Teledyne Analytical InstrumentsHIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORINGSYSTEM.PERSONAL PROTECTIVE EQUIPMENT M
Span gas: An inert gas (nitrogen or argon) with a known trace amount of oxygen. The oxygen content should be 70 – 90% of the full scale on the range o
2) Establish a flowrate in the sample line of from 0.2 to 5 liters/min. Allow the sample to flow long enough to allow a stable low reading to be achi
that an overnight recharge be accomplished every three (3) weeks of continuous use. To recharge the batteries, place the range switch in the “OFF” po
The spare cell should be carefully stored in an area that is not subject to large variations in ambient temperature (75 Deg. F nominal), and in such a
5. TRANSDUCER AND TEMPERATURE COMPENSATION The Micro-Fuel Cell has an inherent positive temperature coefficient, the effects of which have been mini
-11- 7. PRODUCT SPECIFICATION DATA RANGES: 0-2, 10, 100, 1000 PPM Oxygen and Cal range for air calibrat
Accuracy at 25oC and constant pressure after thermal equilibrium has been reached. Basic accuracy +/- 2% of full scale for all ranges except 0-2 PPM r
1 F-51 MICRO-FUSE, 0.5 AMP., FAST-BLOW (FOR 100-115 VAC) 1 C6689 -B2CXL MICRO-FUEL CELL, CLASS B-2CXL 2 B-9905 BATTERY 1 L-79 LAMP 1
C-70246 SCHEMATIC D-70268 OUTLINE C-14627 WIRING DIAGRAM (110 VAC) C-10470 WIRING DIAGRAM (220VAC) DRAWING LIST MODEL 31
INSTRUCTION MANUAL FOR TRACE OXYGEN ANALYZER MODEL 311XL TELEDYNE ANALYTICAL INSTRUMENTS 16830 CHESTNUT STREET CITY OF INDUSTRY, CA 91749
WARNING The sensor(s) used in this instrument uses electrolytes which contain substances that are extremely harmful if touched, swallowed, or inha
TABLE OF CONTENTS 311XL SERIES 1. INTRODUCTION 1 1.1 Description 1 1.2 Method of Analysis 1 1.3 Outstanding Features 1
311XL SERIES TRACE OXYGEN ANALYZER 1. INTRODUCTION 1.1 Description: The Teledyne Analytical Instruments (TAI) Model 311XL and 311TCXL are porta
1.3.2 Reliable Calibration. Although this instrument can be calibrated with air, the use of a calibration gas gives better results on the lower P
A current limiting resistor (R1 and R2) is potted into the end of each battery. This assures that under no circumstances can more than 25 milliampere
2. SUPPORTING EQUIPMENT AND SERVICES 2.1 Sampling Equipment. The customer must provide a means of controlling the pressure and flowrate of the
3.1 Introduction. The Model 311XL is supplied completely assembled and ready for instant use. The Micro-Fuel Cell must be installed prior use. W