Teledyne Analytical Instruments i
OPERATING INSTRUCTIONS
Model 3010MA
Paramagnetic Oxygen Purity
Analyzer
DANGER
Toxic gases and or flammable liquids may be present in this monitoring system.
Personal protective equipment may be required when servicing this instrument.
Hazardous voltages exist on certain components internally which may persist for
a time even after the power is turned off and disconnected.
Only authorized personnel should conduct maintenance and/or servicing. Before
conducting any maintenance or servicing, consult with authorized
supervisor/manager.
P/N M3010MA
7/01/10
Page view 0
1 2 3 4 5 6 ... 87 88

Summary of Contents

Page 1 - Model 3010MA

Teledyne Analytical Instruments i OPERATING INSTRUCTIONS Model 3010MA Paramagnetic Oxygen Purity Analyzer DANGERToxic gases and or flammab

Page 2

Model 3010MA x Teledyne Analytical Instruments

Page 3 - Safety Messages

Paramagnetic O2 Purity Analyzer Introduction Teledyne Analytical Instruments 11 Preface Overview The Teledyne Analytical Instruments Model 3010

Page 4 - CONSEQUENCES

Part 1 Control Unit Model 3010MA 12 Teledyne Analytical Instruments Main Features of the Analyzer The Model 3010MA series Paramagnetic Oxygen Analy

Page 5 - Table of Contents

Paramagnetic O2 Purity Analyzer Introduction Teledyne Analytical Instruments 13 OPERATING INSTRUCTIONS Model 3010MA Paramagnetic Oxygen Purity

Page 6

Part 1 Control Unit Model 3010MA 14 Teledyne Analytical Instruments

Page 7

Paramagnetic O2 Purity Analyzer Introduction Teledyne Analytical Instruments 15 Introduction 1.1 Overview The Teledyne Analytical Instruments

Page 8

Part 1 Control Unit Model 3010MA 16 Teledyne Analytical Instruments • Alarms—Set the alarm setpoints and attributes. • Range— Set up the 3 user d

Page 9 - Figures

Paramagnetic O2 Purity Analyzer Introduction Teledyne Analytical Instruments 17 Standby Button: The Standby feature turns off the display and ou

Page 10 - Model 3010MA

Part 1 Control Unit Model 3010MA 18 Teledyne Analytical Instruments • Network—For future expansion. Not implemented at this printing.  50-Pin Eq

Page 11 - Preface

Paramagnetic O2 Purity Analyzer Operational Theory Teledyne Analytical Instruments 19 Operational Theory 2.1 Introduction The analyzer is compo

Page 12

Model 3010MA ii Teledyne Analytical Instruments Copyright © 2010 TELEDYNE Analytical Instruments All Rights Reserved. No part of this manual may be

Page 13

Part 1 Control Unit Model 3010MA 20 Teledyne Analytical Instruments Figure 2-1: Block Diagram for Model 3010MA Control Unit Figure 2-2: Block Diag

Page 14

Paramagnetic O2 Purity Analyzer Operational Theory Teledyne Analytical Instruments 21 2.2 Electronics and Signal Processing The Model 3010MAParam

Page 15 - Introduction

Part 1 Control Unit Model 3010MA 22 Teledyne Analytical Instruments When an oxygen level is present, the paramagnetic sensor in the Analysis Unit g

Page 16

Paramagnetic O2 Purity Analyzer Operational Theory Teledyne Analytical Instruments 23 Figure 2-4: Block Diagram of the Model 3010MA Electronics

Page 17

Part 1 Control Unit Model 3010MA 24 Teledyne Analytical Instruments

Page 18

Paramagnetic O2 Purity Analyzer Installation Teledyne Analytical Instruments 25 Installation Installation of the Model 3010MA Analyzer includes:

Page 19 - Operational Theory

Part 1 Control Unit Model 3010MA 26 Teledyne Analytical Instruments Figure 3-1: Rack Mountable Control and Analysis Units

Page 20

Paramagnetic O2 Purity Analyzer Installation Teledyne Analytical Instruments 27 All operator controls are mounted on the control panel which is s

Page 21

Part 1 Control Unit Model 3010MA 28 Teledyne Analytical Instruments 3.3 Rear Panel Connections—Control Unit Figure 3-4 shows the Model 3010MA rear

Page 22 - 2.5 Temperature Control

Paramagnetic O2 Purity Analyzer Installation Teledyne Analytical Instruments 29 3.3.2 Electrical Connections—Control Unit CAUTION: FOR SAFE CON

Page 23

Paramagnetic O2 Purity Analyzer Teledyne Analytical Instruments iii Safety Messages Your safety and the safety of others is very important. We have

Page 24

Part 1 Control Unit Model 3010MA 30 Teledyne Analytical Instruments are given where each function is described in the paragraphs below. At the end

Page 25 - Installation

Paramagnetic O2 Purity Analyzer Installation Teledyne Analytical Instruments 31 contact connections. The contact connections are shown in Table 3

Page 26

Part 1 Control Unit Model 3010MA 32 Teledyne Analytical Instruments 36 System Alarm, normally closed contact 20 System Alarm, moving contact

Page 27

Paramagnetic O2 Purity Analyzer Installation Teledyne Analytical Instruments 33 When the contact is OPEN, the analyzer is analyzing, the Remote C

Page 28

Part 1 Control Unit Model 3010MA 34 Teledyne Analytical Instruments 39 Range 2 ID Contact 19 Range 3 ID Contact 18 Range 3 ID Contact 34

Page 29 - FUSE INSTALLATION

Paramagnetic O2 Purity Analyzer Installation Teledyne Analytical Instruments 35 Figure 3-7: FET Series Resistance Below is a table of the comple

Page 30

Part 1 Control Unit Model 3010MA 36 Teledyne Analytical Instruments 12 (-) Remote Span 37 System Alarm, NO Contact 13 38 Range 1 ID Contact

Page 31

Paramagnetic O2 Purity Analyzer Installation Teledyne Analytical Instruments 37 Table 3-6: Commands via RS-232 Input Command Description as<e

Page 32

Part 1 Control Unit Model 3010MA 38 Teledyne Analytical Instruments The Analysis Unit is installed adjacent to the Control Unit usually on the same

Page 33

Paramagnetic O2 Purity Analyzer Installation Teledyne Analytical Instruments 39 3.5.1 Gas Connections—Analysis Unit When making gas connections

Page 34

Model 3010MA iv Teledyne Analytical Instruments POSSIBLY ACCOMPANIED WITH HAZARDOUS CONSEQUENCES. This manual provides information designed to guid

Page 35

Part 1 Control Unit Model 3010MA 40 Teledyne Analytical Instruments 4. Check that sample flow rate is between 300 and 1000 SCCM. CAUTION: FLOW R

Page 36 - RS-232 PORT

Paramagnetic O2 Purity Analyzer Operation Teledyne Analytical Instruments 41 Operation 4.1 Introduction Once the analyzer has been installed, it

Page 37

Part 1 Control Unit Model 3010MA 42 Teledyne Analytical Instruments The Enter button is used to accept any new entries on the VFD screen. The Escap

Page 38

Paramagnetic O2 Purity Analyzer Operation Teledyne Analytical Instruments 43 • Setup Auto-Cal • Assign Password • Initiate Self-Test • Check sof

Page 39 - 3.6 Testing the System

Part 1 Control Unit Model 3010MA 44 Teledyne Analytical Instruments After a password is assigned, the operator must log out to activate it. Until

Page 40 - SENSOR!

Paramagnetic O2 Purity Analyzer Operation Teledyne Analytical Instruments 45 To setup an Auto–Cal cycle: Choose System from the Function buttons.

Page 41 - Operation

Part 1 Control Unit Model 3010MA 46 Teledyne Analytical Instruments 4.3.3.1 ENTERING THE PASSWORD To install a new password or change a previously

Page 42

Paramagnetic O2 Purity Analyzer Operation Teledyne Analytical Instruments 47 4.3.3.2 INSTALLING OR CHANGING THE PASSWORD If you want to install

Page 43 - 4.3 The System Function

Part 1 Control Unit Model 3010MA 48 Teledyne Analytical Instruments AAAAARetypePWDToVerifyWait a moment for the entry screen. You wi

Page 44 - 4.3.1 Setting the Display

Paramagnetic O2 Purity Analyzer Operation Teledyne Analytical Instruments 49 4.3.5 System Self-Diagnostic Test The Model 3010MA has a built-in s

Page 45 - 4.3.3 Password Protection

Paramagnetic O2 Purity Analyzer Teledyne Analytical Instruments v Table of Contents Safety Messages...

Page 46 - ENTERING THE PASSWORD

Part 1 Control Unit Model 3010MA 50 Teledyne Analytical Instruments 4.3.6 Version Screen Move the < > arrow key to More and press Enter. Wit

Page 47

Paramagnetic O2 Purity Analyzer Operation Teledyne Analytical Instruments 51 If you are using password protection, you will need to enter your pa

Page 48 - 4.3.4 Logout

Part 1 Control Unit Model 3010MA 52 Teledyne Analytical Instruments The beginning span value is shown in the upper left corner of the display. As t

Page 49

Paramagnetic O2 Purity Analyzer Operation Teledyne Analytical Instruments 53 4.5 The Alarms Function The Model 3010MA is equipped with 2 fully a

Page 50 - 4.3.7 Display Sensor Output

Part 1 Control Unit Model 3010MA 54 Teledyne Analytical Instruments In latching mode, once the alarm or alarms trigger, they will remain in the ala

Page 51 - 4.4.1 Span Cal

Paramagnetic O2 Purity Analyzer Operation Teledyne Analytical Instruments 55 To set the other parameters use the < > arrow keys to move the

Page 52 - MANUAL MODE SPANNING

Part 1 Control Unit Model 3010MA 56 Teledyne Analytical Instruments However, the digital readout and the RS-232 output of the concentration are una

Page 53 - 4.5 The Alarms Function

Paramagnetic O2 Purity Analyzer Operation Teledyne Analytical Instruments 57 or L—0.00M—95.00H—98.00Mode—FX/HIPress Escape to re-enter the

Page 54 - Dft—NFs—NLtch—N

Part 1 Control Unit Model 3010MA 58 Teledyne Analytical Instruments 98.6 0.3 8.8 98.8 0.4 10.4 99.0 0.5 12 99.2 0.6 13.6 99.4 0.7 15.2 99.6 0.

Page 55 - 4.6 The Range Function

Paramagnetic O2 Purity Analyzer Maintenance Teledyne Analytical Instruments 59 Maintenance 5.1 Routine Maintenance Aside from normal cleaning an

Page 56 - 4.6.3 Fixed Range Analysis

Model 3010MA vi Teledyne Analytical Instruments 4.1 Introduction 414.2 Using the Data Entry and Function Buttons 414.3 The System Function

Page 57 - 4.8 Signal Output

Part 1 Control Unit Model 3010MA 60 Teledyne Analytical Instruments 2. To change between American and European fuses, remove the single retaining

Page 58

Paramagnetic O2 Purity Analyzer Maintenance Teledyne Analytical Instruments 61 Preamp 0 OK 1 Zero Too High 2 Amplifier output doesn’t match tes

Page 59 - Maintenance

Part 1 Control Unit Model 3010MA 62 Teledyne Analytical Instruments The Control Unit Contains:  2 Power Supplies  Motherboard (with microproces

Page 60 - European Fuses

Paramagnetic O2 Purity Analyzer Maintenance Teledyne Analytical Instruments 63 To access the Universal Power Supply and electronic PCB’s, the rea

Page 61

Part 1 Control Unit Model 3010MA 64 Teledyne Analytical Instruments

Page 62

Paramagnetic O2 Purity Analyzer Teledyne Analytical Instruments 65 OPERATING INSTRUCTIONS Model 3010MA Paramagnetic Oxygen Purity Analyzer Part

Page 63 - 5.5 Cleaning

Part 2 Analysis Unit Model 3010MA 66 Teledyne Analytical Instruments

Page 64

Paramagnetic O2 Purity Analyzer Analysis Unit Teledyne Analytical Instruments 67 Analysis Unit 6.1 Overview The Model 3010MA Paramagnetic Oxyge

Page 65

Part 2 Analysis Unit Model 3010MA 68 Teledyne Analytical Instruments 6.2 Paramagnetic Sensor The paramagnetic sensor is sensitive to pressure and

Page 66

Paramagnetic O2 Purity Analyzer Analysis Unit Teledyne Analytical Instruments 69 Figure 6-1: Side view of Paramagnetic Sensor Figure 6-2: Fron

Page 67 - Analysis Unit

Paramagnetic O2 Purity Analyzer Teledyne Analytical Instruments vii 6.2.2 Cross Interference 716.3 Oven and Temperature Controller 746.4

Page 68 - 6.2 Paramagnetic Sensor

Part 2 Analysis Unit Model 3010MA 70 Teledyne Analytical Instruments The Paramagnetic sensor and heating elements are contained within the Analysis

Page 69

Paramagnetic O2 Purity Analyzer Analysis Unit Teledyne Analytical Instruments 71 4 Not used 5 Preamplifier output ~0- 1VDC 6 Not used 7 2

Page 70

Part 2 Analysis Unit Model 3010MA 72 Teledyne Analytical Instruments Figure 6-4: 15-Pin Sensor Cable and Pin Identification

Page 71 - 6.2.2 Cross Interference

Paramagnetic O2 Purity Analyzer Analysis Unit Teledyne Analytical Instruments 73 Example: Cross sensitivity calculation for a gas mixture: The ga

Page 72

Part 2 Analysis Unit Model 3010MA 74 Teledyne Analytical Instruments Cyclo hexane C6H12 -1.56 Nitric oxide NO +40.00 Ethane C2H6 -0.43 n-Octane

Page 73 - 137.00.137--0%

Paramagnetic O2 Purity Analyzer Analysis Unit Teledyne Analytical Instruments 75 6.4 Sample System The paramagnetic sensor used in the Model 301

Page 74

Part 2 Analysis Unit Model 3010MA 76 Teledyne Analytical Instruments The sample system delivers gases to the sensor from the gas inlet located on

Page 75 - 6.4 Sample System

Paramagnetic O2 Purity Analyzer Analysis Unit Teledyne Analytical Instruments 77 side of the sensor for the pressure compensation. This line is f

Page 76 - Note: There is no flow

Part 2 Analysis Unit Model 3010MA 78 Teledyne Analytical Instruments If remote gas selection is desired, the user must supply 3 two-way 24VDC solen

Page 77

Paramagnetic O2 Purity Analyzer Analysis Unit Teledyne Analytical Instruments 79 As mentioned in Chapter 2, the paramagnetic sensor used in this

Page 78

Model 3010MA viii Teledyne Analytical Instruments Tables Table 3-1: Analog Output Connections ... 30 Table 3

Page 79

Part 2 Analysis Unit Model 3010MA 80 Teledyne Analytical Instruments arrow keys to blink More, The voltage output of the cell is displayed on the s

Page 80

Paramagnetic O2 Purity Analyzer Appendix Teledyne Analytical Instruments 81 Appendix A-1 Model 3010MA Specifications Sensor: Paramagnetic Ra

Page 81 - Appendix

Part 2 Analysis Unit Model 3010MA 82 Teledyne Analytical Instruments Approvals: CE mark Area Classification: General Purpose

Page 82 - General Purpose

Paramagnetic O2 Purity Analyzer Appendix Teledyne Analytical Instruments 83 A-2 Recommended 2-Year Spare Parts List Qty Part Number Description

Page 83 -  20 mm, Slow Blow

Part 2 Analysis Unit Model 3010MA 84 Teledyne Analytical Instruments A-3 Drawing List D-73635 Outline Drawing B-73695 Piping Diagram

Page 84 - A-3 Drawing List

Paramagnetic O2 Purity Analyzer Appendix Teledyne Analytical Instruments 85 A.4 Programmable Pressure Compensation Adjustment The standard model

Page 85 - Adjustment

Part 2 Analysis Unit Model 3010MA 86 Teledyne Analytical Instruments 2. Zero (with N2 gas) the analyzer with no restriction on vent (needle valve

Page 86 - 0.993112210More

Paramagnetic O2 Purity Analyzer Appendix Teledyne Analytical Instruments 87 Example 1: With pressure compensation factor set to 0.00 and the ven

Page 87

Part 2 Analysis Unit Model 3010MA 88 Teledyne Analytical Instruments to the pressure compensation factor and change the value by 0.01, then repeat

Page 88

Paramagnetic O2 Purity Analyzer Teledyne Analytical Instruments ix Figures Figure 1-1: Model 3010MA Control Unit Front Panel ......