Trace Oxygen Analyzer
Teledyne Analytical Instruments i
OPERATING INSTRUCTIONS FOR
MODEL 3000TA-XL-EU
Trace Oxygen Analyzer
DANGER
Toxic gases and or flammable liquids may be present in this monitoring system.
Personal protective equipment may be required when servicing this instrument.
Hazardous voltages exist on certain components internally which may persist
for a time even after the power is turned off and disconnected.
Only authorized personnel should conduct maintenance and/or servicing.
Before conducting any maintenance or servicing, consult with authorized
supervisor/manager.
P/N M69603
12/02/13
Page view 0
1 2 3 4 5 6 ... 107 108

Summary of Contents

Page 1 - MODEL 3000TA-XL-EU

Trace Oxygen Analyzer Teledyne Analytical Instruments i OPERATING INSTRUCTIONS FOR MODEL 3000TA-XL-EU Trace Oxygen Analyzer DANGER Toxic gases

Page 2 - 3000TA-XL-EU

3000TA-XL-EU Teledyne Analytical Instruments x 3.3.2.2.1 Analog Outputs 233.3.2.2.2 Alarm Relays 243.3.2.2.3 Digital Remote Cal Inputs 263

Page 3 - Trace Oxygen Analyzer

Appendix 3000TA-XL-EU Teledyne Analytical Instruments 84 Power: Universal power supply 85-250 VAC, at 47-63 Hz. Operating Temperature: 5-35°C

Page 4 - Specific Model Information

Trace Oxygen Analyzer Appendix Teledyne Analytical Instruments 85 A-2 Recommended 2-Year Spare Parts List Qty. Part Number Description 1 C65507A Ba

Page 5 - Important Notice

Appendix 3000TA-XL-EU Teledyne Analytical Instruments 86 A-3 Drawing List D-69604 Outline Drawing A-4 19-inch Relay Rack Panel Mount Figure A-1: S

Page 6 - Safety Messages

Trace Oxygen Analyzer Appendix Teledyne Analytical Instruments 87 A.5 Application notes 3000XL SERIES ANALYZERS APPLICATION NOTES ON PRESSURES AN

Page 7

Appendix 3000TA-XL-EU Teledyne Analytical Instruments 88 damage may occur. For cell pressures above atmospheric, caution must be exercised to avoi

Page 8

Trace Oxygen Analyzer Appendix Teledyne Analytical Instruments 89 Material Safety Data Sheet Section I - Product Identification Product Name:Micro-

Page 9 - Table of Contents

Appendix 3000TA-XL-EU Teledyne Analytical Instruments 90 Section III -Physical Hazards Potential for fire and explosion: The electrolyte in the Mi

Page 10

Trace Oxygen Analyzer Appendix Teledyne Analytical Instruments 91 Section V - Emergency and First Aid Procedures Eye Contact:Flush eyes with water

Page 11

Appendix 3000TA-XL-EU Teledyne Analytical Instruments 92

Page 12

Trace Oxygen Analyzer Teledyne Analytical Instruments xi 4.4.2.1 Auto Mode Spanning 514.4.2.2 Manual Mode Spanning 524.4.3 Span Failure 534.

Page 13 - List of Figures

3000TA-XL-EU Teledyne Analytical Instruments xii 6.5 Major Internal Components 806.6 Cleaning 816.7 Troubleshooting 81Appendix ...

Page 14

Trace Oxygen Analyzer Teledyne Analytical Instruments xiii List of Figures Figure 1-1: Model 3000TA-XL-EU Front Panel 3Figure 1-2: Model 3000TA

Page 15 - List of Tables

3000TA-XL-EU Teledyne Analytical Instruments xiv Figure 5-5: Set Range Window 69Figure 5-6: System Window 70Figure 5-7: Self-Test Window 71

Page 16

Trace Oxygen Analyzer Teledyne Analytical Instruments xv List of Tables Table 3-1: Analog Output Connections 24Table 3-2: Alarm Relay Contact P

Page 18 - 1.4 Model Designations

Trace Oxygen Analyzer Introduction Teledyne Analytical Instruments 1 Introduction 1.1 Overview The Teledyne Analytical Instruments Model 3000TA-XL-

Page 19

Introduction 3000TA-XL-EU Teledyne Analytical Instruments 2  Advanced Micro-Fuel Cell, designed for trace analysis, has a 0-1 ppm low range wit

Page 20

Trace Oxygen Analyzer Introduction Teledyne Analytical Instruments 3 1.5 Front Panel (Operator Interface) The standard 3000TA-XL-EU is housed in a

Page 21

3000TA-XL-EU Teledyne Analytical Instruments ii Copyright © 2013 Teledyne Analytical Instruments All Rights Reserved. No part of this manual may

Page 22

Introduction 3000TA-XL-EU Teledyne Analytical Instruments 4 Data Entry Keys: Six touch-sensitive membrane switches are used to input data to the

Page 23 - Operational Theory

Trace Oxygen Analyzer Introduction Teledyne Analytical Instruments 5 requires unfastening rear panel screws and sliding the unit out of the case.

Page 24

Introduction 3000TA-XL-EU Teledyne Analytical Instruments 6 Note: If you require highly accurate Auto-Cal timing, use external Auto-Cal control

Page 25

Trace Oxygen Analyzer Operational Theory Teledyne Analytical Instruments 7 Operational Theory 2.1 Introduction The analyzer is composed of three su

Page 26

Operational Theory 3000TA-XL-EU Teledyne Analytical Instruments 8 between a battery and a true fuel cell. (All of the reactants are stored exter

Page 27

Trace Oxygen Analyzer Operational Theory Teledyne Analytical Instruments 9 Figure 2-2: Cross Section of a Micro-fuel Cell (not to scale) The anode

Page 28 - 2.3 Sample System

Operational Theory 3000TA-XL-EU Teledyne Analytical Instruments 10 When the oxygen is reduced at the cathode, lead is simultaneously oxidized at

Page 29

Trace Oxygen Analyzer Operational Theory Teledyne Analytical Instruments 11 directly proportional to the concentration of oxygen in the gaseous mix

Page 30

Operational Theory 3000TA-XL-EU Teledyne Analytical Instruments 12 If, however, a low range accuracy better than 100 ppb is required, then a zer

Page 31

Trace Oxygen Analyzer Operational Theory Teledyne Analytical Instruments 13 Figure 2-4: Piping Layout

Page 32

Trace Oxygen Analyzer Teledyne Analytical Instruments iii

Page 33

Operational Theory 3000TA-XL-EU Teledyne Analytical Instruments 14 Figure 2-5A: Flow Diagram—Sample Under Pressure-Standard Figure 2-5B: Flow

Page 34

Trace Oxygen Analyzer Operational Theory Teledyne Analytical Instruments 15 2.4 Electronics and Signal Processing The Model 3000TA-XL-EU Trace Oxyg

Page 35 - Installation

Operational Theory 3000TA-XL-EU Teledyne Analytical Instruments 16 Figure 2-7: Block Diagram of the Model 3000TA-XL-EU Electronics

Page 36

Trace Oxygen Analyzer Operational Theory Teledyne Analytical Instruments 17 In the presence of oxygen the cell generates a current. A current to vo

Page 37 - 3.3 Rear Panel Connections

Operational Theory 3000TA-XL-EU Teledyne Analytical Instruments 18

Page 38

Trace Oxygen Analyzer Installation Teledyne Analytical Instruments 19 Installation Installation of the Model 3000TA-XL-EU Analyzer includes: 1. Un

Page 39 - PRIMARY INPUT POWER

Installation 3000TA-XL-EU Teledyne Analytical Instruments 20 Figure 3-1: Front Panel of the Model 3000TA-XL-EU All operator controls, except t

Page 40

Trace Oxygen Analyzer Installation Teledyne Analytical Instruments 21 3.3 Rear Panel Connections Figure 3-3 shows the Model 3000TA-XL-EU rear panel

Page 41

Installation 3000TA-XL-EU Teledyne Analytical Instruments 22 SAMPLE IN: In the standard model, gas connections are made at the SAMPLE IN and EXHA

Page 42

Trace Oxygen Analyzer Installation Teledyne Analytical Instruments 23 3.3.2.1 PRIMARY INPUT POWER The power cord receptacle and fuse block are loca

Page 43 - 41 Cal Contact

3000TA-XL-EU Teledyne Analytical Instruments iv Specific Model Information The instrument for which this manual was supplied may incorporate one

Page 44

Installation 3000TA-XL-EU Teledyne Analytical Instruments 24  0–1 V DC % of Range: Voltage rises linearly with increasing oxygen, from 0 V at

Page 45

Trace Oxygen Analyzer Installation Teledyne Analytical Instruments 25  Threshold Alarm 1:  Can be configured as high (actuates when concent

Page 46 - RS-232 PORT

Installation 3000TA-XL-EU Teledyne Analytical Instruments 26 43 Threshold Alarm 2, normally open contact 36 System Alarm, normally closed c

Page 47

Trace Oxygen Analyzer Installation Teledyne Analytical Instruments 27 40 Cal Contact 41 Cal Contact 3.3.2.2.5 Remote Calibration Protocol To pr

Page 48 - INSTALLATION

Installation 3000TA-XL-EU Teledyne Analytical Instruments 28 1 ID, Medium range is assigned to Range 2 ID, and High range is assigned to Range 3

Page 49

Trace Oxygen Analyzer Installation Teledyne Analytical Instruments 29 Figure 3-5: Remote Probe Connections The voltage from these outputs is nomina

Page 50 - 3.5 Testing the System

Installation 3000TA-XL-EU Teledyne Analytical Instruments 30 3.3.2.3 RS-232 PORT The digital signal output is a standard, full duplex RS-232 seri

Page 51

Trace Oxygen Analyzer Installation Teledyne Analytical Instruments 31 Byte 8 bits Parity none Stop Bits 1 Message Interval 2 seconds.

Page 52 - Installation 3000TA-XL-EU

Installation 3000TA-XL-EU Teledyne Analytical Instruments 32 MINIMUM AMOUNT OF TIME, INCLUDING DURING INSTALLATION. Use care to minimize the ex

Page 53 - Operation

Trace Oxygen Analyzer Installation Teledyne Analytical Instruments 33 The following procedure can be used to install the Insta-Trace Micro-fuel Cel

Page 54

Trace Oxygen Analyzer Teledyne Analytical Instruments v Important Notice Model 3000TA-XL-EU complies with all of the requirements of the Commonwea

Page 55

Installation 3000TA-XL-EU Teledyne Analytical Instruments 34 9. Purge the system using sample or zero gas. 10. Power the system back up. The Ins

Page 56 - 4.3 The System Function

Trace Oxygen Analyzer Installation Teledyne Analytical Instruments 35  Check that inlet sample pressure is within the accepted range (see Section

Page 57

Installation 3000TA-XL-EU Teledyne Analytical Instruments 36

Page 58

Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 37 Operation 4.1 Introduction Once the analyzer has been installed, it can be confi

Page 59 - ENTERING THE PASSWORD

Operation 3000TA-XL-EU Teledyne Analytical Instruments 38 4.2 Using the Data Entry and Function Buttons Data Entry Buttons: The ◄► arrow buttons

Page 60 - PSWD, and

Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 39 Figure 4-1: Hierarchy of Available Functions Any function can be selected at an

Page 61 - arrow keys to

Operation 3000TA-XL-EU Teledyne Analytical Instruments 40 4.3 The System Function The sub functions of the System function are described below. S

Page 62

Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 41 4.3.1 Tracking the Oxygen Readings During Calibration and Alarm Delay The user

Page 63

Operation 3000TA-XL-EU Teledyne Analytical Instruments 42 The factory default is three minutes, but the delay time is programmable. To adjust to

Page 64 - Version

Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 43 Press ◄► arrows to blink Span (or Zero), then press Enter again. (You won’t be a

Page 65 - AUTO MODE ZEROING

3000TA-XL-EU Teledyne Analytical Instruments vi Safety Messages Your safety and the safety of others is very important. We have provided many im

Page 66 - MANUAL MODE ZEROING

Operation 3000TA-XL-EU Teledyne Analytical Instruments 44 Use the ◄► arrow keys to scroll the blinking over to PSWD, and press Enter to select th

Page 67 - 4.4.2.1 AUTO MODE SPANNING

Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 45 Press Enter to change the password (either the default TETAI or the previously a

Page 68 - MANUAL MODE SPANNING

Operation 3000TA-XL-EU Teledyne Analytical Instruments 46 Wait a moment for the entry screen. You will be given clearance to proceed. A A A

Page 69

Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 47 4.3.5 System Self-Diagnostic Test The Model 3000TA-XL-EU has a built-in self-dia

Page 70

Operation 3000TA-XL-EU Teledyne Analytical Instruments 48 4.3.6 Version Screen Move the ◄► arrow key to More and press Enter. With Version blink

Page 71 - 4.6 The Alarms Function

Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 49 4.4.1 Zero Cal The B-2CXL cell has a zero offset of less than 0.2 ppm oxygen. N

Page 72 - Dft—N Fs—N Ltch—N

Operation 3000TA-XL-EU Teledyne Analytical Instruments 50 #### PPM Zero Slope=#### ppm/s The beginning zero level is shown in the uppe

Page 73 - 4.7 The Range Function

Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 51 Generally, you have a good zero when Slope is less than 0.05 ppm/s for about 30

Page 74 - H—100 Mode—AUTO

Operation 3000TA-XL-EU Teledyne Analytical Instruments 52 Span: Settling: AUTO <ENT> For Next Press Enter to move to the next screen. Calib

Page 75 - 4.8 The Analyze Function

Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 53 Calib. Holding time Cal hold: 5 min This menu allows the operator to set the t

Page 76 - 4.9 Signal Output

Trace Oxygen Analyzer Teledyne Analytical Instruments vii CAUTION: THE ANALYZER SHOULD ONLY BE USED FOR THE PURPOSE AND IN THE MANNER DESCRIBED I

Page 77 - 4.10 The IP Address Function

Operation 3000TA-XL-EU Teledyne Analytical Instruments 54 VFD display “FCAL” will be shown. This message flag will help the operator troubleshoo

Page 78

Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 55 4.5.1 Special Notes on Hydrogen Gas Stream The standard oxygen cells, Model B-2C

Page 79 - Web Implementation

Operation 3000TA-XL-EU Teledyne Analytical Instruments 56 3. Are either or both of the alarms to be configured as failsafe? You can set either o

Page 80

Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 57  Value of the alarm setpoint, AL–1 #### ppm (oxygen); this value can be set

Page 81 - 5.2 Remote Calibration

Operation 3000TA-XL-EU Teledyne Analytical Instruments 58 below 85% of full scale of the next lower range, the instrument will switch to that ran

Page 82

Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 59 4.7.2 Fixed Range Analysis The autoranging mode of the instrument can be overrid

Page 83

Operation 3000TA-XL-EU Teledyne Analytical Instruments 60 4.9 Signal Output The standard Model 3000TA-XL-EU Trace Oxygen Analyzer is equipped wit

Page 84 - 5.3 Set the Alarms Remotely

Trace Oxygen Analyzer Operation Teledyne Analytical Instruments 61 current when using the current outputs) to represent a particular range. The fol

Page 85 - 5.4 Set the Ranges Remotely

Operation 3000TA-XL-EU Teledyne Analytical Instruments 62 Ethernet web server coprocessor. To set the IP address, enter the system menu: Ver S

Page 86 - 5.5 System Menu

Trace Oxygen Analyzer Web Implementation Teledyne Analytical Instruments 63 Web Implementation 5.1 The Embedded Browser The 3000TA-XL-I series inst

Page 87 - 5.6 The Self-Test function

3000TA-XL-EU Teledyne Analytical Instruments viii This is a general purpose instrument designed for use in a nonhazardous area. It is the cust

Page 88

Web Implementation 3000TA-XL-EU Teledyne Analytical Instruments 64 Figure 5-1: Analyzer Status Window—Running This screen shows the relevant s

Page 89 - Maintenance

Trace Oxygen Analyzer Web Implementation Teledyne Analytical Instruments 65  System  Self Test These functions are described in the following s

Page 90

Web Implementation 3000TA-XL-EU Teledyne Analytical Instruments 66 Figure 5-2: Span Calibration Window

Page 91

Trace Oxygen Analyzer Web Implementation Teledyne Analytical Instruments 67 5.2.2 Remote Zero To perform a remote zero, click the Zero button at th

Page 92 - Maintenance 3000TA-XL-EU

Web Implementation 3000TA-XL-EU Teledyne Analytical Instruments 68 5.3 Set the Alarms Remotely To set the concentration alarms, Alarm 1 and Ala

Page 93

Trace Oxygen Analyzer Web Implementation Teledyne Analytical Instruments 69 5.4 Set the Ranges Remotely With the 3000TA-XL-EU-I, it is possible to

Page 94 - 6.3 Fuse Replacement

Web Implementation 3000TA-XL-EU Teledyne Analytical Instruments 70 5.5 System Menu By clicking the System button some settings found in the sys

Page 95

Trace Oxygen Analyzer Web Implementation Teledyne Analytical Instruments 71 5.6 The Self-Test function Clicking on the Self Test button will initia

Page 96

Web Implementation 3000TA-XL-EU Teledyne Analytical Instruments 72

Page 97 - 6.7 Troubleshooting

Trace Oxygen Analyzer Maintenance Teledyne Analytical Instruments 73 Maintenance 6.1 Routine Maintenance Aside from normal cleaning and checking fo

Page 98 - Possible Cause:

Trace Oxygen Analyzer Teledyne Analytical Instruments ix Table of Contents Safety Messages ...

Page 99 - Appendix

Maintenance 3000TA-XL-EU Teledyne Analytical Instruments 74 The spare cell should be carefully stored in an area that is not subject to large var

Page 100

Trace Oxygen Analyzer Maintenance Teledyne Analytical Instruments 75 6.2.3 Removing the Micro-Fuel Cell The Micro-Fuel cell is located inside the s

Page 101

Maintenance 3000TA-XL-EU Teledyne Analytical Instruments 76 Figure 6-1: Removing the Micro-Fuel

Page 102 - A-3 Drawing List

Trace Oxygen Analyzer Maintenance Teledyne Analytical Instruments 77 Figure 6-2: Installing the Insta-Trace Cell 6.2.5 Cell Warranty The Class B-

Page 103 - A.5 Application notes

Maintenance 3000TA-XL-EU Teledyne Analytical Instruments 78 Consult TAI for the appropriate cell for measuring trace oxygen in the presence of CO

Page 104

Trace Oxygen Analyzer Maintenance Teledyne Analytical Instruments 79 3. Replace fuse as shown in Figure 6-4. 4. Reassemble Housing as shown in Fi

Page 105 - Material Safety Data Sheet

Maintenance 3000TA-XL-EU Teledyne Analytical Instruments 80 Preamp 0 OK 1 Zero too high 2 Amplifier output doesn't match test

Page 106

Trace Oxygen Analyzer Maintenance Teledyne Analytical Instruments 81 The 3000TA-XL contains the following major components:  Analysis Section 

Page 107

Maintenance 3000TA-XL-EU Teledyne Analytical Instruments 82 Solution: Turn the analyzer off, then turn back on again. Press the System key whe

Page 108 - Appendix 3000TA-XL-EU

Trace Oxygen Analyzer Appendix Teledyne Analytical Instruments 83 Appendix A-1 Specifications Packaging: General Purpose • Flush panel moun