Trace Oxygen Analyzer
Teledyne Analytical Instruments
OPERATING INSTRUCTIONS FOR
Model 3000TB
Trace Oxygen Analyzer
P/N M66673
10/25/2011
HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING
SYSTEM.
PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM.
HAZARDOUS VOLTAGES EXIST ON CERTAIN COMPONENTS INTERNALLY WHICH MAY PERSIST
FOR A TIME EVEN AFTER THE POWER IS TURNED OFF AND DISCONNECTED.
ONLY AUTHORIZED PERSONNEL SHOULD CONDUCT MAINTENANCE AND/OR SERVICING. BEFORE
CONDUCTING ANY MAINTENANCE OR SERVICING CONSULT WITH AUTHORIZED SUPERVISOR/
MANAGER.
DANGER
Page view 0
1 2 3 4 5 6 ... 77 78

Summary of Contents

Page 1 - Model 3000TB

iTrace Oxygen AnalyzerTeledyne Analytical InstrumentsOPERATING INSTRUCTIONS FORModel 3000TBTrace Oxygen AnalyzerP/N M6667310/25/2011HIGHLY TOXIC AND

Page 2

1-21 Introduction Model 3000TBTeledyne Analytical Instruments• 316 Stainless steel cell block and sample system—all wettedparts.• Advanced Micro-

Page 3

1-3Trace Oxygen Analyzer Introduction 1Teledyne Analytical Instruments1.5 Operator InterfaceFigure 1-1 is an illustration of the front of the Model

Page 4

1-41 Introduction Model 3000TBTeledyne Analytical InstrumentsFigure 1-1: Model 3000TB—Outer Door Open—Showing Control Panel(Pressing the latch bu

Page 5

1-5Trace Oxygen Analyzer Introduction 1Teledyne Analytical InstrumentsFigure 1-2: Model 3000TB—Inner Door Open—Showing Internal Components

Page 6

1-61 Introduction Model 3000TBTeledyne Analytical Instruments1.5.1 DisplaysDigital Meter Display: The meter display is a LED device thatproduces

Page 7 - COMBUSTIBLE GAS USAGE WARNING

1-7Trace Oxygen Analyzer Introduction 1Teledyne Analytical Instruments• Escape Moves VFD display back to the previous screen in aseries. If none rem

Page 8

1-81 Introduction Model 3000TBTeledyne Analytical Instruments3.0 A MAXFigure 1-3: Electrical Connector PanelElectrical Connections: The electrica

Page 9 - Introduction

1-9Trace Oxygen Analyzer Introduction 1Teledyne Analytical Instruments• RS-232 Port Serial digital concentration signal outputand control input.• Re

Page 10 - 1.4 Model Designations

1-101 Introduction Model 3000TBTeledyne Analytical Instruments• Gas Inlet and Outlet One inlet (must be externally valved)and one exhaust out.Opt

Page 11 - 1.5 Operator Interface

2-1Trace Oxygen Analyzer Operational Theory 2Teledyne Analytical InstrumentsOperational Theory2.1 IntroductionThe analyzer is composed of three subs

Page 12 - (Pressing the latch button

iiModel 3000TBTeledyne Analytical InstrumentsCopyright © 2010 Teledyne Analytical InstrumentsAll Rights Reserved. No part of this manual may be reprod

Page 13

2-22 Operational Theory Model 3000TBTeledyne Analytical Instrumentsanalyzed. The Micro-Fuel Cell is therefore a hybrid between a battery and atru

Page 14

2-3Trace Oxygen Analyzer Operational Theory 2Teledyne Analytical InstrumentsAt the top end of the cell is a diffusion membrane of Teflon, whosethick

Page 15 - 1.7 Equipment Interface

2-42 Operational Theory Model 3000TBTeledyne Analytical InstrumentsThe overall reaction for the fuel cell is the SUM of the half reactionsabove,

Page 16 - 3.0 A MAX

2-5Trace Oxygen Analyzer Operational Theory 2Teledyne Analytical Instruments Figure 2-3. Characteristic Input/Output Curve for a Micro-Fuel Ce

Page 17 - please, see page 1-4

2-62 Operational Theory Model 3000TBTeledyne Analytical InstrumentsFigure 2-4 is the flow diagram for the sampling system. In the standardinstru

Page 18

2-7Trace Oxygen Analyzer Operational Theory 2Teledyne Analytical InstrumentsFigure 2-5: Block Diagram of the Model 3000TB Electronics

Page 19 - Operational Theory

2-82 Operational Theory Model 3000TBTeledyne Analytical InstrumentsIn the presence of oxygen the cell generates a current. A current tovoltage am

Page 20 - Figure 2-1: Micro-Fuel Cell

Trace Oxygen Analyzer Installation 33-1Teledyne Analytical InstrumentsInstallationInstallation of the Model 3000TB Analyzer includes:1. Unpacking2. M

Page 21

3 Installation Model 3000TB3-2Teledyne Analytical InstrumentsFigure 3-1: Front View of the Model 3000TB (Simplified)

Page 22

Trace Oxygen Analyzer Installation 33-3Teledyne Analytical InstrumentsFigure 3-2: Required Front Door Clearance3.3 Electrical ConnectionsFigure 3-3 s

Page 23 - 2.3 Sample System

iiiTrace Oxygen AnalyzerTeledyne Analytical InstrumentsModel 3000TB complies with all of the requirements of theCommonwealth of Europe (CE) for Radio

Page 24 - Exhaust Out

3 Installation Model 3000TB3-4Teledyne Analytical InstrumentsFor safe connections, ensure that no uninsulated wire extends outside ofthe connector

Page 25

Trace Oxygen Analyzer Installation 33-5Teledyne Analytical InstrumentsFigure 3-5: Analog Output ConnectionsThe outputs are:0–1 V dc % of Range: Volta

Page 26

3 Installation Model 3000TB3-6Teledyne Analytical InstrumentsTable 3-1: Analog Concentration Output—ExampleVoltage Signal Current Signalppm O2Outp

Page 27 - Installation

Trace Oxygen Analyzer Installation 33-7Teledyne Analytical InstrumentsFigure 3-6: Types of Relay ContactsThe connectors are:Threshold Alarm 1: • Can

Page 28

3 Installation Model 3000TB3-8Teledyne Analytical InstrumentsZERO: Floating input. 5 to 24 V input across the + and – terminalsputs the analyzer i

Page 29 - 3.3 Electrical Connections

Trace Oxygen Analyzer Installation 33-9Teledyne Analytical InstrumentsNote: The remote probe connections (paragraph 3.3.9) providessignals to ensure

Page 30

3 Installation Model 3000TB3-10Teledyne Analytical Instruments• The range in use (HI, MED, LO)• The span of the range (0-100 ppm, etc)• Which alar

Page 31

Trace Oxygen Analyzer Installation 33-11Teledyne Analytical InstrumentsFigure 3-7: Remote Sensor Connector PinoutsFigure 3-8: Remote Solenoid Return

Page 32

3 Installation Model 3000TB3-12Teledyne Analytical Instruments3.4 Installing the Micro-Fuel CellThe Micro-Fuel Cell is not installed in the cell b

Page 33

Trace Oxygen Analyzer Installation 33-13Teledyne Analytical Instruments2. Hold the fitting body steady with a backup wrench, and withanother wrench r

Page 34

ivModel 3000TBTeledyne Analytical InstrumentsTable of Contents1 Introduction1.1 Overview ...

Page 35

3 Installation Model 3000TB3-14Teledyne Analytical InstrumentsEXHAUST OUT: Exhaust connections must be consistent with thehazard level of the cons

Page 36

Trace Oxygen Analyzer Operation 44-1Teledyne Analytical InstrumentsOperation4.1 IntroductionOnce the analyzer has been installed, it can be configure

Page 37

4 Operation Model 3000TB4-2Teledyne Analytical Instruments4.2 Using the Data Entry and FunctionButtonsData Entry Buttons: The < > arrow butto

Page 38 - 3.5 Gas Connections

Trace Oxygen Analyzer Operation 44-3Teledyne Analytical InstrumentsFigure 4-1: Hierarchy of Functions and Subfunctions4.3 The System FunctionThe su

Page 39 - 4 turns

4 Operation Model 3000TB4-4Teledyne Analytical InstrumentsAfter a password is assigned, the operator must log out toactivate it. Until then, anyone

Page 40 - 3.6 Testing the System

Trace Oxygen Analyzer Operation 44-5Teledyne Analytical InstrumentsIn the first line, TRACK or HOLD should be blinking. The operatorcan toggle betwee

Page 41 - Operation

4 Operation Model 3000TB4-6Teledyne Analytical InstrumentsUse < > arrows to blink Auto—Cal, and press Enter. A new screen forSpan/Zero set ap

Page 42 - Model 3000TB

Trace Oxygen Analyzer Operation 44-7Teledyne Analytical InstrumentsPress System to enter the System mode.TRAK/HLD Auto—CalPSWD Logout MoreUse the <

Page 43 - 4.3 The System Function

4 Operation Model 3000TB4-8Teledyne Analytical InstrumentsPress Enter to change the password (either the default TETAI or thepreviously assigned pa

Page 44

Trace Oxygen Analyzer Operation 44-9Teledyne Analytical Instrumentswill immediately switch to the Analyze screen, and you now have access toall instr

Page 45

vTrace Oxygen AnalyzerTeledyne Analytical Instruments3.4 Installing the Micro-Fuel Cell ... 3-123.5 Gas Connect

Page 46

4 Operation Model 3000TB4-10Teledyne Analytical InstrumentsThe self diagnostics are run automatically by the analyzer whenever theinstrument is tur

Page 47

Trace Oxygen Analyzer Operation 44-11Teledyne Analytical InstrumentsTo show negative oxygen readings on the display:- Press the System keyTRAK/HLD Au

Page 48

4 Operation Model 3000TB4-12Teledyne Analytical InstrumentsShut off the gas pressure before connecting it to the analyzer, andbe sure to limit the

Page 49

Trace Oxygen Analyzer Operation 44-13Teledyne Analytical Instruments#### PPM Zero4 Left=### ppm/sThe zeroing process will automatically co

Page 50

4 Operation Model 3000TB4-14Teledyne Analytical InstrumentsCELL FAIL/ ZERO HIGHBefore replacing the cell:a. Check your span gas to make sure it is

Page 51 - Show_Negative=NO

Trace Oxygen Analyzer Operation 44-15Teledyne Analytical InstrumentsUse the ΔΔΔΔΔ∇ arrow keys to enter the oxygen-concentration mode. Usethe < &g

Page 52

4 Operation Model 3000TB4-16Teledyne Analytical InstrumentsOnce the span has begun, the microprocessor samples the output at apredetermined rate. I

Page 53

Trace Oxygen Analyzer Operation 44-17Teledyne Analytical InstrumentsChapter 3, Installation and/or the Interconnection Diagram included at theback of

Page 54

4 Operation Model 3000TB4-18Teledyne Analytical InstrumentsIf you are using password protection, you will need to enter yourpassword to access the

Page 55

Trace Oxygen Analyzer Operation 44-19Teledyne Analytical Instruments4.6 The Range FunctionThe Range function allows the operator to program up to th

Page 56 - 4.5 The Alarms Function

viModel 3000TBTeledyne Analytical Instruments5.4 Fuse Replacement... 5-75.5 System Self Diagnost

Page 57

4 Operation Model 3000TB4-20Teledyne Analytical InstrumentsNote: The ranges must be increasing from low to high, for example,if range 1 is set as 0

Page 58

Trace Oxygen Analyzer Operation 44-21Teledyne Analytical Instruments4.7 The Analyze FunctionThis is the normal operating mode of the analyzer. In th

Page 59 - 4.6 The Range Function

4 Operation Model 3000TB4-22Teledyne Analytical Instruments

Page 60

Trace Oxygen Analyzer Maintenance 55-1Teledyne Analytical InstrumentsMaintenance5.1 Routine MaintenanceAside from normal cleaning and checking for l

Page 61 - 4.7 The Analyze Function

5 Maintenance Model 3000TB5-2Teledyne Analytical InstrumentsFigure 5-1: Major Internal Components5.3 Cell ReplacementThe Micro-Fuel Cell is a sea

Page 62

Trace Oxygen Analyzer Maintenance 55-3Teledyne Analytical Instruments5.3.1 Storing and Handling Replacement CellsTo have a replacement cell availabl

Page 63 - Maintenance

5 Maintenance Model 3000TB5-4Teledyne Analytical Instrumentsa. Check your span gas to make sure it is within specifications.b. Check for leaks do

Page 64 - 5.3 Cell Replacement

Trace Oxygen Analyzer Maintenance 55-5Teledyne Analytical InstrumentsFigure 5-2: Exploded View of Cell Block and Micro-Fuel CellCell BlockMicro-Fue

Page 65 - 5.3.2 When to Replace a Cell

5 Maintenance Model 3000TB5-6Teledyne Analytical Instruments5.3.4 Installing a New Micro-Fuel CellIt is important to minimize the amount of time

Page 66

Trace Oxygen Analyzer Maintenance 55-7Teledyne Analytical Instruments5.3.5 Cell WarrantyThe Class L-2C Micro-Fuel cell is used in the Model 3000TB.

Page 67 - Outer and inner doors

viiTrace Oxygen AnalyzerTeledyne Analytical InstrumentsCOMBUSTIBLE GAS USAGE WARNINGDANGERThis is a general purpose instrument designed for usage in

Page 68

5 Maintenance Model 3000TB5-8Teledyne Analytical Instruments3.0 A MAXFigure 5-3: Removing Fuse Cap and Fuse from Holder2. Replace fuse by reversi

Page 69 - 5.4 Fuse Replacement

Trace Oxygen Analyzer Maintenance 55-9Teledyne Analytical Instruments5.6 TroubleshootingProblem:Erratic readings of the Oxygen concentration as repo

Page 70

5 Maintenance Model 3000TB5-10Teledyne Analytical InstrumentsIncrease flow rate and/or length or vent tubing in order to dilute of minimizethe di

Page 71 - Problem:

A-1Trace Oxygen Analyzer AppendixTeledyne Analytical InstrumentsAppendixA-1 SpecificationsPackaging: NEMA-4, weather-proof, bulkhead mount.Sensor: L-2

Page 72

A-2Appendix Model 3000TBTeledyne Analytical InstrumentsAccuracy: ±2% of full scale at constant temperature.±5% of full scale over operating tempera

Page 73 - Appendix

A-3Trace Oxygen Analyzer AppendixTeledyne Analytical InstrumentsOrders should be sent to:TELEDYNE ANALYTICAL INSTRUMENTS16830 Chestnut StreetCity of I

Page 74 - Qty. Part Number Description

A-4Appendix Model 3000TBTeledyne Analytical InstrumentsA-53000 SERIES ANALYZERSAPPLICATION NOTES ON RESTRICTORS,PRESSURES, AND FLOW RECOMMENDATIONS

Page 75 - A-3 Drawing List

A-5Trace Oxygen Analyzer AppendixTeledyne Analytical Instruments The second function that the restriction device provides is a pressuredrop. This dev

Page 76

A-6Appendix Model 3000TBTeledyne Analytical InstrumentsBY-PASS:To improve the system response, a by-pass can be added to increase the sample flowra

Page 77

viiiModel 3000TBTeledyne Analytical InstrumentsSpecific Model InformationThe instrument for which this manual was supplied may incorporateone or more

Page 78

1-1Trace Oxygen Analyzer Introduction 1Teledyne Analytical InstrumentsIntroduction1.1 OverviewThe Teledyne Analytical Instruments Model 3000TB Trace