Page 1 - Model 6200T

Teledyne Analytical Instruments Operation Manual for Model 6200T UV Fluorescence Total Sulfur Analyzer TELEDYNE ELECTRONIC TECHNOLOGIES Analyt

Page 2

Table of Contents Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments x 4.1.2. Warning Messages ...

Page 3

Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 100 5.4.3. Reporting Range Modes: Single, Dual, Auto Ranges The 620

Page 4

6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 101 5.4.3.1. Single Range mode (SNGL) The default range mode for the anal

Page 5 - ABOUT THIS MANUAL

Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 102 5.4.3.2. Dual Range Mode (DUAL) Selecting Dual Range mode allow

Page 6

6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 103 5.4.3.3. Auto Range Mode (AUTO) In AUTO range mode, the analyzer auto

Page 7 - SAFETY MESSAGES

Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 104 milligrams per cubic meter (mg/m3, MGM). Changing units affects

Page 8 - CONSIGNES DE SÉCURITÉ

6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 105 5.4.5. Dilution Ratio (Option) The dilution ratio is a software optio

Page 9 - TABLE OF CONTENTS

Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 106 IMPORTANT IMPACT ON READINGS OR DATA Once the above settings ha

Page 10

6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 107 To enable or disable passwords, press: SAMPLE ENTER SE

Page 11

Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 108 If the password feature is enabled, then when entering either C

Page 12

6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 109 5.6. SETUP – CLK: Setting the Internal Time-of-Day Clock The 6200T ha

Page 13 - ...321

6200T Total Sulfur Analyzer Table of Contents Teledyne Analytical Instruments xi 7.2.5. Sample Period and Report Period ...

Page 14

Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 110 SAMPLE RANGE = 500.000 PPB SO2 =XXX.X <

Page 15 - LIST OF FIGURES

6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 111 5.7. SETUP – COMM: Communications Ports This section introduces the c

Page 16

Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 112 SETUP X.X COMMUNICATIONS MENU ID INET COM1

Page 17

6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 113 5.8. SETUP – VARS: Variables Setup and Definition Through the SETUP&g

Page 18 - LIST OF TABLES

Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 114 To access and navigate the VARS menu, use the following button

Page 19

6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 115 5.9. SETUP – DIAG: Diagnostics Functions The SETUP>MORE>DIAG me

Page 20

Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 116 To access the DIAG functions press the following buttons:

Page 21 - GENERAL INFORMATION

6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 117 5.9.1. Signal I/O The signal I/O diagnostic mode allows a user to rev

Page 22

Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 118 5.9.2. Analog Output Step Test Analog Output is used as a step

Page 23 - 1.2. Features

6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 119 5.9.3. Analog I/O Configuration Table 6-8 lists the analog I/O functi

Page 24 - 1.4. Options

Table of Contents Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments xii 11. INSTRUMENT MAINTENANCE ...

Page 25 - General Safety Hazard

Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 120 The following DC current output limits apply to the current loo

Page 26

6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 121 5.9.3.1. Analog Output Signal Type and Range Span Selection To select

Page 27 - 1.5. Configurations

Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 122 5.9.3.2. Analog Output Calibration Mode Analog output calibrati

Page 28

6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 123 To automatically calibrate a single analog channel from the DIAG Menu

Page 29

Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 124 To select manual output calibration for a particular channel, a

Page 30

6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 125 5.9.3.3. Manual Analog Output Calibration and Voltage Adjustment For

Page 31

Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 126 To make these manual adjustments, the AOUT auto-calibration fea

Page 32

6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 127 5.9.3.4. Analog Output Offset Adjustment Some analog signal recorders

Page 33

Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 128 5.9.3.5. Current Loop Output Adjustment A current loop option i

Page 34

6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 129 To adjust the zero and span values of the current outputs, activate t

Page 35 - < 0.5 ppb/24 hours

6200T Total Sulfur Analyzer Table of Contents Teledyne Analytical Instruments xiii 12.7.2. Sensor Module Repair & Cleaning ...

Page 36

Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 130 5.9.3.6. AIN Calibration This is the sub-menu to conduct the an

Page 37

6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 131 To adjust settings for the Analog Inputs option parameters press:

Page 38 - 2.3. CE Mark Compliance

Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 132 5.9.4. Optic Test The optic test function tests the response of

Page 39 - 3. GETTING STARTED

6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 133 5.9.5. Electrical Test The electrical test function creates a current

Page 40 - 3.2. Instrument Layout

Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 134 5.9.6. Lamp Calibration An important factor in accurately deter

Page 41

6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 135 5.9.7. Pressure Calibration A sensor at the exit of the sample chambe

Page 42

Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 136 5.9.8. Flow Calibration The flow calibration allows the user to

Page 43

6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 137 5.9.9. Test Channel Output When activated, output channel A3 can be u

Page 44 - 3.2.2. Rear Panel

Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 138 Once a TEST function is selected, the instrument begins to rep

Page 45 - (Call factory for details)

6200T Total Sulfur Analyzer Communications Setup Teledyne Analytical Instruments 139 6. COMMUNICATIONS SETUP AND OPERATION This instrument rear

Page 46

Table of Contents Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments xiv 14.4.1. General Rules ...

Page 47

Communications Setup Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 140 6.2.1. Communication Modes Either of the analyzer’s s

Page 48 - 3.3. Connections and Setup

6200T Total Sulfur Analyzer Communications Setup Teledyne Analytical Instruments 141 To turn on or off the communication modes for either COM1

Page 49 - 3.3.1.1. Connecting Power

Communications Setup Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 142 6.2.2. COMM Port Baud Rate To select the baud rate of

Page 50 - 2 A3 A4

6200T Total Sulfur Analyzer Communications Setup Teledyne Analytical Instruments 143 6.2.3. COMM Port Testing The serial ports can be tested for

Page 51

Communications Setup Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 144 analyzers, monitors, calibrators, etc.) into one comm

Page 52

6200T Total Sulfur Analyzer Communications Setup Teledyne Analytical Instruments 145 Table 6-3: LAN/Internet Default Configuration Properties PR

Page 53

Communications Setup Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 146 Internet Configuration Button Functions BUT

Page 54

6200T Total Sulfur Analyzer Communications Setup Teledyne Analytical Instruments 147 6.5.2. Configuring Ethernet Communication Using Dynamic Hos

Page 55

Communications Setup Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 148 change this name (particularly if you have more than

Page 56

6200T Total Sulfur Analyzer Communications Setup Teledyne Analytical Instruments 149 1. Go to SETUP>MORE>COMM>COM2 menu. 2. Press the

Page 57

6200T Total Sulfur Analyzer Table of Contents Teledyne Analytical Instruments xv LIST OF FIGURES Figure 1-1:  6200T Basic Configuration...

Page 58

Communications Setup Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 150 Example Read/Write Definition window: Example Connec

Page 59

6200T Total Sulfur Analyzer Communications Setup Teledyne Analytical Instruments 151 6.6.2. HESSEN The Hessen protocol is a Multidrop protocol,

Page 60

Communications Setup Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 152 6.6.2.2. Activating Hessen Protocol The first step

Page 61 - RS-232 port

6200T Total Sulfur Analyzer Communications Setup Teledyne Analytical Instruments 153 6.6.2.3. Selecting a Hessen Protocol Type Currently there a

Page 62 - 3.3.2. Pneumatic Connections

Communications Setup Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 154 6.6.2.4. Setting the Hessen Protocol Response Mode Te

Page 63

6200T Total Sulfur Analyzer Communications Setup Teledyne Analytical Instruments 155 IMPORTANT IMPACT ON READINGS OR DATA It is possible to as

Page 64 - Calibrator

Communications Setup Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 156 To assign or reset the status flag bit assignments, a

Page 65

6200T Total Sulfur Analyzer DAS and AICOM Teledyne Analytical Instruments 157 7. DATA ACQUISITION SYSTEM (DAS) AND AICOM The 6200T analyzer cont

Page 66

DAS and AICOM Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 158 7.1. DAS Structure The DAS is designed around the feature of

Page 67 - REACTION

6200T Total Sulfur Analyzer DAS and AICOM Teledyne Analytical Instruments 159 7.1.2. DAS Parameters Data parameters are types of data that may b

Page 68 - Chassis

Table of Contents Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments xvi Figure 5-22: DIAG – Analog Output – Auto Cal or Manual

Page 69

DAS and AICOM Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 160  WARNINGS: Some data may be useful when stored if one of s

Page 70

6200T Total Sulfur Analyzer DAS and AICOM Teledyne Analytical Instruments 161 The Channel Properties, Triggering Events and Data Parameters/Func

Page 71 - Sensor Option

DAS and AICOM Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 162 7.2.1. Viewing DAS Data and Settings DAS data and settings c

Page 72

6200T Total Sulfur Analyzer DAS and AICOM Teledyne Analytical Instruments 163 7.2.2. Editing DAS Data Channels Although DAS configuration is mos

Page 73 - Calibration Gases

DAS and AICOM Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 164 When editing the data channels, the top line of the display

Page 74 - 3.4.1. Startup

6200T Total Sulfur Analyzer DAS and AICOM Teledyne Analytical Instruments 165 7.2.3. Trigger Events To edit the list of data parameters associa

Page 75

DAS and AICOM Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 166 7.2.4. Editing DAS Parameters Data channels can be edited in

Page 76

6200T Total Sulfur Analyzer DAS and AICOM Teledyne Analytical Instruments 167 To configure the parameters for a specific data parameter, follow

Page 77 - 3.4.2. Warning Messages

DAS and AICOM Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 168 SAMPLE PERIOD: Determines how often DAS temporarily records

Page 78 - MTS501 Converter Messages

6200T Total Sulfur Analyzer DAS and AICOM Teledyne Analytical Instruments 169 To define the REPORT PERIOD, follow the instruction shown in Figur

Page 79 - 3.4.3. Functional Checks

6200T Total Sulfur Analyzer Table of Contents Teledyne Analytical Instruments xvii Figure 9-15: Activate O2 Sensor Stability Function ...

Page 80

DAS and AICOM Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 170 Edit Data Channel Menu SETUP X.X 0) CONC: A

Page 81 - 3.4.4. Initial Calibration

6200T Total Sulfur Analyzer DAS and AICOM Teledyne Analytical Instruments 171 7.2.7. RS-232 Report Function The M DAS can automatically report d

Page 82

DAS and AICOM Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 172 To disable a data channel, follow the instruction shown in F

Page 83 - Dilution Ratio Setup

6200T Total Sulfur Analyzer DAS and AICOM Teledyne Analytical Instruments 173 7.2.11. HOLDOFF Feature The DAS HOLDOFF feature allows to prevent

Page 84 - Set the expected SO

DAS and AICOM Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 174  Check on system parameters for trouble-shooting and quali

Page 85

6200T Total Sulfur Analyzer DAS and AICOM Teledyne Analytical Instruments 175 7.4. Remote DAS Configuration via AICOM Editing channels, paramete

Page 86

DAS and AICOM Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 176 Figure 7-15: DAS Configuration Through a Terminal Emulat

Page 87 - OPERATING INSTRUCTIONS

6200T Total Sulfur Analyzer Remote Operation Teledyne Analytical Instruments 177 8. REMOTE OPERATION OF THE ANALYZER This section provides infor

Page 88

Remote Operation Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 178 Table 8-1: Status Output Pin Assignments CONNECTOR PIN S

Page 89 - MODE EXPLANATION

6200T Total Sulfur Analyzer Remote Operation Teledyne Analytical Instruments 179 CONTROL IN A B C D E F U +

Page 90 - 4.1. Sample Mode

Table of Contents Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments xviii LIST OF TABLES Table 1-1: Analyzer Options ...

Page 91

Remote Operation Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 180 8.2.2. Help Commands in Terminal Mode Table 8-3: Terminal

Page 92

6200T Total Sulfur Analyzer Remote Operation Teledyne Analytical Instruments 181 8.2.4. Data Types Data types consist of integers, hexadecimal i

Page 93 - 4.1.2. Warning Messages

Remote Operation Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 182 DDD:HH:MM is the time stamp, the date and time when the

Page 94

6200T Total Sulfur Analyzer Remote Operation Teledyne Analytical Instruments 183 To change this setting, access the COMMUNICATIONS Menu (refer t

Page 95 - 4.3. Setup Mode

Remote Operation Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 184 To initialize the modem, access the COMMUNICATIONS Menu (

Page 96 - 4.3.2. Primary Setup Menu

6200T Total Sulfur Analyzer Remote Operation Teledyne Analytical Instruments 185  LOGON FAILED - Password not given or incorrect  LOGOFF SUC

Page 97 - 5. SETUP MENU

Remote Operation Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 186 This page intentionally left blank.

Page 98 - EXAMPLE:

6200T Total Sulfur Analyzer Calibration Procedures Teledyne Analytical Instruments 187 9. CALIBRATION PROCEDURES This section describes the cali

Page 99

Calibration Procedures Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 188 standard charcoal scrubber that supplies zero air f

Page 100

6200T Total Sulfur Analyzer Calibration Procedures Teledyne Analytical Instruments 189 9.1.1.3. Calibration Gas Standards and Traceability All e

Page 101

6200T Total Sulfur Analyzer Table of Contents Teledyne Analytical Instruments xix Table 10-4: Definition of Level 1 and Level 2 Zero and Span C

Page 102 - IMPORTANT

Calibration Procedures Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 190 OR SAMPLEEXHAUSTMODEL 701Zero Air Generator VENT

Page 103 - 5.4.4. Range Units

6200T Total Sulfur Analyzer Calibration Procedures Teledyne Analytical Instruments 191 This sequence causes the analyzer to prompt for the expe

Page 104

Calibration Procedures Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 192 M-P CAL SO2 STB=X.XXX PPB SO2 =

Page 105

6200T Total Sulfur Analyzer Calibration Procedures Teledyne Analytical Instruments 193 IMPORTANT IMPACT ON READINGS OR DATA If the ZERO or SPAN

Page 106

Calibration Procedures Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 194 9.4. Manual Calibration with Zero/Span Valves Zero

Page 107

6200T Total Sulfur Analyzer Calibration Procedures Teledyne Analytical Instruments 195 STEP TWO: Set the expected SO2 span gas value: This sequ

Page 108

Calibration Procedures Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 196 Step Three: Perform the calibration or calibration

Page 109 - Figure 5-11: SETUP – Clock

6200T Total Sulfur Analyzer Calibration Procedures Teledyne Analytical Instruments 197 9.5. Manual Calibration with IZS Option Under the best co

Page 110 - SECONDARY SETUP MENU

Calibration Procedures Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 198 To perform a manual calibration check of an analyze

Page 111

6200T Total Sulfur Analyzer Calibration Procedures Teledyne Analytical Instruments 199 STEP TWO: Perform the zero/span check. Wait until STAB

Page 112 - 5.7.2. INET (Ethernet)

Model 6200T Total Sulfur Analyzer ii Copyright © 2013 Teledyne Analytical Instruments All Rights Reserved. No part of this manual may be reproduced, t

Page 114

Calibration Procedures Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 200 9.7. Manual Calibration in DUAL or AUTO Reporting R

Page 115

6200T Total Sulfur Analyzer Calibration Procedures Teledyne Analytical Instruments 201 calibrate the analyzer until the contact is opened. At th

Page 116 - Figure 5-16: DIAG Menu

Calibration Procedures Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 202 Note The CALIBRATE attribute must always be set to

Page 117 - 5.9.1. Signal I/O

6200T Total Sulfur Analyzer Calibration Procedures Teledyne Analytical Instruments 203 To program the sample sequence shown in Table 9-4: SAMPL

Page 118

Calibration Procedures Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 204 With dynamic calibration turned on, the state of th

Page 119

6200T Total Sulfur Analyzer Calibration Procedures Teledyne Analytical Instruments 205 9.10. Calibration of Optional Sensors This section presen

Page 120 - ANALOG OUT

Calibration Procedures Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 206 Figure 9-14: O2 Span Gas Concentration Set Up

Page 121

6200T Total Sulfur Analyzer Calibration Procedures Teledyne Analytical Instruments 207 9.10.1.3. Activate O2 Sensor Stability Function To change

Page 122

Calibration Procedures Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 208 9.10.1.4. O2 Zero/Span Calibration To perform the z

Page 123 - 5-16), press:

6200T Total Sulfur Analyzer Calibration Procedures Teledyne Analytical Instruments 209 9.10.2. CO2 Sensor Calibration Calibration begins with co

Page 124

6200T Total Sulfur Analyzer Part I Teledyne Analytical Instruments 21 PART I GENERAL INFORMATION

Page 125

Calibration Procedures Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 210 9.10.2.2. Set CO2 Span Gas Concentration Set the ex

Page 126

6200T Total Sulfur Analyzer Calibration Procedures Teledyne Analytical Instruments 211 9.10.2.3. Activate CO2 Sensor Stability Function To chan

Page 127

Calibration Procedures Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 212 9.10.2.4. CO2 Zero/Span Calibration To perform the

Page 128

6200T Total Sulfur Analyzer EPA Protocol Calibration Teledyne Analytical Instruments 213 10. EPA PROTOCOL CALIBRATION 10.1. Calibration Requirem

Page 129

EPA Protocol Calibration Model 6200T Total Sulfur Analyzer 214 Teledyne Analytical Instruments Table 10-1: Activity Matrix for Calibration Equipm

Page 130 - 5.9.3.6. AIN Calibration

6200T Total Sulfur Analyzer EPA Protocol Calibration Teledyne Analytical Instruments 215 a NIST referenced voltage source or meter. Data recordi

Page 131 - Units and whether to display

EPA Protocol Calibration Model 6200T Total Sulfur Analyzer 216 Teledyne Analytical Instruments 10.1.7. Summary of Quality Assurance Checks The fo

Page 132 - 5.9.4. Optic Test

6200T Total Sulfur Analyzer EPA Protocol Calibration Teledyne Analytical Instruments 217 Table 10-4: Definition of Level 1 and Level 2 Zero and

Page 133 - 5.9.5. Electrical Test

EPA Protocol Calibration Model 6200T Total Sulfur Analyzer 218 Teledyne Analytical Instruments 10.3. ZERO and SPAN Checks A system of Level 1 and

Page 134 - 5.9.6. Lamp Calibration

6200T Total Sulfur Analyzer EPA Protocol Calibration Teledyne Analytical Instruments 219  Alternate units: ensure that the ppb units are selec

Page 135 - 5.9.7. Pressure Calibration

Part I Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 22 This page intentionally left blank.

Page 136 - 5.9.8. Flow Calibration

EPA Protocol Calibration Model 6200T Total Sulfur Analyzer 220 Teledyne Analytical Instruments Wait until STABIL falls below 0.5 ppb. This may tak

Page 137 - 5.9.9. Test Channel Output

6200T Total Sulfur Analyzer EPA Protocol Calibration Teledyne Analytical Instruments 221 2. Quality Assurance Handbook for Air Pollution Meas

Page 138

EPA Protocol Calibration Model 6200T Total Sulfur Analyzer 222 Teledyne Analytical Instruments This page intentionally left blank. .

Page 139 - (DTE DCE)

6200T Total Sulfur Analyzer Part III Teledyne Analytical Instruments 223 PART III MAINTENANCE AND SERVICE

Page 140 - 6.2.1. Communication Modes

Part III Model 6200T Total Sulfur Analyzer 224 Teledyne Analytical Instruments This page intentionally left blank.

Page 141 - EDIT

6200T Total Sulfur Analyzer Maintenance Teledyne Analytical Instruments 225 11. INSTRUMENT MAINTENANCE Predictive diagnostic functions including

Page 142 - 6.2.2. COMM Port Baud Rate

Maintenance Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 226 11.1. Maintenance Schedule Table 11-1 is the recommended main

Page 143 - 6.3. RS-232

6200T Total Sulfur Analyzer Maintenance Teledyen Analytical Instruments 227 11.2. Pre

Page 144 - 6.5. Ethernet

Maintenance Model 6200T Total Sulfur Analyzer 228 Teledyne Analytical Instruments 11.3. Maintenance Procedures The following procedures need to be

Page 145

6200T Total Sulfur Analyzer Maintenance Teledyen Analytical Instruments 229 4. Replace

Page 146 - Network Administrator

6200T Total Sulfur Analyzer Introduction Teledyne Analytical Instruments 23 1. INTRODUCTION, FEATURES AND OPTIONS This section provides an overv

Page 147 - SETUP X.X DHCP: ON

Maintenance Model 6200T Total Sulfur Analyzer 230 Teledyne Analytical Instruments 1. Locate the scrubber on the outside rear panel. 2. Remove t

Page 148 - 6.5.3. USB Port

6200T Total Sulfur Analyzer Maintenance Teledyen Analytical Instruments 231 Gas Line f

Page 149 - 6.6.1. MODBUS

Maintenance Model 6200T Total Sulfur Analyzer 232 Teledyne Analytical Instruments WARNING RISK OF ELECTRICAL SHOCK Some operations need to be car

Page 150 - Example MODBUS Poll window:

6200T Total Sulfur Analyzer Maintenance Teledyen Analytical Instruments 233 3. Pressur

Page 151 - 6.6.2. HESSEN

Maintenance Model 6200T Total Sulfur Analyzer 234 Teledyne Analytical Instruments 11.3.8.1. Checking the Scrubber for Leaks Leaks in the outer tub

Page 152

6200T Total Sulfur Analyzer Maintenance Teledyen Analytical Instruments 235 9. Wait 5

Page 153

Maintenance Model 6200T Total Sulfur Analyzer 236 Teledyne Analytical Instruments This page intentionally left blank.

Page 154

6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 237 12. TRO

Page 155

Troubleshooting Model 6200T Total Sulfur Analyzer 238 Teledyne Analytical Instruments Problem Corrective Action NOT HEATING: PV/SV switch to PV.

Page 156 - 6.6.2.7. Instrument ID

6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 239 12.1.1

Page 157

Introduction Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 24  Front panel USB ports for peripheral devices  Digital sta

Page 158 - 7.1. DAS Structure

Troubleshooting Model 6200T Total Sulfur Analyzer 240 Teledyne Analytical Instruments To view or clear a warning message press: SAMPLE

Page 159 - 7.1.3. DAS Triggering Events

6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 241 Warning

Page 160 - 7.2. Default DAS Channels

Troubleshooting Model 6200T Total Sulfur Analyzer 242 Teledyne Analytical Instruments 12.1.2. Fault Diagnosis with Test Functions Besides being us

Page 161 - Same parameters &

6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 243 TEST FU

Page 162 - MENU BUTTON FUNCTIONS

Troubleshooting Model 6200T Total Sulfur Analyzer 244 Teledyne Analytical Instruments Toggles parameter ON/OFF Exit returns to DIAG display &a

Page 163

6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 245 12.2. S

Page 164

Troubleshooting Model 6200T Total Sulfur Analyzer 246 Teledyne Analytical Instruments 12.2.3. Relay Board Status LEDs The most important status LE

Page 165 - 7.2.3. Trigger Events

6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 247 exists,

Page 166 - Press SET> until…

Troubleshooting Model 6200T Total Sulfur Analyzer 248 Teledyne Analytical Instruments  Carry out an optical test using the OPTIC TEST procedure

Page 167

6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 249 12.4.5.

Page 168

6200T Total Sulfur Analyzer Introduction Teledyne Analytical Instruments 25 OPTION OPTION NUMBER DESCRIPTION/NOTES REFERENCE 21 Rack mount brac

Page 169 - 7.2.6. Number of Records

Troubleshooting Model 6200T Total Sulfur Analyzer 250 Teledyne Analytical Instruments 12.5. Other Performance Problems Dynamic problems (i.e. prob

Page 170 - Press SET> until…

6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 251 determi

Page 171 - 7.2.9. Starting Date

Troubleshooting Model 6200T Total Sulfur Analyzer 252 Teledyne Analytical Instruments AC Configuration of the pump is accomplished via an in-line,

Page 172

6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 253 12.6.4.

Page 173 - 7.2.11. HOLDOFF Feature

Troubleshooting Model 6200T Total Sulfur Analyzer 254 Teledyne Analytical Instruments 12.6.7. Motherboard 12.6.7.1. A/D functions A basic check o

Page 174 - DAS feature

6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 255 1. Con

Page 175

Troubleshooting Model 6200T Total Sulfur Analyzer 256 Teledyne Analytical Instruments If the analyzer stops during initialization (the front panel

Page 176

6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 257 12.6.10

Page 177 - 8.1.1. Status Outputs

Troubleshooting Model 6200T Total Sulfur Analyzer 258 Teledyne Analytical Instruments  Check the analyzer's power supply, the relay board’s

Page 178 - 8.1.2. Control Inputs

6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 259 Table 1

Page 179

Introduction Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 26 OPTION OPTION NUMBER DESCRIPTION/NOTES REFERENCE SO2 IZS Perme

Page 180 - 8.2.3. Command Syntax

Troubleshooting Model 6200T Total Sulfur Analyzer 260 Teledyne Analytical Instruments option is enabled if the TEST functions show a parameter nam

Page 181 - 8.2.5. Status Reporting

6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 261 12.7. S

Page 182 - 8.3. Remote Access by Modem

Troubleshooting Model 6200T Total Sulfur Analyzer 262 Teledyne Analytical Instruments 12.7.2. Sensor Module Repair & Cleaning CAUTION - GENER

Page 183

6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 263 12.7.2.

Page 184

Troubleshooting Model 6200T Total Sulfur Analyzer 264 Teledyne Analytical Instruments 12.7.2.2. Cleaning the Sample chamber IMPORTANT IMPACT ON RE

Page 185 - V RS-232_PASS=NNNNNN

6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 265 The sam

Page 186

Troubleshooting Model 6200T Total Sulfur Analyzer 266 Teledyne Analytical Instruments 2. Remove the sample chamber from the PMT lens and filter h

Page 187 - 9. CALIBRATION PROCEDURES

6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 267 8. Usi

Page 188 - 9.1.1.2. Span Gas

Troubleshooting Model 6200T Total Sulfur Analyzer 268 Teledyne Analytical Instruments 2. Unplug J4 connector from the motherboard to allow tool a

Page 189 - 9.2. Manual Calibration

6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 269 Lamp Ag

Page 190 - STEP TWO: Set the expected SO

6200T Total Sulfur Analyzer Introduction Teledyne Analytical Instruments 27 OPTION OPTION NUMBER DESCRIPTION/NOTES REFERENCE Hessen protocol net

Page 191

Troubleshooting Model 6200T Total Sulfur Analyzer 270 Teledyne Analytical Instruments not adequate). 10. 1. Set the analyzer display to show the

Page 192

6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 271 .< 6

Page 193

Troubleshooting Model 6200T Total Sulfur Analyzer 272 Teledyne Analytical Instruments to this detail could twist and potentially disconnect the la

Page 194 - (At high

6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 273 PMT Hou

Page 195 - span gas value:

Troubleshooting Model 6200T Total Sulfur Analyzer 274 Teledyne Analytical Instruments 11. If the cold block or TEC is to be changed disconnect the

Page 196

6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 275 22. Al

Page 197

Troubleshooting Model 6200T Total Sulfur Analyzer 276 Teledyne Analytical Instruments 15. Set the front panel display to show STABIL (refer to Se

Page 198

6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 277 12.8. M

Page 199

Troubleshooting Model 6200T Total Sulfur Analyzer 278 Teledyne Analytical Instruments 5% of Full Scale is observed. Note the Thermal Converter te

Page 200 - Standard Procedure

6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 279 12.8.3.

Page 201

Introduction Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 28 M501 TS Figure 1-1: 6200T Basic Configuration

Page 202

Troubleshooting Model 6200T Total Sulfur Analyzer 280 Teledyne Analytical Instruments You will need the following tools:  Nutdriver, 5/16  Nut

Page 203 - Default

6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 281 Figure

Page 204 - 9.9. Calibration Quality

Troubleshooting Model 6200T Total Sulfur Analyzer 282 Teledyne Analytical Instruments Figure 12-23: Tie-Wrap Hold-Down Location 16. Form the

Page 205 - Sensor Calibration

6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 283 22. Aft

Page 206 - Figure 9-14: O

Troubleshooting Model 6200T Total Sulfur Analyzer 284 Teledyne Analytical Instruments QUESTION ANSWER How often do I need to change the particulat

Page 207 - Sensor Stability Function

6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 285 13. PRINCIPLES OF OPERATION This section describes the p

Page 208 - Zero/Span Calibration

Principles of Operation Model 6200T Total Sulfur Analyzer 286 Teledyne Analytical Instruments The amount SO2 converted to excited SO2* in the samp

Page 209 - 9.10.2. CO

6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 287 The amount of detectable UV given off by the decay of th

Page 210 - Span Gas Concentration

Principles of Operation Model 6200T Total Sulfur Analyzer 288 Teledyne Analytical Instruments The net result is that any variation in UV fluoresce

Page 211 - 9.10.2.3. Activate CO

6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 289 Figure 13-2: UV Light Path 13.1.3. UV Source Lamp The s

Page 212 - Zero/Span Calibration

6200T Total Sulfur Analyzer Introduction Teledyne Analytical Instruments 29 M501 TS Figure 1-2: 6200T with IZS/Permeation Tube Option

Page 213 - 10. EPA PROTOCOL CALIBRATION

Principles of Operation Model 6200T Total Sulfur Analyzer 290 Teledyne Analytical Instruments A PMT is typically a vacuum tube containing a variet

Page 214

6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 291 LAMP OUTPUT (Arbitrary Untis) 105104102202.5 307.6 481.

Page 215 - 10.1.6. Record Keeping

Principles of Operation Model 6200T Total Sulfur Analyzer 292 Teledyne Analytical Instruments 13.1.7.2. PMT Optical Filter The PMT used in the 620

Page 216 - NOT MET

6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 293 13.1.8. Optical Lenses Two optical lenses are used to fo

Page 217

Principles of Operation Model 6200T Total Sulfur Analyzer 294 Teledyne Analytical Instruments 13.1.9. Measurement Interferences It should be noted

Page 218 - 10.3. ZERO and SPAN Checks

6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 295 concentrations of some or all of these may be very high,

Page 219 - 10.4.2. Precision Check

Principles of Operation Model 6200T Total Sulfur Analyzer 296 Teledyne Analytical Instruments Figure 13-7: Oxygen Sensor - Principles of Opera

Page 220

6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 297 13.3. Carbon Dioxide (CO2) Sensor Principles of Operatio

Page 221

Principles of Operation Model 6200T Total Sulfur Analyzer 298 Teledyne Analytical Instruments 13.3.2. CO2 Operation within the 6200T Analyzer The

Page 222

6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 299 13.4. Pneumatic Operation IMPORTANT IMPACT ON READINGS O

Page 223 - MAINTENANCE AND SERVICE

6200T Total Sulfur Analyzer Configuration Teledyne Analytical Instruments iii INFORMATION ABOUT THE SPECIFIC CONFIGURATION OF YOUR MODEL 6200T

Page 224

Introduction Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 30 M501 TS Figure 1-3: 6200T with M702 Calibrator Option

Page 225 - 11. INSTRUMENT MAINTENANCE

Principles of Operation Model 6200T Total Sulfur Analyzer 300 Teledyne Analytical Instruments 13.4.2. Flow Rate Control The 6200T uses a special f

Page 226

6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 301 The actual flow rate of gas through the orifice (volume

Page 227 - 11.2. Predictive Diagnostics

Principles of Operation Model 6200T Total Sulfur Analyzer 302 Teledyne Analytical Instruments 30 minutes, it takes a certain amount of time for th

Page 228 - 11.3. Maintenance Procedures

6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 303 13.5. Electronic Operation Touchscreen Display I2C Bu

Page 229 - ATTENTION

Principles of Operation Model 6200T Total Sulfur Analyzer 304 Teledyne Analytical Instruments The motherboard is directly mounted to the rear pane

Page 230

6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 305 13.5.1. CPU The unit’s CPU card, installed on the mother

Page 231

Principles of Operation Model 6200T Total Sulfur Analyzer 306 Teledyne Analytical Instruments In the unlikely event that the flash chip should fai

Page 232

6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 307 13.5.2.1. Sample Chamber The main electronic components

Page 233 - Kicker)

Principles of Operation Model 6200T Total Sulfur Analyzer 308 Teledyne Analytical Instruments 13.5.3. Photo Multiplier Tube (PMT) The 6200T uses a

Page 234

6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 309 Figure 13-18: Basic PMT Design A significant perform

Page 235 - Customer Service

6200T Total Sulfur Analyzer Introduction Teledyne Analytical Instruments 31 1.6. The M501TS – Total Reduced Sulfur Converter The M501TS oxidizes

Page 236

Principles of Operation Model 6200T Total Sulfur Analyzer 310 Teledyne Analytical Instruments 13.5.4. PMT Cooling System The performance of the an

Page 237 - Circuit breaker Tripped?

6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 311 13.5.4.2. TEC Control Board The TEC control printed circ

Page 238

Principles of Operation Model 6200T Total Sulfur Analyzer 312 Teledyne Analytical Instruments COOLER (Closed Loop) To Motherboard PMT Preamp

Page 239

6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 313 The optical test (OTEST) feature causes an LED inside th

Page 240 - CONFIG INITIALIZED

Principles of Operation Model 6200T Total Sulfur Analyzer 314 Teledyne Analytical Instruments 13.5.7.3. Status LEDs & Watch Dog Circuitry Zero

Page 241

6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 315 As a safety measure, special circuitry on the relay boar

Page 242

Principles of Operation Model 6200T Total Sulfur Analyzer 316 Teledyne Analytical Instruments This measurement is stored in the analyzer. Memory a

Page 243 - DRK LMP

6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 317 first two of these outputs (A1 and A2). Refer to Section

Page 244

Principles of Operation Model 6200T Total Sulfur Analyzer 318 Teledyne Analytical Instruments situation before turning the analyzer back on. AC P

Page 245 - CPU Status LED

6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 319 13.6. Front Panel/Display Interface Users can input data

Page 246 - 12.3. Gas Flow Problems

Introduction Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 32 SAMPLEINCONVERTERTHERMALSWITCHPOWERTEMPERATURECONTROLLERTHERM

Page 247 - 12.4. Calibration Problems

Principles of Operation Model 6200T Total Sulfur Analyzer 320 Teledyne Analytical Instruments 13.7. Software Operation The instrument’s core modul

Page 248

6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 321 If necessary, these filter lengths of these two modes ma

Page 249 - 12.4.6. Non-Linear Response

Principles of Operation Model 6200T Total Sulfur Analyzer 322 Teledyne Analytical Instruments sample gas normalized to standard atmospheric pressu

Page 250 - 12.6. Subsystem Checkout

6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 323 This page intentionally left blank.

Page 252 - 12.6.3. I

6200T Total Sulfur Analyzer ESD Primer Teledyen Analytical Instruments 325 14. A PRIMER ON ELECTRO-STATIC DISCHARGE Teledyne Analytical Instrume

Page 253 - 12.6.6. Relay Board

ESD Primer Model 6200T Total Sulfur Analyzer 326 Teledyne Analytical Instruments Table 14-1: Static Generation Voltages for Typical Activities MEA

Page 254 - 12.6.7. Motherboard

6200T Total Sulfur Analyzer ESD Primer Teledyen Analytical Instruments 327  When static charges accumulated on a sensitive device discharges f

Page 255 - 12.6.8. CPU

ESD Primer Model 6200T Total Sulfur Analyzer 328 Teledyne Analytical Instruments components are prevented from building static electric charges th

Page 256 - 12.6.9. RS-232 Communication

6200T Total Sulfur Analyzer ESD Primer Teledyen Analytical Instruments 329 charges from building up on the device or assembly and nearby static

Page 257 - 12.6.11. PMT Sensor

6200T Total Sulfur Analyzer Introduction Teledyne Analytical Instruments 33 The sample containing the converted SO2 is then handed back to the a

Page 258

ESD Primer Model 6200T Total Sulfur Analyzer 330 Teledyne Analytical Instruments 3. Open the casing of the analyzer and begin work. Up to this p

Page 259 - 12.6.17. IZS Option

6200T Total Sulfur Analyzer ESD Primer Teledyen Analytical Instruments 331  If you are at an anti-ESD workbench, lay the container down on the

Page 260 - 12.6.19. PMT Temperature

ESD Primer Model 6200T Total Sulfur Analyzer 332 Teledyne Analytical Instruments Analytical Instruments’s Customer Service in anti-ESD bins, tubes

Page 261 - 12.7. Service Procedures

6200T Total Sulfur Analyzer Glossary Teledyen Analytical instruments 333 15. GLOSSARY Note: Some terms in this glossary may not occur elsewhere

Page 262

Glossary Model 6200T Total Sulfur Analyzer 334 Teledyne Analytical Instruments Term Description/Definition O2 molecular oxygen O3 ozone SO2 sulfur

Page 263

6200T Total Sulfur Analyzer Glossary Teledyen Analytical instruments 335 Term Description/Definition I2C bus a clocked, bi-directional, serial

Page 264

Glossary Model 6200T Total Sulfur Analyzer 336 Teledyne Analytical Instruments Term Description/Definition PTFE Polytetrafluoroethylene, a very i

Page 265

6200T Total Sulfur Analyzer Spare Parts Teledyen Analytical instruments 337 16. SPARE PARTS This section presents the Spare Parts Lists for the

Page 266

Spare Parts Model 6200T Total Sulfur Analyzer 338 Teledyne Analytical Instruments PART NUMBER DESCRIPTION 040010000 ASSY, FAN REAR PANEL

Page 267

6200T Total Sulfur Analyzer Spare Parts Teledyen Analytical instruments 339 PART NUMBER DESCRIPTION HW0000005 FOOT

Page 269

Spare Parts Model 6200T Total Sulfur Analyzer 340 Teledyne Analytical Instruments 018080000 AKIT, DESSICANT BAGGIES, (12) 039620100

Page 270

6200T Total Sulfur Analyzer Instrument Test and Calibration Record Teledyen Analytical instruments 341 17. INSTRUMENT TEST & CALIBRATION REC

Page 271

Instrument Test and Calibration Record Model 6200T Total Sulfur Analyzer 342 Teledyne Analytical Instruments Table 17-2. For test and calibration in

Page 272

6200T Total Sulfur Analyzer Instrument Test and Calibration Record Teledyen Analytical instruments 343 Table 17-2. Test and Calibration Values

Page 273

Instrument Test and Calibration Record Model 6200T Total Sulfur Analyzer 344 Teledyne Analytical Instruments Table 17-3. Test and Calibrations Val

Page 274

6200T Total Sulfur Analyzer Specifications Teledyne Analytical Instruments 35 2. SPECIFICATIONS, APPROVALS & COMPLIANCE This section present

Page 275

Specifications Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 36 Parameter Description Certifications EN61326 (1997 w/A1: 98)

Page 276 - MPORTANT

6200T Total Sulfur Analyzer Specifications Teledyne Analytical Instruments 37 Table 2-4: M501TS Converter Specifications Specification Value Un

Page 277

Specifications Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 38 Dynamic span OFF Dynamic zero OFF Dilution factor OFF AutoC

Page 278

6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 39 3. GETTING STARTED This section addresses the procedures for unpa

Page 279

Configuration Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments iv Selected Options for the Model 6200T Calibrator Option: This op

Page 280

Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 40 1. Remove the locking screw located in the top, center of

Page 281

6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 41  Thumb drive (not included) to download updates to instruction

Page 282

Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 42 The analyzer’s front panel liquid crystal display screen in

Page 283

6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 43 Figure 3-3: Display/Touch Control Screen Mapped to Menu Charts

Page 284 - 12.10. Technical Assistance

Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 44 The controller has been set up at the factory. Should furt

Page 285 - *2SOhvSO

6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 45 Table 3-3: Rear Panel Description—Analyzer and Converter Compon

Page 286 - SOaxexp1IIa

Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 46 3.2.3. Internal Chassis Layout Figure 3-5A illustrates the

Page 287 - = Amount of excited SO

6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 47 SAMPLEINCONVERTERTHERMALSWITCHPOWERTEMPERATURECONTROLLERTHERMOCO

Page 288 - 13.1.2. The UV Light Path

Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 48 3.3. Connections and Setup This section presents the electr

Page 289 - 13.1.5. The PMT

6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 49 3.3.1.1. Connecting Power Attach the power cords to the analyzer

Page 290 - 13.1.7. Optical Filters

6200T Total Sulfur Analyzer About This Manual Teledyne Analytical Instruments v ABOUT THIS MANUAL This manual describes operation, specification

Page 291 - BEFORE AFTER

Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 50 Table 3-5: Analog Input Pin Assignments PIN DESCRIPTION DA

Page 292 - 13.1.7.2. PMT Optical Filter

6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 51 3.3.1.4. Current Loop Analog Outputs (Option 41) Setup If your a

Page 293 - 13.1.8. Optical Lenses

Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 52 Figure 3-8: Current Loop Option Installed on the Motherboa

Page 294 - 13.1.9.3. Dilution

6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 53 3.3.1.5. Connecting the Status Outputs The status outputs report

Page 295 - 13.2. Oxygen (O

Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 54 Table 3-7: Status Output Signals REAR PANEL LABEL STATUS DE

Page 296 - 13.2.2. O

6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 55 CONTROL IN A B C D E F U + S

Page 297 - 13.3. Carbon Dioxide (CO

Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 56 3.3.1.7. Connecting the Concentration Alarm Relay (Option 6

Page 298 - Sensor

6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 57 SO2 Alarm 2 = 100 PPM SO2 Alarm 1 = 20 PPM SO2 Alarm 2 = 100 PPM

Page 299 - CRITICAL

Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 58 RS-232 Connection For RS-232 communications with data term

Page 300 - 13.4.2. Flow Rate Control

6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 59 Figure 3-13: CPU Connector Pin-Outs for RS-232 Mode RS-232 COMM

Page 301

About This Manual Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments vi This page intentionally left blank.

Page 302 - 13.4.4. Pneumatic Sensors

Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 60 DB9 connectors on the analyzer’s rear panel to connect the

Page 303 - CPU Card

6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 61 3. Close the instrument. 4. Referring to Figure 3-15 use straig

Page 304

Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 62 3.3.2. Pneumatic Connections This section provides not only

Page 305 - 13.5.1. CPU

6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 63 Dust Plugs: Remove dust plugs from rear panel exhaust and su

Page 306 - 13.5.2. Sensor Module

Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 64 One primary application of this instrument is to analyze CO

Page 307 - 13.5.2.1. Sample Chamber

6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 65 Figure 3-16: Pneumatic Connections–Basic Configuration–Using Bo

Page 308

Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 66 Figure 3-17: Pneumatic Connections–Basic Configuration–Us

Page 309

6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 67 3.3.2.2. Pneumatic Layout for Basic Configuration SAMPLE PRESSURE

Page 310 - 13.5.4. PMT Cooling System

Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 68 3.3.2.3. Pneumatic Layout for Zero/Span Valves Option Figur

Page 311 - 13.5.5. PMT Preamplifier

6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 69  remotely through the RS-232/485 serial I/O ports (refer to App

Page 312 - PMT Preamp PCA

6200T Total Sulfur Analyzer Safety Messages Teledyne Analytical Instruments vii SAFETY MESSAGES Important safety messages are provided throughou

Page 313 - 13.5.7. Relay Board

Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 70 ZERO CAL Sample/Cal Open to zero/span valve Zero/Span Ope

Page 314 - D6 YELLOW Unused N/A N/A

6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 71 3.3.2.7. TS and Zero Air Scrubbers There are two charcoal scrubbe

Page 315 - 13.5.8. Motherboard

Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 72 3.3.2.9. Pneumatic Layout with CO2 Sensor Option Figure 3-2

Page 316 - 13.5.9. Analog Outputs

6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 73 For analyzers without an IZS or external zero air scrubber option

Page 317 - 13.5.12. Power up Circuit

Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 74 Span Gas for Multipoint Calibration Some applications, such

Page 318

6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 75 The converter has not been tested for conversion efficiency or re

Page 319 - Touchscreen

Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 76 Table includes typical values for a converter set up for o

Page 320 - 13.7.1. Adaptive Filter

6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 77 3.4.2. Warning Messages Because internal temperatures and other

Page 321

Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 78 Table 3-13 lists brief descriptions of the warning messages

Page 322 - A)-HG(" 29.92

6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 79 Table 3-14: Possible Startup Warning Messages – 6200T Analyzers

Page 323

Safety Messages Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments viii CONSIGNES DE SÉCURITÉ Des consignes de sécurité important

Page 324

Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 80 To view the current values of these parameters press the fo

Page 325

6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 81 3.4.4. Initial Calibration To perform the following calibration y

Page 326 - 10-25% RH

Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 82 While these are the default settings for the 6200T analyzer

Page 327

6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 83 Dilution Ratio Setup If the dilution ratio option is enabled on y

Page 328 - Ground Point

Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 84 Set SO2 Span Gas Concentration Set the expected SO2 span ga

Page 329 - Maintenance

6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 85 Zero/Span Calibration To perform the zero/span calibration proce

Page 330

Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 86 3.4.4.2. Calibration Procedure for the O2 Option If your an

Page 331

6200T Total Sulfur Analyzer Part II Teledyne Analytical Instruments 87 PART II OPERATING INSTRUCTIONS

Page 332

Part II Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 88 This page intentionally left blank.

Page 333 - 15. GLOSSARY

6200T Total Sulfur Analyzer Overview of Operating Modes Teledyne Analytical Instruments 89 4. OVERVIEW OF OPERATING MODES To assist in navigatin

Page 334

6200T Total Sulfur Analyzer Table of Contents Teledyne Analytical Instruments ix TABLE OF CONTENTS Selected Versions of the Model 6200T ...

Page 335

Overview of Operating Modes Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 90 LO CAL A1 Unit is performing LOW SPAN (midpoin

Page 336

6200T Total Sulfur Analyzer Overview of Operating Modes Teledyne Analytical Instruments 91 Table 4-2: Test Functions Defined DISPLAY PARAMETER

Page 337 - 16. SPARE PARTS

Overview of Operating Modes Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 92 To view the TEST Functions press the following

Page 338 - DESCRIPTION

6200T Total Sulfur Analyzer Overview of Operating Modes Teledyne Analytical Instruments 93 4.1.2. Warning Messages The most common instrument fa

Page 339

Overview of Operating Modes Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 94 To view and clear warning messages, press: SA

Page 340

6200T Total Sulfur Analyzer Overview of Operating Modes Teledyne Analytical Instruments 95 4.2. Calibration Mode Pressing the CAL button switche

Page 341

Overview of Operating Modes Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 96 4.3.2. Primary Setup Menu Table 4-4: Primary Se

Page 342 - Table 17-2

6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 97 5. SETUP MENU The SETUP menu is used to set instrument parameters for

Page 343

Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 98 5.4.1. Available Analog Output Signals The analyzer has three ac

Page 344

6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 99 concentration of SO2 is typically less than 500 ppb, the full scale of