Troubleshooting & Repair Model GFC7001E Carbon Dioxide Analyzer
Teledyne Analytical Instruments 304
13.6.4. DISK-ON-MODULE REPLACEMENT PROCEDURE
Replacing the Disk-on-Module (DOM) will cause loss of all DAS data; it also may cause loss of some instrument
configuration parameters unless the replacement DOM carries the exact same firmware version. Whenever
changing the version of installed software, the memory must be reset. Failure to ensure that memory is reset can
cause the analyzer to malfunction, and invalidate measurements. After the memory is reset, the A/D converter
must be re-calibrated, and all information collected in Step 1 below must be re-entered before the instrument will
function correctly. Also, zero and span calibration should be performed.
1. Document all analyzer parameters that may have been changed, such as range, auto-cal, analog
output, serial port and other settings before replacing the DOM
2. Turn off power to the instrument, fold down the rear panel by loosening the mounting screws.
3. When looking at the electronic circuits from the back of the analyzer, locate the Disk-on-Module in the
right most socket of the CPU board.
4. The DOM should carry a label with firmware revision, date and initials of the programmer.
5. Remove the nylon fastener that mounts the DOM over the CPU board, and lift the DOM off the CPU. Do
not bend the connector pins.
6. Install the new Disk-on-Module, making sure the notch at the end of the chip matches the notch in the
socket.
7. It may be necessary to straighten the pins somewhat to fit them into the socket. Press the DOM all the
way in and reinsert the offset clip.
8. Close the rear panel and turn on power to the machine.
9. If the replacement DOM carries a firmware revision, re-enter all of the setup information.
13.7. TECHNICAL ASSISTANCE
If this manual and its troubleshooting / repair sections do not solve your problems, technical assistance may be
obtained from:
TELEDYNE ELECTRONIC TECHNOLOGIES
Analytical Instruments
16830 Chestnut Street
City of Industry, CA 91748
Telephone: (626) 934-1500
Fax: (626) 961-2538
Web: www.teledyne-ai.com
Before you contact Teledyne Customer service, fill out the problem report form in Appendix C, which is also
available online for electronic submission at http://www.teledyne-api.com/forms/
.