3300ZB Teledyne Analytical Instruments iv CAUTION: THE ANALYZER SHOULD ONLY BE USED FOR THE PURPOSE AND IN THE MANNER DESCRIBED IN THIS MA
Trace Oxygen Analyzer Teledyne Analytical Instruments v Table of Contents Safety Messages ...
Model 3300ZB Teledyne Analytical Instruments vi Operation ... 1
Trace Oxygen Analyzer Teledyne Analytical Instruments vii List of Figures Figure 1-1: Front Panel ...
Model 3300ZB Teledyne Analytical Instruments viii DANGER COMBUSTIBLE GAS USAGE WARNING This is a general purpose instrument designed for use
Trace Oxygen Analyzer Teledyne Analytical Instruments ix (This page intentionally left blank)
Comments to this Manuals