Getting Started Model GFC7001E Carbon Dioxide Analyzer
Teledyne Analytical Instruments 47
3.5.1. STARTUP
After the electrical and pneumatic connections are made, turn on the instrument. The pump and exhaust fan
should start immediately. The display should immediately display a single, horizontal dash in the upper left
corner of the display. This will last approximately 30 seconds while the CPU loads the operating system.
Once the CPU has completed this activity it will begin loading the analyzer firmware and configuration data.
During this process, astring of messages will appear on the analyzer’s front panel display.
System waits 3 seconds then
automatically begins its
initialization routine.
.
System is checking the validity and
functionality of the Terminus
memory and firmware.
If at this point,
appears, contact Teledyne
Instruments customer service
.
The instrument is loading
configuration and calibration data
from the flash chip.
The startup process may hesitate at this point if:
· The Ethernet option is installed;
· DHCP mode is turned on and;
· The instrument is not connected to a
functioning network.
The instrument is loading the
system firmware.
The analyzer should automatically switch to Sample Mode after completing the boot-up sequence and start
monitoring CO gas.