Teledyne GFC 7001E User Manual Page 217

  • Download
  • Add to my manuals
  • Print
  • Page
    / 321
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 216
EPA Calibration Protocol Model GFC7001E Carbon Dioxide Analyzer
Teledyne Analytical Instruments 217
10. EPA CALIBRATION PROTOCOL
10.1. CALIBRATION REQUIREMENTS
If the GFC 7001E is to be used for EPA SLAMS monitoring, it must be calibrated in accordance with the
instructions in this section.
The USEPA strongly recommends that you obtain a copy of the publication Quality Assurance Handbook for Air
Pollution Measurement Systems Volume 2: Part 1, Ambient (abbreviated, Q.A. Handbook Volume II). This
manual can be purchased from:
USEPA Order Number: EPA454R98004; or NTIS Order Number: PB99 129876.
National Technical Information Service (phone 800-553-6847) or Center for Environmental Research
Information or the U.S. Government Printing Office at http://www.gpo.gov. The Handbook can also be
located on line by searching for the title at http://www.epa.gov.
Special attention should be paid to Section 2.6 of that which covers CO analyzers of this type.
Specific regulations regarding the use and operation of ambient CO analyzers can be found in
Reference 1 at the end of this Section.
A bibliography and references relating to CO monitoring are listed in Section 10.6.
10.1.1. CALIBRATION OF EQUIPMENT - GENERAL GUIDELINES
In general, calibration is the process of adjusting the gain and offset of the GFC 7001E against some recognized
standard. In this section the term dynamic calibration is used to express a multipoint check against known
standards and involves introducing gas samples of known concentration into the instrument in order to adjust the
instrument to a predetermined sensitivity and to produce a calibration relationship.
This relationship is derived from the instrumental response to successive samples of different known
concentrations. As a minimum, three reference points and a zero point are recommended to define this
relationship.
All monitoring instrument systems are subject to some drift and variation in internal parameters and cannot be
expected to maintain accurate calibration over long periods of time. Therefore, it is necessary to dynamically
check the calibration relationship on a predetermined schedule. Zero and span checks
must be used to
document that the data remains within control limits. These checks are also used in data reduction and
validation.
Calibration can be done by either diluting high concentration CO standards with zero air or using individual tanks
of known concentration. Details of documentation, forms and procedures should be maintained with each
analyzer and also in a central backup file as described in Section 2.6.2 of the Quality Assurance Handbook.
The reliability and usefulness of all data derived from any analyzer depends primarily upon its state of calibration.
To ensure accurate measurements of the CO levels:
1. The analyzer must be calibrated at the time of installation and recalibrated as necessary.
2. In order to insure that high quality, accurate measurement information is obtained at all times, the
analyzer must be calibrated prior to use.
3. Calibrations should be carried out at the field-monitoring site.
4. The analyzer should be in operation for at least several hours (preferably overnight) before calibration
so that it is fully warmed up and its operation has stabilized.
Page view 216
1 2 ... 212 213 214 215 216 217 218 219 220 221 222 ... 320 321

Comments to this Manuals

No comments