Operation Series 4060
52 Teledyne Analytical Instruments
To initiate a self-test:
<UP/DOWN >From the MAIN MENU scroll to the SELF-TEST
function.
<Enter> Activate the SELF-TEST function by pressing <Enter>.
This brings up the SELF-TEST initialization screen.
________________________________________________
Begin Self-Test?
ENTER=Yes ESCAPE=NO
________________________________________________
<Enter or Escape> Start the diagnostic testing routine by pressing
<Enter> or cancel out by pressing <Escape>.
If you pressed <Enter> the self-test routine will begin and
after a few moments the results will appear onscreen. The
module is functioning properly if it is followed by GOOD
otherwise it is followed by BAD. In this case, please
contact TAI Customer Service for an explanation of the
problem.
To return the analyzer to the MAIN MENU, press
<Enter> after the results screen.
If you pressed <Escape> you will be returned to the
Analyze Mode.
4.6.9 The Alarms Function
The Model 4060 is equipped with two alarms and a system failure
alarm relay. Each alarm relay has a set of form "C" contacts rated for 3
amperes resistive load at 250 VAC. See Figure in Chapter 3, Installation
and/or the Interconnection Diagram included at the back of this manual
for relay terminal connections. The alarm relay contacts are accessible to
the user from 50-pin Equipment Connector. See Section 3.3.2.5 and
Table 3-2.
The system failure alarm has a fixed configuration described in
Chapter 3 Installation.
The concentration alarms can be configured from the ALARM
function screen as follows: