Remote OperationTeledyne API – Model T300/T300M CO Analyzer
Teledyne Analytical Instruments 191
8.4. PASSWORD SECURITY FOR SERIAL REMOTE
COMMUNICATIONS
In order to provide security for remote access of the GFC7001T/GFC7001TM, a
LOGON feature can be enabled to require a password before the instrument will accept
commands. This is done by turning on the SECURITY MODE (Mode 4, Table 6-1).
Once the SECURITY MODE is enabled, the following items apply.
A password is required before the port will respond or pass on commands.
If the port is inactive for one hour, it will automatically logoff, which can also be
achieved with the LOGOFF command.
Three unsuccessful attempts to log on with an incorrect password will cause
subsequent logins to be disabled for 1 hour, even if the correct password is used.
If not logged on, the only active command is the '?' request for the help screen.
The following messages will be returned at logon:
LOGON SUCCESSFUL - Correct password given
LOGON FAILED - Password not given or incorrect
LOGOFF SUCCESSFUL - Connection terminated successfully
To log on to the GFC7001T/GFC7001TM Analyzer with SECURITY MODE feature
enabled, type:
LOGON 940331
940331 is the default password. To change the default password, use the variable RS-
232_PASS issued as follows:
V RS-232_PASS=NNNNNN
Where N is any numeral between 0 and 9.