Teledyne 4060 User Manual Page 43

  • Download
  • Add to my manuals
  • Print
  • Page
    / 79
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 42
Series 4060 Operation
Teledyne Analytical Instruments 43
4.3.1 Air
1. Set the air source regulator to 40 psig.
2. Adjust the analyzer air regulator until the air pressure gauge
reads the recommended air pressure of 7.0 psig.
After the air is flowing through the sensor and warm-up time has
been completed, activate the following gases:
4.3.2 Carrier Gas
Set the carrier gas source regulator to 80 psig and adjust the analyzer
sample regulator until the sample pressure gauge reads the
recommended sample pressure of 7.0 psig.
4.3.3 Span Gas
1. Feed span gas to the analyzer (see Section 4.6.5). Gas switching
can also be handled through the autocal valves (if the option is
added), which can also be manually activated as described in
Section 4.6.5.
2. Observe that the analyzer sample flow meter reads from 0.3 to
1.0 SCFH.
4.3.4 Fuel
1. Open the main valve on the fuel source and set the fuel pressure
regulator to 40 psig.
2. Adjust the fuel regulator until its pressure gauge reads the
recommended pressure of 4.0 psig.
Note: Adjust fuel settings only when the red LED (flame failure
light) is off.
4.4 Flame Ignition
Observe that after warm up count down timer reaches zero (timer to
preheat the sensor), the amber heater lamp is blinking (indicating that
the temperature controller is maintaining the temperature setpoint) and
the red Flame Out lamp is on. See Figure 1-1.
The Model 4060 will automatically attempt a flame ignition
sequence following the warm up period which has been preset at the
Page view 42
1 2 ... 38 39 40 41 42 43 44 45 46 47 48 ... 78 79

Comments to this Manuals

No comments