INSTRUCTION MANUAL MODEL 200AH NITROGEN OXIDES ANALYZER ©TELEDYNE INSTRUMENTS ADVANCED POLLUTION INSTRUMENTATION DIVISION (T-API) 9480 CA
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 viii LIST OF TABLES TABLE 2-1: FINAL TEST AND CALIBRATION VALUES ...
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 7-8 Table 7-7: Manual Zero Check Procedure - Z/S Valves Step Number Actio
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 7-9 Table 7-8: Manual Span Check Procedure - Z/S Valves Step Number Actio
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 7-10 7.5. Use of Zero/Span Valves with Remote Contact Closure The Zero/Sp
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 7-11 7.6. NO Only Mode Calibration For details on the operation of this m
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 7-12 Table 7-10: Calibration Requirements for AutoRange or Remote Range No
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 7-13 Table 7-11: Calibration Quality Check Step Number Action Comment 1.
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 7-14 Table 7-12: Converter Efficiency - Automatic Calculation Step Number
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 7-15 NOTE It is important to note that if you are using blended calibratio
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 7-16 THIS PAGE IS INTENTIONALLY LEFT BLANK
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 8-1 8. MAINTENANCE 8.1. Maintenance Schedule NOTE The operations outlin
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 ix TABLE 9-4: DIAGNOSTIC MODE - SIGNAL I/O ...
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 8-2 8.2. Replacing the Sample Particulate Filter The particulate filter s
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 8-3 Figure 8-1: Replacing the Particulate Filter
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 8-4 8.3. Replacing the Converter The Converter is located in the center o
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 8-5 Figure 8-2: Converter Assembly
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 8-6 8.4. Cleaning the Reaction Cell The reaction cell should be cleaned w
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 8-7 10. Reassemble in the reverse order of disassembly, being sure to seat
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 8-8 Figure 8-4: Sample/Bypass Flow Control Assembly
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 8-9 8.5. Pneumatic Line Inspection Particulate matter in the pneumatic li
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 8-10 Figure 8-5: Pneumatic Diagram
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 8-11 Figure 8-6: Pneumatic Diagram with Zero/Span Valves
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 x THIS PAGE IS INTENTIONALLY LEFT BLANK
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 8-12 Figure 8-7: Pneumatic Diagram with Oxygen Sensor
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 8-13 8.6. Leak Check Procedure If a leak checker is not available, it is
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 8-14 8.8. Prom Replacement Procedure Preparation: If any setup changes su
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-1 9. TROUBLESHOOTING AND ADJUSTMENTS NOTE The operations outlined in t
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-2 4. Pay special attention to the slopes, the slopes are the software eq
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-3 The Table 9-1 indicates possible fault conditions that could cause the
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-4 Table 9-1: Test Functions (Continued) Test Function Factory Set-Up C
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-5 Table 9-1: Test Functions (Continued) Test Function Factory Set-Up C
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-6 Table 9-1: Test Functions (Continued) Test Function Factory Set-Up C
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-7 Table 9-1: Test Functions (Continued) Test Function Factory Set-Up C
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 1-1 1. HOW TO USE THIS MANUAL The Model 200AH has been designed to produc
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-8 9.1.2. Fault Diagnosis with WARNING Messages The M200AH monitors seve
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-9 Table 9-2: Front Panel Warning Messages Message Description SYSTEM RE
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-10 Table 9-2: Front Panel Warning Messages (Continued) Message Descript
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-11 Table 9-3: Summary of Diagnostic Modes DIAG Mode Description SIGNAL
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-12 9.1.3.1. Signal I/O Diagnostics Table 9-4: Diagnostic Mode - Signal
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-13 Table 9-4: Diagnostic Mode - Signal I/O (Continued) No. Signal Cont
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-14 Table 9-4: Diagnostic Mode - Signal I/O (Continued) No. Signal Cont
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-15 Table 9-4: Diagnostic Mode - Signal I/O (Continued) No. Signal Cont
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-16 9.1.3.2. Electric Test This function injects a constant voltage betw
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-17 9.1.3.4. Ozone Gen Power This diagnostic manually turns the power of
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 1-2 THIS PAGE IS INTENTIONALLY LEFT BLANK
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-18 Table 9-5: Model 200AH Variables No. Name Units Default Value Value
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-19 9.1.5. Test Channel Analog Output Many of the TEST functions have an
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-20 Table 9-6: Test Channel Readings (Continued) TEST Channel Minimum*
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-21 PMT Calibration Procedure: NOTE In this procedure a range of 100 ppm a
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-22 10. Adjust S2, the HVPS coarse adjustment, on the preamp board to the
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-23 Figure 9-1: Span Calibration Voltage
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-24 9.2. Performance Problems When the response from a span check is out
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-25 9.2.2. Flow Check 1. Check TEST function RCEL - this is the absolut
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-26 2. Check instrument electronics. A. Do ELEC TEST procedure in DIAGNO
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-27 3. HVPS noisy - see HVPS test procedure. See Section 9.3.8.5. 4. Def
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 2-1 2. GETTING STARTED 2.1. Unpacking CAUTION Your safety and the safet
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-28 9.2.7. Unstable Zero Common causes are: 1. Leak in pneumatic system
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-29 A. A leak that admits gas containing NO. B. Contaminated reaction ce
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-30 7. Ozone concentration too low: A. Wet air in generator - need to di
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-31 9.3. Subsystem Troubleshooting and Adjustments 9.3.1. Computer, Di
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-32 Testing and operation of the CPU RS-232 port is described in Section 9
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-33 Figure 9-2: CPU Board Jumper Settings
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-34 9.3.1.3. Front Panel Keyboard The keyboard consists of 8 keys and 3
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-35 Figure 9-3: RS-232 Pin Assignments
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-36 9.3.2.1. RS-232 Port If the RS-232 port is not working, check the fo
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-37 Check that the voltage level on Pin 8 of the Analyzer is between +5 V
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 2-2 WARNING Analyzer Exhaust – O3 Scrubber – Pump Pack Danger – Analyzer
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-38 A. Turn on the instrument. B. Enter the SETUP-MORE-DIAG menu. C. Se
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-39 9.3.3.1. ADC/DAC Calibration Procedure Due to the stability of moder
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-40 10 V output or non-isolated current loop Switch State Comment 3 ON
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-41 C. The M200AH will display a voltage near 1% of the voltage range set
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-42 Figure 9-4: V/F Board Jumper Settings
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-43 9.3.4. Status/Temp Board The Status/Temp Board is a multifunction b
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-44 The thermocouple amplifier for the converter temperature has a gain ad
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-45 9.3.4.4. 4-20 mA Current Output There are two types of current loop
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-46 Table 9-9: Power Supply Module Subassemblies Module Description Line
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-47 Figure 9-5: Power Supply Module Layout
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 2-3 Figure 2-1: Removal of Shipping Screws & Check for Correct Power
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-48 Figure 9-6: Electrical Block Diagram
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-49 PSM Diagnostic procedures: The Linear Power Supply board can be tested
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-50 Common faults in the ozone generator are: 1. A leak or some other fai
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-51 9.3.6.1. The PermaPure Drier The PermaPure drier is constructed of 2
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-52 Figure 9-7: Ozone Generator Subsystem
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-53 9.3.7. Flow/Pressure Sensor The flow/pressure sensor board consists
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-54 To adjust the SAMPLE PRESSURE: 1. Go to DIAG mode by pressing SETUP-M
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-55 Figure 9-8: Flow/Pressure Sensor
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-56 Figure 9-9: NOX Sensor Module
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-57 Figure 9-10: NOX Sensor Module
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 2-4 Figure 2-2: Rear Panel
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-58 9.3.8. NOX Sensor Module Figure 9-9 and Figure 9-10 show the assembl
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-59 9.3.8.4. PMT Cooler The PMT cooler uses a Peltier cooler supplied wi
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-60 Figure 9-11: PMT Cooler Subsystem
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-61 9.3.8.5. High Voltage P.S. The HVPS is located in the interior of th
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-62 Figure 9-12: High Voltage Power Supply
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-63 9.3.9. Z/S Valves The Z/S Valves is an option in the instrument. Be
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 9-64 It is also possible to manually toggle each of the valves in the DIAGN
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 A-1 APPENDIX A ELECTRICAL SCHEMATICS Table A-1: Electrical Schematics P
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 A-2 Table A-1: Electrical Schematics (Continued) Part No. Name 02034 A-S
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 B-1 APPENDIX B SPARE PARTS AND EXPENDABLES NOTE Use of replacement part
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 2-5 Figure 2-3: Rear Panel with O2 Option
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 B-2 THIS PAGE IS INTENTIONALLY LEFT BLANK
SPARE PARTS LISTM200AHPart Number Description000940100 Orifice, 3 mil, 60 cc (IZS)000940500 Orifice, 7 mil, 250 cc (84
SPARE PARTS LISTM200AH017130000 Model 200AH Level 1 Spare Parts Kit (for 10 units) with MINI-HICON017130100 Model 200A
LEVEL 1 SPARES KIT(WITH MINI-HICON CONVERTER)M200AHPart Number Description000940100 CD, ORIFICE, .003 GREEN 000940500
LEVEL 1 SPARES KIT(WITH MOLYBDENUM CONVERTER)M200AHPart Number Description000940100 CD, ORIFICE, .003 GREEN 000940500
EXPENDABLES KITM200AHPart Number Description002270100 AKIT, GASKETS, WINDOW, (12) 009690300 AKIT, TFE FLTR
THIS PAGE IS INTENTIONALLY LEFT BLANK
THIS PAGE IS INTENTIONALLY LEFT BLANK
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 2-6 Figure 2-4: Inlet and Exhaust Venting Recommendations
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 2-7 2.3. Initial Operation 1. Turn on the instrument power. 2. The dis
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 2-8 7. When the instrument is warmed up, re-check the TEST functions again
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 2-9 Step 2 - Calibrate the instrument for NOX: Zero/Span Calibration Proced
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 2-10 Zero/Span Calibration Procedure (Continued) Step For Units w/o O2 Opt
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 2-11 Step 3 - Review Quality of calibration: Calibration Quality Check Proc
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 2-12 Figure 2-5: Front Panel
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 2-13 Figure 2-6: Assembly Layout
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 2-14 Figure 2-7: Assembly Layout – O2 Sensor Option
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 2-15 Table 2-1: Final Test and Calibration Values TEST Values Observed Val
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 i SAFETY MESSAGES Your safety and the safety of others is very important. We
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 2-16 Table 2-1: Final Test and Calibration Values (Continued) Parameter Ob
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 3-1 3. SPECIFICATIONS, WARRANTY 3.1. Specifications NOX Specificatio
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 3-2 3.2. Warranty WARRANTY POLICY (02024c) Prior to shipment, Teledyne AP
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 4-1 4. THE M200AH NOX ANALYZER 4.1. Principle of Operation 4.1.1. NOX
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 4-2 concentration is measured and stored internally. The valve is then swit
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 4-3 Figure 4-1: Paramagnetic Sensor Refer to Figure 4-1. The measuring s
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 4-4 Figure 4-2: Block Diagram
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 4-5 4.2. Operation Summary 4.2.1. NOX Sensor Module, Reaction Cell, Det
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 4-6 4.2.3. Pneumatic Sensor Board The sensor board consists of 2 pressure
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 4-7 4.2.5. V/F Board The computer CPU board communicates via 2 major hard
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 ii THIS PAGE IS INTENTIONALLY LEFT BLANK
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 4-8 Table 4-2: Front Panel Status LED's LED State Meaning Green Off
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 4-9 An external pump comes as standard equipment. The Pump Pack includes a
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 4-10 Figure 4-3: External Pump Pack
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 4-11 4.2.9. Ozone Generator Because of the instability of ozone, it is ne
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 4-12 THIS PAGE IS INTENTIONALLY LEFT BLANK
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-1 5. SOFTWARE FEATURES The M200AH control software has two major operat
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-2 Figure 5-1: Sample Menu Tree
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-3 Figure 5-2: Setup Menu Tree
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-4 Figure 5-3: Setup Menu Tree- Units with O2 Sensor
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-5 Figure 5-4: Setup Menu Tree – Units with O2 Sensor
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 iii Table of Contents SAFETY MESSAGES ...
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-6 5.1.1. Sample Menu Table 5-1: M200AH Sample Menu Structure Level 1 L
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-7 5.1.2. Set-Up Menu Table 5-2: M200AH Setup Menu Level 1 Level 2 L
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-8 Table 5-2: M200AH Setup Menu (Continued) Level 1 Level 2 Level 3 Le
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-9 Table 5-3: M200AH Setup Menu Level 1 Level 2 Level 3 Level 4 Level
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-10 Table 5-4: M200AH Setup Menu Level 1 Level 2 Level 3 Level 4 Leve
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-11 5.2. Sample Mode 5.2.1. Test Functions NOTE In any of the followin
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-12 Signal Stability (STABIL, NOX STB) The instrument noise is computed by
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-13 DC Power Supply Voltage (DCPS) The DCPS voltage is a composite of the
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-14 If the analyzer is equipped with the molybdenum converter option the T
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-15 5.2.2. CAL, CALS, CALZ, Calibration Functions The calibration and ze
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 iv 5.3.4. Range Menu...
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-16 5.2.2.5. NO, NOX and O2 Cal Concentration Before the M200AH can be s
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-17 Because the span calibration calculation includes zero calibration dat
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-18 5.3.3. Data Acquisition System (DAS) The M200A contains a flexible a
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-19 The function of each of the default Data Channels is described below:
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-20 5.3.4. Range Menu For NOX measurements, the instrument operates on a
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-21 To set up Auto Range press SETUP-RNGE-MODE-AUTO, then press ENTR. To s
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-22 3. SELECT RANGE The range selection is the same with dilution as wit
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-23 5.3.4.7. Recorder Offset If necessary, a recorder offset can be intr
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-24 5.3.8. Communications Menu The COMM menu allows the RS-232 BAUD rate
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-25 5.4. M200AH Operating Modes The M200AH has two main operating modes.
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 v 7.12.1. Calibration Gasses ...
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-26 This is the most common mode of operation. It provides continuous outp
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-27 5.5. 4-20 mA Current Loop The analog signal output is available on
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-28 5.7. RS-232 Interface The RS-232 communications protocol allows the
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-29 Table 5-9: RS-232 Port Setup - Front Panel Decimal Value Description
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-30 940331 is the default password. The password can be changed to any num
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-31 Table 5-11: RS-232 Terminal Mode Editing Keys Key Function CR (carri
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-32 Table 5-12: RS-232 Command Summary Commands Definition ? Print help
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-33 Table 5-13: RS-232 Command Summary Terminal Mode Editing Keys Defini
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-34 Table 5-14: RS-232 Interface Command Types First Character Message T
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-35 5.7.3. TEST Commands and Messages Table 5-15: RS-232 Test Messages
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 vi APPENDIX A ELECTRICAL SCHEMATICS ...
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-36 The T command lists TEST messages. Examples of the T command are: T LI
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-37 Whenever a warning message is reported on the analyzer display, if the
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-38 Whenever the analyzer starts or finishes a calibration, it issues a st
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-39 The following is a summary of the Diagnostic commands. Table 5-19: RS
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-40 To print records from a DAS data channel enter: D REPORT “name” RECORD
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-41 All of the default Data Channels sample more than one parameter, for t
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-42 Where: name = name of the variable value = current value of variable
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 6-1 6. OPTIONAL HARDWARE AND SOFTWARE Optional equipment offered with the
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 6-2 6.2. Zero/Span Valves Option The Zero/Span Valve option consists of a
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 6-3 6.2.1. Autocal - Setup Zero/Span Valves The Zero/Span valves can be s
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 vii LIST OF FIGURES FIGURE 2-1: REMOVAL OF SHIPPING SCREWS & CHECK FOR
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 6-4 Table 6-3: Action of MODE Field in AutoCal MODE CALIBRATE = ON CALIB
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 6-5 6.4. Isolated 4-20 mA Current Loop Option The M200AH comes equipped w
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 6-6 The M501 contains 2 converters, a high temperature stainless steel and
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 7-1 7. CALIBRATION AND ZERO/SPAN CHECKS There are several ways to check a
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 7-2 Table 7-1: Types of Zero/Span Check and Calibration Section Type of C
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 7-3 7.1. Manual NOX Zero/Span Check or Cal With Zero/Span Gas in the Samp
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 7-4 Enter the expected NOx and NO span gas concentrations: Table 7-3: Ente
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 7-5 Table 7-4: Manual Span Calibration Procedure - Span Gas Thru Sample Po
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 7-6 7.2. Manual O2 Zero/Span Check or Cal With Zero/Span Gas in the Sampl
Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 7-7 Table 7-6: Manual O2 Span Calibration Procedure - Span Gas Thru Sample