Teledyne GFC7001T User Manual Page 180

  • Download
  • Add to my manuals
  • Print
  • Page
    / 326
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 179
DAC and APICONTeledyne API – Model T300/T300M CO Analyzer
Teledyne Analytical Instruments 180
7.1.11. HOLDOFF FEATURE
The DAS HOLDOFF feature prevents data collection during calibration operations and
at certain times when the quality of the analyzer’s CO measurements may not be certain
(e.g. while the instrument is warming up). In this case, the length of time that the
HOLDOFF feature is active is determined by the value of the internal variable (VARS),
DAS_HOLDOFF.
To set the length of the DAS_HOLDOFF period, go to the SETUP>MORE>VARS
menu (Section 5.8) and EDIT the “0) DAS_HOLD_OFF…” parameter.
To enable or disable the HOLDOFF feature for an individual channel, follow the
instruction shown in Section 7.1.4 then press: