Teledyne MX300 User Manual Page 1

Browse online or download User Manual for Oxygen Equipment Teledyne MX300. Teledyne MX300 User's Manual [el] [es] [pt] [tr]

  • Download
  • Add to my manuals
  • Print
  • Page
    / 50
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 0
Portable Oxygen Monitor
Teledyne Analytical Instruments i
OPERATING & SERVICE INSTRUCTIONS FOR
MX300
PORTABLE OXYGEN MONITOR
TYPE B EQUIPMENT: Equipment providing a particular degree
of protection against electric shock,
particularly regarding—
Allowable LEAKAGE CURRENT
Reliability of the protective earth
connection (if present).
P/N M75387
ECO 03-0166
REV 0
Page view 0
1 2 3 4 5 6 ... 49 50

Summary of Contents

Page 1 - PORTABLE OXYGEN MONITOR

Portable Oxygen Monitor Teledyne Analytical Instruments i OPERATING & SERVICE INSTRUCTIONS FOR MX300 PORTABLE OXYGEN MONITOR

Page 2

Introduction MX300 Teledyne Analytical Instruments 10 Options for alternate display configuration). An integral battery life indicator is displayed

Page 3 - About This Manual

Portable Oxygen Monitor Introduction Teledyne Analytical Instruments 11 1.2 Features The MX300 is a compact, versatile instrument capable of rapi

Page 4 - Safety Messages

Introduction MX300 Teledyne Analytical Instruments 12 Note: Contact the factory for retrofitting an existing instrument for 3-digit LCD display. Fo

Page 5 - Table of Contents

Portable Oxygen Monitor Introduction Teledyne Analytical Instruments 13 activates the audible and visual alarms if the oxygen level exceeds those

Page 6

Introduction MX300 Teledyne Analytical Instruments 14 alarm set points if appropriate. Using a microprocessor allows for easier setting of alarms, a

Page 7 - List of Figures

Portable Oxygen Monitor Operation Teledyne Analytical Instruments 15 Operation Note: Upon receipt, INSPECT THE ENTIRE UNIT FOR DAMAGE. Check the

Page 8 - List of Tables

Operation MX300 Teledyne Analytical Instruments 16 2.1.1 Sensor Installation or Replacement Note: The R17MED oxygen sensor must be installed before

Page 9 - Introduction

Portable Oxygen Monitor Operation Teledyne Analytical Instruments 17 Figure 2-2: Sensor Cable Connection to Monitor Note: When the AX/MX300 in

Page 10 - 1.1 Applicable Standards

Operation MX300 Teledyne Analytical Instruments 18 2.1.2 Mounting The MX300 can be mounted in several ways depending on the optional equipment orde

Page 11 - 1.3 Options

Portable Oxygen Monitor Operation Teledyne Analytical Instruments 19 2.1.3 Battery Installation Note: Three “AA” alkaline batteries must be inst

Page 12 - 1.5 Theory of Operation

MX300 Teledyne Analytical Instruments ii Copyright © 2003 Teledyne Analytical Instruments All Rights Reserved No part of this manual may be reprodu

Page 13 - 1.5.2 Signal Processing

Operation MX300 Teledyne Analytical Instruments 20 4. Re-install the battery compartment door. When the unit is first turned on the display will mo

Page 14

Portable Oxygen Monitor Operation Teledyne Analytical Instruments 21 3. If the LCD is not flashing, press the LOCK/UNLOCK key to unlock the keys

Page 15 - Operation

Operation MX300 Teledyne Analytical Instruments 22 Figure 2-8: Setting the Alarms calibration. Air calibration should only be used for monitoring ox

Page 16

Portable Oxygen Monitor Operation Teledyne Analytical Instruments 23 you to select a LO alarm value less than 18%. Note: The MX300 Portable Oxyg

Page 17

Operation MX300 Teledyne Analytical Instruments 24 Figure 2-9: 0-1 VDC or RS 232 Digital Output Port If your instrument is set for analog (0-1 V

Page 18 - 2.1.2 Mounting

Portable Oxygen Monitor Operation Teledyne Analytical Instruments 25 2.2 Use 2.2.1 Procedure Note: Prior to use, always test the batteries and al

Page 19 - 2.1.3 Battery Installation

Operation MX300 Teledyne Analytical Instruments 26 R17MED sensor so that it does not interfere with the rapid exchange of gases to and from the sens

Page 20 - 2.1.3 Calibration

Portable Oxygen Monitor Operation Teledyne Analytical Instruments 27 2.3.2 Temperature The R17MED oxygen sensor adjusts for ambient temperature c

Page 21

Operation MX300 Teledyne Analytical Instruments 28 2.3.4 Discrepancy in Readings The MX300 instrument should be used to measure the oxygen concentra

Page 22 - 2.1.4 Alarms

Portable Oxygen Monitor Operation Teledyne Analytical Instruments 29 Table 2-1: Oxygen Reading Error in a Mixture of Anesthetic Gas Gas or Vapor

Page 23

Portable Oxygen Monitor Teledyne Analytical Instruments iii About This Manual The MX300 Operator's Manual provides both introductory and de

Page 24 - APPROVED DEVICE

Operation MX300 Teledyne Analytical Instruments 30 After exposure to nitrous oxide mixtures, the sensor should be left in 100% oxygen overnight (e.g

Page 25 - 2.2.1 Procedure

Portable Oxygen Monitor Operation Teledyne Analytical Instruments 31 manufacturer’s instructions). Due to the varying conditions imposed on the m

Page 26 - 2.3.1 Humidity

Operation MX300 Teledyne Analytical Instruments 32 • Recalibrate after replacing the sensor. • Use properly installed alkaline batteries only. •

Page 27 - 2.3.3 Pressure

Portable Oxygen Monitor Operation Teledyne Analytical Instruments 33 • Expose the LCD to excessive sunlight. • Expose the unit to a condensing

Page 28 - 2.3.5 Anesthetic Gases

Operation MX300 Teledyne Analytical Instruments 34 Blank

Page 29

Portable Oxygen Monitor Service Manual Teledyne Analytical Instruments 35 Service Manual 3.1 General Service Information The Teledyne Model MX300

Page 30 - 2.3.6 Cleaning

Service Manual MX300 Teledyne Analytical Instruments 36 The MX300 instrument incorporates a battery test feature that provides the approximate amou

Page 31 - 2.4 Do’s and Don’ts

Portable Oxygen Monitor Service Manual Teledyne Analytical Instruments 37 3.6 Alarms To set the alarms on the MX300 monitor, refer to Section 2.

Page 32

Service Manual MX300 Teledyne Analytical Instruments 38 3.7.3 Pressure Virtually all gas sensors and monitors measure the partial pressure, not the

Page 33

Portable Oxygen Monitor Service Manual Teledyne Analytical Instruments 39 Note: The MSDS on this material is located in the Appendix. It is also

Page 34

MX300 Teledyne Analytical Instruments iv Safety Messages Your safety and the safety of others are very important. Please carefully read the followi

Page 35 - Service Manual

Service Manual MX300 Teledyne Analytical Instruments 40 Symptom Why What To Do The sensor does not react to changes in oxygen concentration, or t

Page 36 - 3.5 Calibration

Portable Oxygen Monitor Service Manual Teledyne Analytical Instruments 41 Symptom Why What To Do The unit has stopped working and the LCD is di

Page 37 - 3.7.2 Temperature

Service Manual MX300 Teledyne Analytical Instruments 42 3.9 Watchdog The MX300 is equipped with a watchdog circuit that continuously monitors the

Page 38 - 3.7.3 Pressure

Portable Oxygen Monitor Service Manual Teledyne Analytical Instruments 43 5 - Silence 6 - High Alarm 7 - Alarm Test 8 - Key Lock 7(2) 7 (The

Page 39 - 3.8 Troubleshooting

Service Manual MX300 Teledyne Analytical Instruments 44 3.11 Return Authorization for Service For any service beyond sensor and battery replacement

Page 40 - Service Manual MX300

Portable Oxygen Monitor Appendix Teledyne Analytical Instruments 45 Appendix A.1 Specifications Range: 0-100% oxygen Accuracy: +2% of full sc

Page 41 - Symptom Why What To Do

Appendix MX300 Teledyne Analytical Instruments 46 A.2 Spare Parts List QTY PART NO DESCRIPTION 1 C43690-R17MED Micro-Fuel Cell R17MED with flow d

Page 42 - 3.9 Watchdog

Portable Oxygen Monitor Appendix Teledyne Analytical Instruments 47 Note: Orders for replacement parts should include the part number (if availa

Page 43

Appendix MX300 Teledyne Analytical Instruments 48 BLANK

Page 44

Portable Oxygen Monitor index Teledyne Analytical Instruments 49 Index accessories, 46 accuracy, 45 address, 44, 47 alarm setpoints, 22 alarm set

Page 45 - Appendix

Portable Oxygen Monitor Teledyne Analytical Instruments v Table of Contents Safety Messages ...

Page 46 - A.3 Optional Accessories

Index MX300 Teledyne Analytical Instruments 50 sensor, 10, 12 sensor cable, 45 sensor failure alarm, 10 sensor life, 45 sensor maintenance, 36 senso

Page 47

MX300 Teledyne Analytical Instruments vi 2.3.1 Humidity 26 2.3.2 Temperature 27 2.3.3 Pressure 27 2.3.4 Discrepancy in Readings 28 2.3.5 Anesth

Page 48

Portable Oxygen Monitor Teledyne Analytical Instruments vii List of Figures Figure 1-1: MX300 Front View...

Page 49

MX300 Teledyne Analytical Instruments viii List of Tables Table 2-1: Oxygen Reading Error in a Mixture of Anesthetic Gas . 29 Table 3-1 Troubleshoo

Page 50 - Index MX300

Portable Oxygen Monitor Introduction Teledyne Analytical Instruments 9 Introduction Teledyne Analytical Instruments MX300 Oxygen Monitor with ala

Comments to this Manuals

No comments