Teledyne Analytical Instruments Operation Manual for Model 6200T UV Fluorescence Total Sulfur Analyzer TELEDYNE ELECTRONIC TECHNOLOGIES Analyt
Table of Contents Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments x 4.1.2. Warning Messages ...
Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 100 5.4.3. Reporting Range Modes: Single, Dual, Auto Ranges The 620
6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 101 5.4.3.1. Single Range mode (SNGL) The default range mode for the anal
Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 102 5.4.3.2. Dual Range Mode (DUAL) Selecting Dual Range mode allow
6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 103 5.4.3.3. Auto Range Mode (AUTO) In AUTO range mode, the analyzer auto
Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 104 milligrams per cubic meter (mg/m3, MGM). Changing units affects
6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 105 5.4.5. Dilution Ratio (Option) The dilution ratio is a software optio
Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 106 IMPORTANT IMPACT ON READINGS OR DATA Once the above settings ha
6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 107 To enable or disable passwords, press: SAMPLE ENTER SE
Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 108 If the password feature is enabled, then when entering either C
6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 109 5.6. SETUP – CLK: Setting the Internal Time-of-Day Clock The 6200T ha
6200T Total Sulfur Analyzer Table of Contents Teledyne Analytical Instruments xi 7.2.5. Sample Period and Report Period ...
Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 110 SAMPLE RANGE = 500.000 PPB SO2 =XXX.X <
6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 111 5.7. SETUP – COMM: Communications Ports This section introduces the c
Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 112 SETUP X.X COMMUNICATIONS MENU ID INET COM1
6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 113 5.8. SETUP – VARS: Variables Setup and Definition Through the SETUP&g
Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 114 To access and navigate the VARS menu, use the following button
6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 115 5.9. SETUP – DIAG: Diagnostics Functions The SETUP>MORE>DIAG me
Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 116 To access the DIAG functions press the following buttons:
6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 117 5.9.1. Signal I/O The signal I/O diagnostic mode allows a user to rev
Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 118 5.9.2. Analog Output Step Test Analog Output is used as a step
6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 119 5.9.3. Analog I/O Configuration Table 6-8 lists the analog I/O functi
Table of Contents Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments xii 11. INSTRUMENT MAINTENANCE ...
Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 120 The following DC current output limits apply to the current loo
6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 121 5.9.3.1. Analog Output Signal Type and Range Span Selection To select
Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 122 5.9.3.2. Analog Output Calibration Mode Analog output calibrati
6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 123 To automatically calibrate a single analog channel from the DIAG Menu
Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 124 To select manual output calibration for a particular channel, a
6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 125 5.9.3.3. Manual Analog Output Calibration and Voltage Adjustment For
Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 126 To make these manual adjustments, the AOUT auto-calibration fea
6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 127 5.9.3.4. Analog Output Offset Adjustment Some analog signal recorders
Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 128 5.9.3.5. Current Loop Output Adjustment A current loop option i
6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 129 To adjust the zero and span values of the current outputs, activate t
6200T Total Sulfur Analyzer Table of Contents Teledyne Analytical Instruments xiii 12.7.2. Sensor Module Repair & Cleaning ...
Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 130 5.9.3.6. AIN Calibration This is the sub-menu to conduct the an
6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 131 To adjust settings for the Analog Inputs option parameters press:
Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 132 5.9.4. Optic Test The optic test function tests the response of
6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 133 5.9.5. Electrical Test The electrical test function creates a current
Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 134 5.9.6. Lamp Calibration An important factor in accurately deter
6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 135 5.9.7. Pressure Calibration A sensor at the exit of the sample chambe
Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 136 5.9.8. Flow Calibration The flow calibration allows the user to
6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 137 5.9.9. Test Channel Output When activated, output channel A3 can be u
Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 138 Once a TEST function is selected, the instrument begins to rep
6200T Total Sulfur Analyzer Communications Setup Teledyne Analytical Instruments 139 6. COMMUNICATIONS SETUP AND OPERATION This instrument rear
Table of Contents Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments xiv 14.4.1. General Rules ...
Communications Setup Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 140 6.2.1. Communication Modes Either of the analyzer’s s
6200T Total Sulfur Analyzer Communications Setup Teledyne Analytical Instruments 141 To turn on or off the communication modes for either COM1
Communications Setup Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 142 6.2.2. COMM Port Baud Rate To select the baud rate of
6200T Total Sulfur Analyzer Communications Setup Teledyne Analytical Instruments 143 6.2.3. COMM Port Testing The serial ports can be tested for
Communications Setup Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 144 analyzers, monitors, calibrators, etc.) into one comm
6200T Total Sulfur Analyzer Communications Setup Teledyne Analytical Instruments 145 Table 6-3: LAN/Internet Default Configuration Properties PR
Communications Setup Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 146 Internet Configuration Button Functions BUT
6200T Total Sulfur Analyzer Communications Setup Teledyne Analytical Instruments 147 6.5.2. Configuring Ethernet Communication Using Dynamic Hos
Communications Setup Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 148 change this name (particularly if you have more than
6200T Total Sulfur Analyzer Communications Setup Teledyne Analytical Instruments 149 1. Go to SETUP>MORE>COMM>COM2 menu. 2. Press the
6200T Total Sulfur Analyzer Table of Contents Teledyne Analytical Instruments xv LIST OF FIGURES Figure 1-1: 6200T Basic Configuration...
Communications Setup Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 150 Example Read/Write Definition window: Example Connec
6200T Total Sulfur Analyzer Communications Setup Teledyne Analytical Instruments 151 6.6.2. HESSEN The Hessen protocol is a Multidrop protocol,
Communications Setup Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 152 6.6.2.2. Activating Hessen Protocol The first step
6200T Total Sulfur Analyzer Communications Setup Teledyne Analytical Instruments 153 6.6.2.3. Selecting a Hessen Protocol Type Currently there a
Communications Setup Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 154 6.6.2.4. Setting the Hessen Protocol Response Mode Te
6200T Total Sulfur Analyzer Communications Setup Teledyne Analytical Instruments 155 IMPORTANT IMPACT ON READINGS OR DATA It is possible to as
Communications Setup Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 156 To assign or reset the status flag bit assignments, a
6200T Total Sulfur Analyzer DAS and AICOM Teledyne Analytical Instruments 157 7. DATA ACQUISITION SYSTEM (DAS) AND AICOM The 6200T analyzer cont
DAS and AICOM Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 158 7.1. DAS Structure The DAS is designed around the feature of
6200T Total Sulfur Analyzer DAS and AICOM Teledyne Analytical Instruments 159 7.1.2. DAS Parameters Data parameters are types of data that may b
Table of Contents Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments xvi Figure 5-22: DIAG – Analog Output – Auto Cal or Manual
DAS and AICOM Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 160 WARNINGS: Some data may be useful when stored if one of s
6200T Total Sulfur Analyzer DAS and AICOM Teledyne Analytical Instruments 161 The Channel Properties, Triggering Events and Data Parameters/Func
DAS and AICOM Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 162 7.2.1. Viewing DAS Data and Settings DAS data and settings c
6200T Total Sulfur Analyzer DAS and AICOM Teledyne Analytical Instruments 163 7.2.2. Editing DAS Data Channels Although DAS configuration is mos
DAS and AICOM Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 164 When editing the data channels, the top line of the display
6200T Total Sulfur Analyzer DAS and AICOM Teledyne Analytical Instruments 165 7.2.3. Trigger Events To edit the list of data parameters associa
DAS and AICOM Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 166 7.2.4. Editing DAS Parameters Data channels can be edited in
6200T Total Sulfur Analyzer DAS and AICOM Teledyne Analytical Instruments 167 To configure the parameters for a specific data parameter, follow
DAS and AICOM Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 168 SAMPLE PERIOD: Determines how often DAS temporarily records
6200T Total Sulfur Analyzer DAS and AICOM Teledyne Analytical Instruments 169 To define the REPORT PERIOD, follow the instruction shown in Figur
6200T Total Sulfur Analyzer Table of Contents Teledyne Analytical Instruments xvii Figure 9-15: Activate O2 Sensor Stability Function ...
DAS and AICOM Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 170 Edit Data Channel Menu SETUP X.X 0) CONC: A
6200T Total Sulfur Analyzer DAS and AICOM Teledyne Analytical Instruments 171 7.2.7. RS-232 Report Function The M DAS can automatically report d
DAS and AICOM Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 172 To disable a data channel, follow the instruction shown in F
6200T Total Sulfur Analyzer DAS and AICOM Teledyne Analytical Instruments 173 7.2.11. HOLDOFF Feature The DAS HOLDOFF feature allows to prevent
DAS and AICOM Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 174 Check on system parameters for trouble-shooting and quali
6200T Total Sulfur Analyzer DAS and AICOM Teledyne Analytical Instruments 175 7.4. Remote DAS Configuration via AICOM Editing channels, paramete
DAS and AICOM Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 176 Figure 7-15: DAS Configuration Through a Terminal Emulat
6200T Total Sulfur Analyzer Remote Operation Teledyne Analytical Instruments 177 8. REMOTE OPERATION OF THE ANALYZER This section provides infor
Remote Operation Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 178 Table 8-1: Status Output Pin Assignments CONNECTOR PIN S
6200T Total Sulfur Analyzer Remote Operation Teledyne Analytical Instruments 179 CONTROL IN A B C D E F U +
Table of Contents Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments xviii LIST OF TABLES Table 1-1: Analyzer Options ...
Remote Operation Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 180 8.2.2. Help Commands in Terminal Mode Table 8-3: Terminal
6200T Total Sulfur Analyzer Remote Operation Teledyne Analytical Instruments 181 8.2.4. Data Types Data types consist of integers, hexadecimal i
Remote Operation Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 182 DDD:HH:MM is the time stamp, the date and time when the
6200T Total Sulfur Analyzer Remote Operation Teledyne Analytical Instruments 183 To change this setting, access the COMMUNICATIONS Menu (refer t
Remote Operation Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 184 To initialize the modem, access the COMMUNICATIONS Menu (
6200T Total Sulfur Analyzer Remote Operation Teledyne Analytical Instruments 185 LOGON FAILED - Password not given or incorrect LOGOFF SUC
Remote Operation Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 186 This page intentionally left blank.
6200T Total Sulfur Analyzer Calibration Procedures Teledyne Analytical Instruments 187 9. CALIBRATION PROCEDURES This section describes the cali
Calibration Procedures Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 188 standard charcoal scrubber that supplies zero air f
6200T Total Sulfur Analyzer Calibration Procedures Teledyne Analytical Instruments 189 9.1.1.3. Calibration Gas Standards and Traceability All e
6200T Total Sulfur Analyzer Table of Contents Teledyne Analytical Instruments xix Table 10-4: Definition of Level 1 and Level 2 Zero and Span C
Calibration Procedures Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 190 OR SAMPLEEXHAUSTMODEL 701Zero Air Generator VENT
6200T Total Sulfur Analyzer Calibration Procedures Teledyne Analytical Instruments 191 This sequence causes the analyzer to prompt for the expe
Calibration Procedures Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 192 M-P CAL SO2 STB=X.XXX PPB SO2 =
6200T Total Sulfur Analyzer Calibration Procedures Teledyne Analytical Instruments 193 IMPORTANT IMPACT ON READINGS OR DATA If the ZERO or SPAN
Calibration Procedures Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 194 9.4. Manual Calibration with Zero/Span Valves Zero
6200T Total Sulfur Analyzer Calibration Procedures Teledyne Analytical Instruments 195 STEP TWO: Set the expected SO2 span gas value: This sequ
Calibration Procedures Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 196 Step Three: Perform the calibration or calibration
6200T Total Sulfur Analyzer Calibration Procedures Teledyne Analytical Instruments 197 9.5. Manual Calibration with IZS Option Under the best co
Calibration Procedures Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 198 To perform a manual calibration check of an analyze
6200T Total Sulfur Analyzer Calibration Procedures Teledyne Analytical Instruments 199 STEP TWO: Perform the zero/span check. Wait until STAB
Model 6200T Total Sulfur Analyzer ii Copyright © 2013 Teledyne Analytical Instruments All Rights Reserved. No part of this manual may be reproduced, t
Calibration Procedures Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 200 9.7. Manual Calibration in DUAL or AUTO Reporting R
6200T Total Sulfur Analyzer Calibration Procedures Teledyne Analytical Instruments 201 calibrate the analyzer until the contact is opened. At th
Calibration Procedures Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 202 Note The CALIBRATE attribute must always be set to
6200T Total Sulfur Analyzer Calibration Procedures Teledyne Analytical Instruments 203 To program the sample sequence shown in Table 9-4: SAMPL
Calibration Procedures Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 204 With dynamic calibration turned on, the state of th
6200T Total Sulfur Analyzer Calibration Procedures Teledyne Analytical Instruments 205 9.10. Calibration of Optional Sensors This section presen
Calibration Procedures Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 206 Figure 9-14: O2 Span Gas Concentration Set Up
6200T Total Sulfur Analyzer Calibration Procedures Teledyne Analytical Instruments 207 9.10.1.3. Activate O2 Sensor Stability Function To change
Calibration Procedures Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 208 9.10.1.4. O2 Zero/Span Calibration To perform the z
6200T Total Sulfur Analyzer Calibration Procedures Teledyne Analytical Instruments 209 9.10.2. CO2 Sensor Calibration Calibration begins with co
6200T Total Sulfur Analyzer Part I Teledyne Analytical Instruments 21 PART I GENERAL INFORMATION
Calibration Procedures Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 210 9.10.2.2. Set CO2 Span Gas Concentration Set the ex
6200T Total Sulfur Analyzer Calibration Procedures Teledyne Analytical Instruments 211 9.10.2.3. Activate CO2 Sensor Stability Function To chan
Calibration Procedures Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 212 9.10.2.4. CO2 Zero/Span Calibration To perform the
6200T Total Sulfur Analyzer EPA Protocol Calibration Teledyne Analytical Instruments 213 10. EPA PROTOCOL CALIBRATION 10.1. Calibration Requirem
EPA Protocol Calibration Model 6200T Total Sulfur Analyzer 214 Teledyne Analytical Instruments Table 10-1: Activity Matrix for Calibration Equipm
6200T Total Sulfur Analyzer EPA Protocol Calibration Teledyne Analytical Instruments 215 a NIST referenced voltage source or meter. Data recordi
EPA Protocol Calibration Model 6200T Total Sulfur Analyzer 216 Teledyne Analytical Instruments 10.1.7. Summary of Quality Assurance Checks The fo
6200T Total Sulfur Analyzer EPA Protocol Calibration Teledyne Analytical Instruments 217 Table 10-4: Definition of Level 1 and Level 2 Zero and
EPA Protocol Calibration Model 6200T Total Sulfur Analyzer 218 Teledyne Analytical Instruments 10.3. ZERO and SPAN Checks A system of Level 1 and
6200T Total Sulfur Analyzer EPA Protocol Calibration Teledyne Analytical Instruments 219 Alternate units: ensure that the ppb units are selec
Part I Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 22 This page intentionally left blank.
EPA Protocol Calibration Model 6200T Total Sulfur Analyzer 220 Teledyne Analytical Instruments Wait until STABIL falls below 0.5 ppb. This may tak
6200T Total Sulfur Analyzer EPA Protocol Calibration Teledyne Analytical Instruments 221 2. Quality Assurance Handbook for Air Pollution Meas
EPA Protocol Calibration Model 6200T Total Sulfur Analyzer 222 Teledyne Analytical Instruments This page intentionally left blank. .
6200T Total Sulfur Analyzer Part III Teledyne Analytical Instruments 223 PART III MAINTENANCE AND SERVICE
Part III Model 6200T Total Sulfur Analyzer 224 Teledyne Analytical Instruments This page intentionally left blank.
6200T Total Sulfur Analyzer Maintenance Teledyne Analytical Instruments 225 11. INSTRUMENT MAINTENANCE Predictive diagnostic functions including
Maintenance Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 226 11.1. Maintenance Schedule Table 11-1 is the recommended main
6200T Total Sulfur Analyzer Maintenance Teledyen Analytical Instruments 227 11.2. Pre
Maintenance Model 6200T Total Sulfur Analyzer 228 Teledyne Analytical Instruments 11.3. Maintenance Procedures The following procedures need to be
6200T Total Sulfur Analyzer Maintenance Teledyen Analytical Instruments 229 4. Replace
6200T Total Sulfur Analyzer Introduction Teledyne Analytical Instruments 23 1. INTRODUCTION, FEATURES AND OPTIONS This section provides an overv
Maintenance Model 6200T Total Sulfur Analyzer 230 Teledyne Analytical Instruments 1. Locate the scrubber on the outside rear panel. 2. Remove t
6200T Total Sulfur Analyzer Maintenance Teledyen Analytical Instruments 231 Gas Line f
Maintenance Model 6200T Total Sulfur Analyzer 232 Teledyne Analytical Instruments WARNING RISK OF ELECTRICAL SHOCK Some operations need to be car
6200T Total Sulfur Analyzer Maintenance Teledyen Analytical Instruments 233 3. Pressur
Maintenance Model 6200T Total Sulfur Analyzer 234 Teledyne Analytical Instruments 11.3.8.1. Checking the Scrubber for Leaks Leaks in the outer tub
6200T Total Sulfur Analyzer Maintenance Teledyen Analytical Instruments 235 9. Wait 5
Maintenance Model 6200T Total Sulfur Analyzer 236 Teledyne Analytical Instruments This page intentionally left blank.
6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 237 12. TRO
Troubleshooting Model 6200T Total Sulfur Analyzer 238 Teledyne Analytical Instruments Problem Corrective Action NOT HEATING: PV/SV switch to PV.
6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 239 12.1.1
Introduction Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 24 Front panel USB ports for peripheral devices Digital sta
Troubleshooting Model 6200T Total Sulfur Analyzer 240 Teledyne Analytical Instruments To view or clear a warning message press: SAMPLE
6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 241 Warning
Troubleshooting Model 6200T Total Sulfur Analyzer 242 Teledyne Analytical Instruments 12.1.2. Fault Diagnosis with Test Functions Besides being us
6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 243 TEST FU
Troubleshooting Model 6200T Total Sulfur Analyzer 244 Teledyne Analytical Instruments Toggles parameter ON/OFF Exit returns to DIAG display &a
6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 245 12.2. S
Troubleshooting Model 6200T Total Sulfur Analyzer 246 Teledyne Analytical Instruments 12.2.3. Relay Board Status LEDs The most important status LE
6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 247 exists,
Troubleshooting Model 6200T Total Sulfur Analyzer 248 Teledyne Analytical Instruments Carry out an optical test using the OPTIC TEST procedure
6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 249 12.4.5.
6200T Total Sulfur Analyzer Introduction Teledyne Analytical Instruments 25 OPTION OPTION NUMBER DESCRIPTION/NOTES REFERENCE 21 Rack mount brac
Troubleshooting Model 6200T Total Sulfur Analyzer 250 Teledyne Analytical Instruments 12.5. Other Performance Problems Dynamic problems (i.e. prob
6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 251 determi
Troubleshooting Model 6200T Total Sulfur Analyzer 252 Teledyne Analytical Instruments AC Configuration of the pump is accomplished via an in-line,
6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 253 12.6.4.
Troubleshooting Model 6200T Total Sulfur Analyzer 254 Teledyne Analytical Instruments 12.6.7. Motherboard 12.6.7.1. A/D functions A basic check o
6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 255 1. Con
Troubleshooting Model 6200T Total Sulfur Analyzer 256 Teledyne Analytical Instruments If the analyzer stops during initialization (the front panel
6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 257 12.6.10
Troubleshooting Model 6200T Total Sulfur Analyzer 258 Teledyne Analytical Instruments Check the analyzer's power supply, the relay board’s
6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 259 Table 1
Introduction Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 26 OPTION OPTION NUMBER DESCRIPTION/NOTES REFERENCE SO2 IZS Perme
Troubleshooting Model 6200T Total Sulfur Analyzer 260 Teledyne Analytical Instruments option is enabled if the TEST functions show a parameter nam
6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 261 12.7. S
Troubleshooting Model 6200T Total Sulfur Analyzer 262 Teledyne Analytical Instruments 12.7.2. Sensor Module Repair & Cleaning CAUTION - GENER
6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 263 12.7.2.
Troubleshooting Model 6200T Total Sulfur Analyzer 264 Teledyne Analytical Instruments 12.7.2.2. Cleaning the Sample chamber IMPORTANT IMPACT ON RE
6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 265 The sam
Troubleshooting Model 6200T Total Sulfur Analyzer 266 Teledyne Analytical Instruments 2. Remove the sample chamber from the PMT lens and filter h
6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 267 8. Usi
Troubleshooting Model 6200T Total Sulfur Analyzer 268 Teledyne Analytical Instruments 2. Unplug J4 connector from the motherboard to allow tool a
6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 269 Lamp Ag
6200T Total Sulfur Analyzer Introduction Teledyne Analytical Instruments 27 OPTION OPTION NUMBER DESCRIPTION/NOTES REFERENCE Hessen protocol net
Troubleshooting Model 6200T Total Sulfur Analyzer 270 Teledyne Analytical Instruments not adequate). 10. 1. Set the analyzer display to show the
6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 271 .< 6
Troubleshooting Model 6200T Total Sulfur Analyzer 272 Teledyne Analytical Instruments to this detail could twist and potentially disconnect the la
6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 273 PMT Hou
Troubleshooting Model 6200T Total Sulfur Analyzer 274 Teledyne Analytical Instruments 11. If the cold block or TEC is to be changed disconnect the
6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 275 22. Al
Troubleshooting Model 6200T Total Sulfur Analyzer 276 Teledyne Analytical Instruments 15. Set the front panel display to show STABIL (refer to Se
6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 277 12.8. M
Troubleshooting Model 6200T Total Sulfur Analyzer 278 Teledyne Analytical Instruments 5% of Full Scale is observed. Note the Thermal Converter te
6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 279 12.8.3.
Introduction Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 28 M501 TS Figure 1-1: 6200T Basic Configuration
Troubleshooting Model 6200T Total Sulfur Analyzer 280 Teledyne Analytical Instruments You will need the following tools: Nutdriver, 5/16 Nut
6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 281 Figure
Troubleshooting Model 6200T Total Sulfur Analyzer 282 Teledyne Analytical Instruments Figure 12-23: Tie-Wrap Hold-Down Location 16. Form the
6200T Total Sulfur Analyzer Troubleshooting Teledyen Analytical Instruments 283 22. Aft
Troubleshooting Model 6200T Total Sulfur Analyzer 284 Teledyne Analytical Instruments QUESTION ANSWER How often do I need to change the particulat
6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 285 13. PRINCIPLES OF OPERATION This section describes the p
Principles of Operation Model 6200T Total Sulfur Analyzer 286 Teledyne Analytical Instruments The amount SO2 converted to excited SO2* in the samp
6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 287 The amount of detectable UV given off by the decay of th
Principles of Operation Model 6200T Total Sulfur Analyzer 288 Teledyne Analytical Instruments The net result is that any variation in UV fluoresce
6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 289 Figure 13-2: UV Light Path 13.1.3. UV Source Lamp The s
6200T Total Sulfur Analyzer Introduction Teledyne Analytical Instruments 29 M501 TS Figure 1-2: 6200T with IZS/Permeation Tube Option
Principles of Operation Model 6200T Total Sulfur Analyzer 290 Teledyne Analytical Instruments A PMT is typically a vacuum tube containing a variet
6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 291 LAMP OUTPUT (Arbitrary Untis) 105104102202.5 307.6 481.
Principles of Operation Model 6200T Total Sulfur Analyzer 292 Teledyne Analytical Instruments 13.1.7.2. PMT Optical Filter The PMT used in the 620
6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 293 13.1.8. Optical Lenses Two optical lenses are used to fo
Principles of Operation Model 6200T Total Sulfur Analyzer 294 Teledyne Analytical Instruments 13.1.9. Measurement Interferences It should be noted
6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 295 concentrations of some or all of these may be very high,
Principles of Operation Model 6200T Total Sulfur Analyzer 296 Teledyne Analytical Instruments Figure 13-7: Oxygen Sensor - Principles of Opera
6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 297 13.3. Carbon Dioxide (CO2) Sensor Principles of Operatio
Principles of Operation Model 6200T Total Sulfur Analyzer 298 Teledyne Analytical Instruments 13.3.2. CO2 Operation within the 6200T Analyzer The
6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 299 13.4. Pneumatic Operation IMPORTANT IMPACT ON READINGS O
6200T Total Sulfur Analyzer Configuration Teledyne Analytical Instruments iii INFORMATION ABOUT THE SPECIFIC CONFIGURATION OF YOUR MODEL 6200T
Introduction Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 30 M501 TS Figure 1-3: 6200T with M702 Calibrator Option
Principles of Operation Model 6200T Total Sulfur Analyzer 300 Teledyne Analytical Instruments 13.4.2. Flow Rate Control The 6200T uses a special f
6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 301 The actual flow rate of gas through the orifice (volume
Principles of Operation Model 6200T Total Sulfur Analyzer 302 Teledyne Analytical Instruments 30 minutes, it takes a certain amount of time for th
6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 303 13.5. Electronic Operation Touchscreen Display I2C Bu
Principles of Operation Model 6200T Total Sulfur Analyzer 304 Teledyne Analytical Instruments The motherboard is directly mounted to the rear pane
6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 305 13.5.1. CPU The unit’s CPU card, installed on the mother
Principles of Operation Model 6200T Total Sulfur Analyzer 306 Teledyne Analytical Instruments In the unlikely event that the flash chip should fai
6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 307 13.5.2.1. Sample Chamber The main electronic components
Principles of Operation Model 6200T Total Sulfur Analyzer 308 Teledyne Analytical Instruments 13.5.3. Photo Multiplier Tube (PMT) The 6200T uses a
6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 309 Figure 13-18: Basic PMT Design A significant perform
6200T Total Sulfur Analyzer Introduction Teledyne Analytical Instruments 31 1.6. The M501TS – Total Reduced Sulfur Converter The M501TS oxidizes
Principles of Operation Model 6200T Total Sulfur Analyzer 310 Teledyne Analytical Instruments 13.5.4. PMT Cooling System The performance of the an
6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 311 13.5.4.2. TEC Control Board The TEC control printed circ
Principles of Operation Model 6200T Total Sulfur Analyzer 312 Teledyne Analytical Instruments COOLER (Closed Loop) To Motherboard PMT Preamp
6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 313 The optical test (OTEST) feature causes an LED inside th
Principles of Operation Model 6200T Total Sulfur Analyzer 314 Teledyne Analytical Instruments 13.5.7.3. Status LEDs & Watch Dog Circuitry Zero
6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 315 As a safety measure, special circuitry on the relay boar
Principles of Operation Model 6200T Total Sulfur Analyzer 316 Teledyne Analytical Instruments This measurement is stored in the analyzer. Memory a
6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 317 first two of these outputs (A1 and A2). Refer to Section
Principles of Operation Model 6200T Total Sulfur Analyzer 318 Teledyne Analytical Instruments situation before turning the analyzer back on. AC P
6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 319 13.6. Front Panel/Display Interface Users can input data
Introduction Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 32 SAMPLEINCONVERTERTHERMALSWITCHPOWERTEMPERATURECONTROLLERTHERM
Principles of Operation Model 6200T Total Sulfur Analyzer 320 Teledyne Analytical Instruments 13.7. Software Operation The instrument’s core modul
6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 321 If necessary, these filter lengths of these two modes ma
Principles of Operation Model 6200T Total Sulfur Analyzer 322 Teledyne Analytical Instruments sample gas normalized to standard atmospheric pressu
6200T Total Sulfur Analyzer Principles of Operation Teledyen Analytical Instruments 323 This page intentionally left blank.
6200T Total Sulfur Analyzer ESD Primer Teledyen Analytical Instruments 325 14. A PRIMER ON ELECTRO-STATIC DISCHARGE Teledyne Analytical Instrume
ESD Primer Model 6200T Total Sulfur Analyzer 326 Teledyne Analytical Instruments Table 14-1: Static Generation Voltages for Typical Activities MEA
6200T Total Sulfur Analyzer ESD Primer Teledyen Analytical Instruments 327 When static charges accumulated on a sensitive device discharges f
ESD Primer Model 6200T Total Sulfur Analyzer 328 Teledyne Analytical Instruments components are prevented from building static electric charges th
6200T Total Sulfur Analyzer ESD Primer Teledyen Analytical Instruments 329 charges from building up on the device or assembly and nearby static
6200T Total Sulfur Analyzer Introduction Teledyne Analytical Instruments 33 The sample containing the converted SO2 is then handed back to the a
ESD Primer Model 6200T Total Sulfur Analyzer 330 Teledyne Analytical Instruments 3. Open the casing of the analyzer and begin work. Up to this p
6200T Total Sulfur Analyzer ESD Primer Teledyen Analytical Instruments 331 If you are at an anti-ESD workbench, lay the container down on the
ESD Primer Model 6200T Total Sulfur Analyzer 332 Teledyne Analytical Instruments Analytical Instruments’s Customer Service in anti-ESD bins, tubes
6200T Total Sulfur Analyzer Glossary Teledyen Analytical instruments 333 15. GLOSSARY Note: Some terms in this glossary may not occur elsewhere
Glossary Model 6200T Total Sulfur Analyzer 334 Teledyne Analytical Instruments Term Description/Definition O2 molecular oxygen O3 ozone SO2 sulfur
6200T Total Sulfur Analyzer Glossary Teledyen Analytical instruments 335 Term Description/Definition I2C bus a clocked, bi-directional, serial
Glossary Model 6200T Total Sulfur Analyzer 336 Teledyne Analytical Instruments Term Description/Definition PTFE Polytetrafluoroethylene, a very i
6200T Total Sulfur Analyzer Spare Parts Teledyen Analytical instruments 337 16. SPARE PARTS This section presents the Spare Parts Lists for the
Spare Parts Model 6200T Total Sulfur Analyzer 338 Teledyne Analytical Instruments PART NUMBER DESCRIPTION 040010000 ASSY, FAN REAR PANEL
6200T Total Sulfur Analyzer Spare Parts Teledyen Analytical instruments 339 PART NUMBER DESCRIPTION HW0000005 FOOT
Spare Parts Model 6200T Total Sulfur Analyzer 340 Teledyne Analytical Instruments 018080000 AKIT, DESSICANT BAGGIES, (12) 039620100
6200T Total Sulfur Analyzer Instrument Test and Calibration Record Teledyen Analytical instruments 341 17. INSTRUMENT TEST & CALIBRATION REC
Instrument Test and Calibration Record Model 6200T Total Sulfur Analyzer 342 Teledyne Analytical Instruments Table 17-2. For test and calibration in
6200T Total Sulfur Analyzer Instrument Test and Calibration Record Teledyen Analytical instruments 343 Table 17-2. Test and Calibration Values
Instrument Test and Calibration Record Model 6200T Total Sulfur Analyzer 344 Teledyne Analytical Instruments Table 17-3. Test and Calibrations Val
6200T Total Sulfur Analyzer Specifications Teledyne Analytical Instruments 35 2. SPECIFICATIONS, APPROVALS & COMPLIANCE This section present
Specifications Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 36 Parameter Description Certifications EN61326 (1997 w/A1: 98)
6200T Total Sulfur Analyzer Specifications Teledyne Analytical Instruments 37 Table 2-4: M501TS Converter Specifications Specification Value Un
Specifications Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 38 Dynamic span OFF Dynamic zero OFF Dilution factor OFF AutoC
6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 39 3. GETTING STARTED This section addresses the procedures for unpa
Configuration Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments iv Selected Options for the Model 6200T Calibrator Option: This op
Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 40 1. Remove the locking screw located in the top, center of
6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 41 Thumb drive (not included) to download updates to instruction
Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 42 The analyzer’s front panel liquid crystal display screen in
6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 43 Figure 3-3: Display/Touch Control Screen Mapped to Menu Charts
Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 44 The controller has been set up at the factory. Should furt
6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 45 Table 3-3: Rear Panel Description—Analyzer and Converter Compon
Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 46 3.2.3. Internal Chassis Layout Figure 3-5A illustrates the
6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 47 SAMPLEINCONVERTERTHERMALSWITCHPOWERTEMPERATURECONTROLLERTHERMOCO
Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 48 3.3. Connections and Setup This section presents the electr
6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 49 3.3.1.1. Connecting Power Attach the power cords to the analyzer
6200T Total Sulfur Analyzer About This Manual Teledyne Analytical Instruments v ABOUT THIS MANUAL This manual describes operation, specification
Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 50 Table 3-5: Analog Input Pin Assignments PIN DESCRIPTION DA
6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 51 3.3.1.4. Current Loop Analog Outputs (Option 41) Setup If your a
Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 52 Figure 3-8: Current Loop Option Installed on the Motherboa
6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 53 3.3.1.5. Connecting the Status Outputs The status outputs report
Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 54 Table 3-7: Status Output Signals REAR PANEL LABEL STATUS DE
6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 55 CONTROL IN A B C D E F U + S
Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 56 3.3.1.7. Connecting the Concentration Alarm Relay (Option 6
6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 57 SO2 Alarm 2 = 100 PPM SO2 Alarm 1 = 20 PPM SO2 Alarm 2 = 100 PPM
Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 58 RS-232 Connection For RS-232 communications with data term
6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 59 Figure 3-13: CPU Connector Pin-Outs for RS-232 Mode RS-232 COMM
About This Manual Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments vi This page intentionally left blank.
Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 60 DB9 connectors on the analyzer’s rear panel to connect the
6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 61 3. Close the instrument. 4. Referring to Figure 3-15 use straig
Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 62 3.3.2. Pneumatic Connections This section provides not only
6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 63 Dust Plugs: Remove dust plugs from rear panel exhaust and su
Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 64 One primary application of this instrument is to analyze CO
6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 65 Figure 3-16: Pneumatic Connections–Basic Configuration–Using Bo
Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 66 Figure 3-17: Pneumatic Connections–Basic Configuration–Us
6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 67 3.3.2.2. Pneumatic Layout for Basic Configuration SAMPLE PRESSURE
Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 68 3.3.2.3. Pneumatic Layout for Zero/Span Valves Option Figur
6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 69 remotely through the RS-232/485 serial I/O ports (refer to App
6200T Total Sulfur Analyzer Safety Messages Teledyne Analytical Instruments vii SAFETY MESSAGES Important safety messages are provided throughou
Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 70 ZERO CAL Sample/Cal Open to zero/span valve Zero/Span Ope
6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 71 3.3.2.7. TS and Zero Air Scrubbers There are two charcoal scrubbe
Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 72 3.3.2.9. Pneumatic Layout with CO2 Sensor Option Figure 3-2
6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 73 For analyzers without an IZS or external zero air scrubber option
Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 74 Span Gas for Multipoint Calibration Some applications, such
6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 75 The converter has not been tested for conversion efficiency or re
Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 76 Table includes typical values for a converter set up for o
6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 77 3.4.2. Warning Messages Because internal temperatures and other
Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 78 Table 3-13 lists brief descriptions of the warning messages
6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 79 Table 3-14: Possible Startup Warning Messages – 6200T Analyzers
Safety Messages Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments viii CONSIGNES DE SÉCURITÉ Des consignes de sécurité important
Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 80 To view the current values of these parameters press the fo
6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 81 3.4.4. Initial Calibration To perform the following calibration y
Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 82 While these are the default settings for the 6200T analyzer
6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 83 Dilution Ratio Setup If the dilution ratio option is enabled on y
Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 84 Set SO2 Span Gas Concentration Set the expected SO2 span ga
6200T Total Sulfur Analyzer Getting Started Teledyne Analytical Instruments 85 Zero/Span Calibration To perform the zero/span calibration proce
Getting Started Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 86 3.4.4.2. Calibration Procedure for the O2 Option If your an
6200T Total Sulfur Analyzer Part II Teledyne Analytical Instruments 87 PART II OPERATING INSTRUCTIONS
Part II Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 88 This page intentionally left blank.
6200T Total Sulfur Analyzer Overview of Operating Modes Teledyne Analytical Instruments 89 4. OVERVIEW OF OPERATING MODES To assist in navigatin
6200T Total Sulfur Analyzer Table of Contents Teledyne Analytical Instruments ix TABLE OF CONTENTS Selected Versions of the Model 6200T ...
Overview of Operating Modes Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 90 LO CAL A1 Unit is performing LOW SPAN (midpoin
6200T Total Sulfur Analyzer Overview of Operating Modes Teledyne Analytical Instruments 91 Table 4-2: Test Functions Defined DISPLAY PARAMETER
Overview of Operating Modes Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 92 To view the TEST Functions press the following
6200T Total Sulfur Analyzer Overview of Operating Modes Teledyne Analytical Instruments 93 4.1.2. Warning Messages The most common instrument fa
Overview of Operating Modes Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 94 To view and clear warning messages, press: SA
6200T Total Sulfur Analyzer Overview of Operating Modes Teledyne Analytical Instruments 95 4.2. Calibration Mode Pressing the CAL button switche
Overview of Operating Modes Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 96 4.3.2. Primary Setup Menu Table 4-4: Primary Se
6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 97 5. SETUP MENU The SETUP menu is used to set instrument parameters for
Setup Menu Model 6200T Total Sulfur Analyzer Teledyne Analytical Instruments 98 5.4.1. Available Analog Output Signals The analyzer has three ac
6200T Total Sulfur Analyzer Setup Menu Teledyne Analytical Instruments 99 concentration of SO2 is typically less than 500 ppb, the full scale of